Page 1
GE Healthcare Technical Publications Direction 5272220-100 Rev. 2 LOGIQ C5/C5 PRO Quick Guide Copyright 2008 - 2009 By General Electric Co. Operating Documentation...
Regulatory Requirement LOGIQ C5/C5 PRO complies with regulatory requirements of the following European Directive 93/42/EEC concerning medical devices. Manufacturer GE Medical Systems (China) Co., Ltd. No. 19 Changjiang Road Wuxi National Hi-Tech Development Zone Jiangsu, P.R.China 214028 Tel: +86 510 85225888; Fax: +86 510 85226688...
Please verify that you are using the latest revision of this document. Information pertaining to this document is maintained on MyWorkshop/ ePDM (GE Healthcare electronic Product Data Management). If you need to know the latest revision, contact your distributor, local GE Sales...
Page 4
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 LOGIQ C5/C5 PRO Control Panel Tour 13. Left, Right Key:activate the Left or Right displayed image. Power On/Off Switch 14. Freeze. Press Freeze to freeze the image or 2. Probe Key:switch the probes return to scanning.
Page 5
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 LOGIQ C5/C5 PRO Keys Top/Sub Menu Key Tour Function Keys - Programmable Keys How to Program your hot keys • F1 = Help (Enter Online Help / User Manual) <Utility> - <Admin> - <Function Key>, <Hot Key>...
Page 6
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 LOGIQ C5/C5 PRO Monitor Display Tour 1. Institution/Hospital Name, Date, Time, Operator 11. Measurement Results Window. 21. Card iconTrackball Functionality Status: Scroll, Identification, system status (real-time of M&A (Measurement and Analysis), Position, 12.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 System Power Power On 3. Unwrap the power cable. Make sure to allow LEDs sufficient slack in the cable so that the plug is To connect the system to the electrical supply: not pulled out of the wall if the system is moved 1.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Power Off Starting an Exam To power down the system: You need to select a pre-configured dataflow that sets up the ultrasound system to work according to 1. Press the Power On/Off switch at the front of the services associated to the dataflow.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Starting an Exam New Patient Patient Entry Menu To start a new patient’s exam, Image Management Window [1] EZBackup, EZMove [3] 1. Press Patient. Press the New Patient button on Access to this patient’s exam history and image One-step method to backup (move and delete the Patient menu.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 B/M Mode Image Optimize B Mode Control Panel Controls Power Output Frame Average Auto Optimize Optimizes image quality and allows user to reduce Temporal filter that averages frames together. This Automatic Tissue Optimization optimizes the image beam intensity.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 B/M Mode Image Optimize (continued) B/M Mode Scanning Hints Auto Optimize. Improves imaging performance while reducing optimization time. Frequency. Changes system parameters to best optimize for a particular patient type. Maps. There is an inter-dependency between gray maps, gain, and dynamic range.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Color Flow/Doppler Image Optimize Sample Volume Gate Length 6. Increase Frame Average. 7. Increase the Packet Size. Sizes the sample volume gate. 8. Reduce the ROI to the smallest reasonable Baseline size.
Page 13
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Color Flow/Doppler Image Optimize (continued) Figure 12. PWD Top/Sub Menu Controls Figure 11. CFM Top/Sub Menu Controls...
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Basic Measurements NOTE: The following instructions assume that you Circumference/Area (Ellipse) Measurement Circumference/Area (Trace) Measurement first scan the patient and then press Freeze. 1. Press Measure once; an active caliper 1. Press Measure twice; a trace caliper displays.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Volume Velocity Measurement Worksheets 1. To make a volume calculation, do one of the 1. Press Measure; an active caliper with a vertical Measurement/Calculation worksheets are available following: dotted line displays.
DO NOT use a keyboard. 1. Ensure the LOGIQ C5/C5 PRO is in freeze transducer which appears damaged until 4. Ensure the cable is free. mode. If necessary, press the Freeze key.
Page 17
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Probe Application Table 1: Probe Indications for Use Probe Application E7C-RC E6C-RS 6.5C--RS 4C-RS 3.8C-RC 7.5L-RC 7L-RC 3S-RC Abdomen Small Parts Obstetrics Gynecology Pediatrics Neonatal Urology Cardiac Endocavity Transcranial Intraoperative Vascular...
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Probe Features Table 2: LOGIQ C5/C5 PRO Probe Features Probe Feature E7C-RC E6C-RS 6.5C--RS 4C-RS 3.8C-RC 7.5L-RC 7L-RC 3S-RC LOGIQ View Virtual Convex Virtual Apex Easy 3D Tru Access Using Probes...
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Specification Table 5: System Probe Definitions Doppler Frequency (MHz) Center Image Probe Designation Frequency (MHz) Normal Penetration E7C-RC 6.5 + 10% E6C-RC 6.5 + 10% 6.5C-RC 6.5 + 10% 4C-RC 3.75+10% 3.8C-RC...
Page 20
• DO NOT kink, tightly coil, or apply excessive force on the probe cable. Insulation failure may result. • Electrical leakage checks should be performed on a routine basis by GE Service or qualified hospital personnel. Refer to the service manual for leakage check procedures.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Probe Cleaning, After Each Use Probe Immersion Levels Probe Disinfection Agents 1. Disconnect probe from ultrasound console and Ultrasound probes can be disinfected using liquid remove all coupling gel from probe by wiping chemical germicides.
If the patient name is on the patient record list, Trackball to the patient’s name to highlight the Connectivity on the LOGIQ C5/C5 PRO is based on Select the image on the clipboard, then press the name, (or perform a search to locate the the Dataflow concept.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Configuring Connectivity Login as Administrator. Press the right Utility tab. Device Services (better known as Destinations) Select the Connectivity tab. Configure the menus from left to right, starting with TCP/IP first.
Page 24
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Services (cont’d) Verify a Service (Jdicom Image Storage) Buttons Select the service, press verify button You can assign print buttons to a device or to a e. In the Services drop-down menu, select dataflow.
Page 25
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Dataflow DICOM Status Creates a Dataflow, (‘WL-LA-DServ -- Worklist, To check the status of all DICOM jobs or redirect Local Archive, DICOM* Server, for example). DICOM jobs, press F4 to open the spooler.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Using CINE Activating CINE Start Frame/End Frame Press Freeze, then roll the Trackball to activate Press the Start Frame Softkey to move to the Adjusting the CINE Loop Playback Speed CINE. To start CINE Loop playback, press Run/ beginning of the CINE Loop.
Page 27
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Easy 3D Acquiring a 3D Scan Manipulating the 3D Scan Performing a Surface Render 1. Optimize the B-Mode image. Ensure even gel Imagine you are able to manipulate the 3D volume From the 3D Touch Panel, press 3D, then press coverage.
Page 28
Home Page. Select ‘Operator’ in option of User Level, enter correct password ‘uls’, then press Okay. Figure 22. Service Login Window 2. Now user can enter Global Service User interface. Choose Diagnostics, and then we Figure 23. Service Screen enter LOGIQ C5/C5 PRO Diagnostics menu.
Page 29
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Precaution Levels Icon description Various levels of safety precautions may be found on the equipment and different levels of concern are identified by one of the following flag words and icons which precede the precautionary statement.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Hazard Symbols - Icon Description Potential hazards are indicated by the following icons: Table 5: Potential Hazards Icon Potential Hazard Usage Source Biological • Patient/user infection due to contaminated equipment. • Cleaning and care instructions ISO 7000 •...
Applications training is available through the local GE representative. Added confidence in the equipment operation can be gained by establishing a quality assurance program.
See the Bioeffects section of the Acoustic Output chapter in the Advanced Reference Manual for more information. Training It is recommended that all users receive proper training in applications before performing them in a clinical setting. Please contact the local GE representative for training assistance.
Only approved and recommended peripherals and accessories should be used. All peripherals and accessories must be securely WARNING connected to the LOGIQ C5/C5 PRO. The concerns listed below can seriously affect the safety of equipment and personnel during a diagnostic ultrasound examination.
Page 34
Refer to package labeling to determine latex content and FDA’s March 29, 1991 Medical Alert on latex products. Archived data is managed at the individual sites. Performing data backup (to any device) is recommended. CAUTION Do not use high-frequency surgical equipment with LOGIQ C5/C5PRO system. CAUTION...
Page 35
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Device Labels Label Icon Description The following table describes the purpose and location of safety labels and other important information provided on the equipment Table 6: Label Icons Label/Icon Purpose/Meaning Location Identification and Rating Plate Manufacture’s name and address...
Page 36
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Table 6: Label Icons Label/Icon Purpose/Meaning Location Type CF Defib-Proof Applied Part (heart in the box with paddle) symbol is in accordance with IEC 60878-02- ECG Module This symbol indicates that waste electrical and electronic equipment must not be disposed of as unsorted Bottom municipal waste and must be collected seperately.
Page 37
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Table 6: Label Icons Label/Icon Purpose/Meaning Location There is a pinch point on the LCD monitor. Take care to avoid injuring hands of fingers when flipping down the On the back of LCD LCD monitor.
Page 38
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Classifications Type of protection against electric shock LOGIQ C5/C5 PRO console is Class I Equipment (*1) Degree of protection against electric shock Type BF Applied part (*3) (for Probes marked with BF symbol)
Page 39
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 *4. Type CF Applied Part TYPE CF APPLIED PART providing a degree of protection higher than that for Type BF Applied Part against electric shock particularly regarding allowable LEAKAGE CURRENTS. Table 8: Type CF Equipment...
Page 40
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 EMC (Electromagnetic Compatibility) NOTE: This equipment generates, uses and can radiate radio frequency energy. The equipment may cause radio frequency interference to other medical and non- medical devices and radio communications. To provide reasonable protection against such interference, this product complies with emissions limits for a Group 1, Class A Medical Devices Directive as stated in EN 60601-1-2.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 EMC Performance All types of electronic equipment may characteristically cause electromagnetic interference with other equipment, either transmitted through air or connecting cables. The term EMC (Electromagnetic Compatibility) indicates the capability of equipment to curb electromagnetic influence from other equipment and at the same time not affect other equipment with similar electromagnetic radiation from itself.
2. Locate the equipment as far away as possible from other electronic equipment. 3. Be sure to use only the cables provided by or designated by GE Medical Systems. Connect these cables following the installation procedures (i.e. wire power cables separately from signal cables).
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 General Notice 1. Designation of Peripheral Equipment Connectable to This Product. The equipment indicated on Chapter 3 of the Basic User Manual can be hooked up to the product without compromising its EMC performance.
Direction 5272220-100 Rev. 2 Peripheral Update for EC countries The following is intended to provide the users in EC countries with updated information concerning the connection of the LOGIQ C5/C5 PRO to image recording and other devices or communication networks.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Peripheral Update for EC countries (continued) Declaration of Emissions This system is suitable for use in the following environment. The user must assure that it is used only in the electromagnetic environment as specified.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Declaration of Immunity This system is suitable for use in the following environment. The user must assure that the system is used according to the specified guidance and only in the electromagnetic environment listed.
If devices are connected without the approval of GE, the warranty will be INVALID. Any device connected to the LOGIQ C5/C5 PRO must conform to one or more of the requirements listed below: 1. IEC standard or equivalent standards appropriate to devices.
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Acoustic Output Located on the upper right section of the system display monitor, the acoustic output display provides the operator with a real-time indication of acoustic levels being generated by the system. See the Acoustic Output chapter in the Advanced Reference Manual for more information. This display is based on NEMA/AIUM Standards for Real-time Display of Thermal and Mechanic Acoustic Output Indices on Diagnostic Ultrasound Equipment.
Acoustic Output Default Levels In order to ensure that an exam does not start at a high output level, the LOGIQ C5/C5 PRO initiates scanning at a reduced default output level. This reduced level is a programmable preset and depends upon the exam category and probe selected. It takes effect when the system is powered on or New Patient is selected.
Page 51
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Warning Label Locations Console Labels Figure 25. Label Location 1. Possible shock hazard. Do not remove covers or panels. No user serviceable parts are inside. Refer servicing to qualified service personnel.
Page 56
LOGIQ C5/C5 PRO Quick Guide Direction 5272220-100 Rev. 2 Fax Back Form You can order printed documentation by faxing this form to Coakley-Tech. Customer Name Telephone Number Mailing Address Fax To: Fax Number: Attention: Coakley-Tech (414) 389-9130 Norm Keene In the publications list below, please select only those printed manuals required.