GE Healthcare LOGIQ C3/C5 Premium Basic Service Manual Part Number: 5341787-100 Revision: 8...
Page 3
• VERSUCHEN SIE NICHT, DAS GERÄT ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
Page 4
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
Page 5
ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝΔΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
Page 6
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN...
Page 7
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL ŠĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS...
Page 8
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
Page 9
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA.
Page 10
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
Page 11
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL (ZH-CN) (KO)
Page 12
GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 13
Copyright and Trade Secret laws; the violation of which can result in civil damages and criminal prosecution. If you are not party to such a license agreement or a GE Healthcare Employee, you must exit this Material now.
Page 14
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Revision History Revision Date Reason for change 2009/05/02 Initial Release. 2010/03/29 Update the Renewal Part List 2010/05/27 Add PHI information 2011/05/18 Add probe new probes information...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 1 Introduction Overview............1 - 1 Purpose of Chapter 1 .
Page 16
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 2 Site Preparations Overview ............2 - 1 Purpose of chapter 2 .
Page 17
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 3 System Setup Overview............3 - 1 Purpose of Chapter 3 .
Page 18
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose for Chapter 4 .
Page 19
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview............5 - 1 Purpose of Chapter 5 .
Page 20
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of this chapter 6 .
Page 21
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
Page 22
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
Page 23
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ C3/C5 Premium. The service provider must read and understand all the information presented in this manual before installing or servicing a unit.
LOGIQ C3/C5 Premium Models Covered by this Manual Table 1-2 LOGIQ C3 Model Designations (For Software Version R2.0.1, R 2.0.2, R2.0.3 ) Part Number Description 5346769 LOGIQ C3 Premium Console for Europe 5346771 LOGIQ C3 Premium Console for India 5346772 LOGIQ C3 Premium Console for Argentina 5346774...
Page 27
Part Number Description 5433023 LOGIQ C3 Premium Console for Europe (220V) 5433034 LOGIQ C3 Premium Console for Latin America (220V) 5433035 LOGIQ C3 Premium Console for Latin America (110V) Table 1-7 LOGIQ C5 Model Designations (For Software Version R2.1.0) Part Number...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. Table 1-10: Product Icons...
Page 31
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 1-10: Product Icons Purpose/Meaning Location Label/Icon This symbol indicates that Bottom waste electrical and electronic equipment must not be disposed of as unsorted municipal waste and must be collected separately.
Page 32
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 1-10: Product Icons Purpose/Meaning Location Label/Icon Do not place any objects on Back of LCD the monitor. Do not force the LCD monitor Back of LCD with your hands.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-3-3 Mechanical Safety (cont’d) NOTE: Special care should be taken when transporting the unit in a vehicle: • Secure the unit in an upright position. •...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 1-4 EMC, EMI, and ESD 1-4-1 Electromagnetic Compatibility (EMC) Electromagnetic compatibility describes a level of performance of a device within its electromagnetic environment. This environment consists of the device itself and its surroundings including other equipment, power sources and persons with which the device must interface.
Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
ANUAL 1-6-2 System Manufacturer Table 1-12 System Manufacturer Manufacturer FAX Number GE Medical Systems (China) Co., Ltd. TEL: +86 510-85225888 No.19, Changjiang Road, Wuxi National Hi-Tech Development Zone, FAX: +86 510-85226688 Jiangsu, P.R. China 214028 1-6-3 Factory Site Table 1-13...
Page 40
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 1 - 16 Section 1-6 - Customer Assistance...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 2 Site Preparations Section 2-1 Overview 2-1-1 Purpose of chapter 2 This chapter provides the information required to plan and prepare for the installation of LOGIQ C3/C5 Premium.Included are descriptions of the facility and electrical needs to be met by the purchaser of the...
Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
Page 43
2-2-2-5 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
Never place a label where RF gaskets meet the unit. Otherwise, the gap created will permit gaskets touch metal RF leakage. Or, if a label has been found in such a position, move the label. Use GE specified harnesses The interconnect cables are grounded and require ferrite beads and other shielding. Also, and peripherals cable length, material, and routing are all important;...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation:10° to 40° C Storage:-5° to 50° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32).
All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-3-3 Recommended and Alternate Ultrasound Room Layout Recommended standard floor plan and a minimal floor plan for ultrasound equipment: DEDICATED ALALOG TELEPHONE LINE FOR CONNECTION TO INSITE 18 IN.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-3-4 Networking Pre-installation Requirements 2-3-4-1 Stand Alone Scanner (without Network Connection) None. 2-3-4-2 Scanner Connected to Hospital’s Network Supported networks: Wire LAN 2-3-4-3 Purpose of DICOM Network Function DICOM services provide the operator with clinically useful features for moving images and patient information over a hospital network.
Page 50
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) LOGIQ C3/ IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses...
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 3 System Setup Section 3-1 Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-2Setup Reminders 3-2-1 Average Installation Time Table 3-2 Average Installation Time Description Average Installation Time Comments 0.5 hour Unpacking the scanner 0.5 hour Scanner wo/options Dependant on the configuration that is required 0.5 hour...
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the LOGIQ C3/C5 Premium weighs approximately 75 kg without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-2-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE DANGER AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH THE UNIT! Two people should unpack the unit because of its weight.
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment When a new system arrives, check that any components are not damaged and are not in short supply. If shipping damage or shortage occurs, contact the address shown in Chapter 1.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-3-1 Moving into Position Do not tilt the unit more than 5 degrees to avoid tipping it over. CAUTION To avoid injury by tipping over. Set the monitor to the lowest position before moving.
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-4 Preparing for Installation 3-4-1 Verify Customer Order Compare items received by the customer to that which is listed on the delivery order. Report any items that are missing, back ordered or damaged.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-5-2-1 Back-end Processor Power Down (cont’d) 2.) The System-Exit window is displayed. Figure 3-5 System Exit Window 3.) Using the Trackball or Select key, select Shutdown. 4.) The shutdown process takes a few seconds and is complete when the power status LED is turned amber.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-6 System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the LOGIQ C3/C5 Premium unit are summarized in Table 3-4 on page 3-10 The Size of LOGIQ C3/C5 Premium.
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-1-2 Weight Table 3-5 Weight of LOGIQ C3/C5 Premium With Monitor and Without Other Peripherals Model Weight [kg] Weight [lb] LOGIQ C3/C5 Approximately 75 Approximately 165 Premium 3-6-2...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-4 Connecting Cables Equipment damage possibility. Be sure to use the following recommended connecting cables to WARNING WARNING connect recording devices and a network with LOGIQ C3/C5 Premium console.
Page 63
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-2 Left Panel Connector 1 USB port is Located on the left panel. Figure 3-7 Left Connector Panel 1.) 1 USB Ports (Printers, Footswitch, etc) 3-6-5-3 Printer Box Rear Panel connector Figure 3-8 Printer Box Rear Connector Panel 1.) 1 USB Ports only for B/W Printer...
Page 64
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-4 Pin assignment for each connector Table 3-9 Pin Assignments of External VGA Pin No. Signal Pin No. Signal GREEN BLUE Table 3-10 Pin Assignments of USB Pin No.
Page 65
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-5 Connect Peripherals A.) Connect B/W printer to the system. B/W Printer can be properly connected using USB Ports. Figure 3-9 Connect B/W printer to the system B.) Connect UP-D23 color printer to the system.UP-D23 Color Printer can be properly connected using USB Ports.
Page 66
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-5 Connect Peripherals (cont’d) C.) Connect the B/W printer (Sony UP-897MD)to the system The B/W printer can be properly connected using the Composite Out Port. Figure 3-11 Connect B/W Printer (Sony UP-897MD) to the system D.) Connect Foot Switch to the system.
Page 67
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-5 Connect Peripherals (cont’d) E.) Connect the USB Memory to the system. The USB Memory can be properly connected using USB ports. Figure 3-13 USB Memory Connection F.) ECG can be properly connected using USB ports. (ECG only supported by LOGIQ C5 Premium)
Page 68
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-5 Connect Peripherals (cont’d) G.) Connect HP 470 Officejet printer to the system. HP 470b Officejet Printer can be properly connected using USB Ports. Figure 3-15 HP 470b Officejet printer to the system H.) Connect the USB HDD to the system.
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-6 Available Probes See in specification in the LOGIQ C3/C5 Premium User Reference Manual for Probes and intended use. Table 3-14 List of Probes on LOGIQ C3/C5 Premium...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-8 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
GE H VIVID FIVE EALTHCARE 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-9 Loading Base Image Software Refer to Section 8-4 "Loading Base Image Software" on page 8-6. Chapter 3 System Setup 3 - 21...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-10Software Version check out 3-10-1 Functional Check-out 1.) Power on LOGIQ C3/C5 Premium scanner and wait until system booting to main screen. 2.) Press UTILITY key on control panel.
User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 System Setup...
Page 74
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 3 - 24 Section 3-11 - Paperwork...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ C3/C5 Premium console, diagnostics by using the built-in service software, and power supply.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes.
Page 77
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-2-1Back-end Processor Power Down (cont’d) 2.) The System-Exit window is displayed. Figure 4-20 System Exit Window 3.) Using the Trackball or Select key, select Shutdown. 4.) The shutdown process takes a few seconds and the power off sequence is complete when the power status LED is turned amber.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-3-2 Loading Presets from an CD-R disk (or USB memory device) 1.) Insert the CD-R disk with the archived Presets into the DVD-RW. 2.) Access to the Utility Menu, and select System. The Restore sheet will be shown on the LCD display.
10.)Ellipse key 11.)Depth key 12.)Reverse, Print, Store keys 13.)Left, Right key 14.)Freeze key 15.)Trackball 16.)Imaging/Measurement keys 17.)TGC 18.)Gel Holder 19.)Probe Holder NOTE: LOGIQ C3 Premium does not support PW Mode and CF Mode. 4 - 6 Section 4-3 - General Procedure...
Page 81
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-5-2 LOGIQ C3/C5 Premium SoftMenu Key Tour 1.) Top Menu Controls op Menu Controls Figure 4-23 SoftMenu Key Tour 2.) Sub menu Controls: Activates Sub Menu for Modes and toggles/changes functions A.) Press F5 on the keyboard to activate Sub Menu,...
Page 82
13. Imaging Parameters by Mode (current mode highlighted). 3. Acoustic Output Readout, 14. Focal Zone. 4. GE Symbol: Probe Orientation Marker. Coincides with a 15. Trackball Functionality Status: Scroll, M&A (Measurement probe orientation marking on the probe. and Analysis), Position, Size, Scan Area Width and Tilt.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-6 B Mode Checks 4-3-6-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-19, in Chapter 3 Installation to the System probe connector.
Page 84
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-6-2 B Mode OP Panel Controls Table 4-3 B Mode Control Panel Controls Step Task Expected Result(s) Press B Mode key B Mode Starts Adjust the field of view. Increasing the depth...
Page 85
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-6-3 B Mode Softmenu Table 4-4 B Mode Softmenu Step Task Expected Result(s) Enables gray scale image colorization. To Activate Colorize deactivate, reselect a Gray Map.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-7 M Mode Controls 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-19, in Chapter 3 Installation to the System probe connector.
Page 87
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-7-2 M Mode OP Panel Controls Table 4-5 M Mode OP Panel Controls Step Task Expected Result(s) Press M Mode key M Mode Starts Controls the amount of echo information displayed in an image.
Page 88
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-7-3 M Mode Softmenu Table 4-6 M Mode Softmenu Step Task Expected Result(s) Changes the speed at which the time line is Adjust Sweep Speed swept. The following speed values are available, 1, 2, 3, 4, 6, 8, 12, 16.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-8 Doppler Mode Checks 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-19, in Chapter 3 Installation to the System probe connector.
Page 90
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-8-2 Doppler Mode OP Panel Controls Table 4-7 Doppler Mode OP Panel Controls Step Task Expected Result(s) Press PW Mode key PW Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
Page 91
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-8-3 Doppler Mode OP Panel Controls Table 4-8 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Selects a level below which echoes will not be...
Page 92
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-8 Doppler Mode Touch Panel Controls Step Task Expected Result(s) Velocity scale determines pulse repetition frequency. If the sample volume gate range Change PRF (Pulse Repetition Frequencies)
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-9 Color Flow Mode Checks NOTE: Color Flow Mode only support by LOGIQ C5 Premium 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-19, in Chapter 3 Installation to the System probe connector.
Page 94
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-9-2 Color Flow Mode OP Panel Controls Table 4-9 Color Flow Mode OP Panel Controls Step Task Expected Result(s) Press CFM-Mode key CFM Mode Starts Amplifies the overall strength of the echoes processed in the Color Flow window.
Page 95
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-9-3 Color Flow Mode Softmenu Key Table 4-10 Color Flow Mode Softmenu Key Step Task Expected Result(s) Threshold assigns the gray scale level at Threshold which color information stops.
Page 96
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-10 Color Flow Mode Softmenu Key Step Task Expected Result(s) Optimizes image quality and allows user to reduce beam intensity. 10% increments Power output between 0-100%. Values greater than 0.1 are displayed.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-10 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-10-1 Distance and Tissue Depth Measurements 1.) Press MEASURE once; an active caliper displays.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-10-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. The worksheets are selected from the Measurement Touch Panel.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-12-2 Quickly Move to Start/End Frame Press First to move to the first CINE frame; press Last to move to the last CINE frame. 4-3-12-3 Start Frame/End Frame Press the Start Frame Two-Button Softkey to move to the beginning of the CINE Loop.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14 Backup and Restore Database, Preset Configurations and Images 4-3-14-1 Formatting Media NOTE: The system may crash if use a DVD for Archive/Backup/EZBackup/Restore, so always use a CD-R or USB HDD for Archive/Backup/EZBackup/Restore.You only can use DVD for upgrade.
Page 101
4.) On the LCD display, select BACKUP/RESTORE. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 5.) In the Backup list, select Patient Archive, Report Archive and User Defined Configuration.
Page 102
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-2Backup System Presets and Configurations (cont’d) Check here to backup presets and configurations Figure 4-36 Backup/Restore Menu 4 - 28 Section 4-3 - General Procedure...
Page 103
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-3 Restore System Presets and Configurations The restore procedure overwrites the existing database on the local hard drive. Make sure to insert CAUTION the correct CD (or USB memory device).
Page 104
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-4 Archiving Images 1.) Insert the archive media. To format the archive media, select the Utility button on the Keyboard. 2.) Select Connectivity, then Removable Media.
Page 105
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-4 Archiving Images (cont’d) 7.) Press PATIENT and set the Dataflow to store images directly to CD-ROM. 8.) From the image screen, press EZBACKUP/MOVE. The Move Images pop-up appears.
Page 106
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-4 Archiving Images (cont’d) 11.)Press Finish after Backup/Move complete. Figure 4-42 EZBackup/Move complete All databases, presets and images should now be saved to removable media. 4-3-14-5 Full Backup 1.) Connect the USB HDD to system.
Page 107
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 4.) Return to scan mode. Press Ctrl + B, the Emergency Disk Making window pop-up appears. Figure 4-44 Emergency Disk Making Pop-up Menu 5.) Press Next on pop-up message.
Page 108
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 7.) Press Finish after Emergency Disk Making complete. Figure 4-46 Emergency Disk Making Complete All databases, presets and images should now be saved to Emergency Disk.
Page 109
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-14-5 Full Backup (cont’d) 10.)A recovery information message appears. Press Next to continue. Figure 4-48 Recovery Information message 11.)Press Finish after Emergency Disk Recovery complete. Figure 4-49 Emergency Disk Recovery Complete...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-11 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ C3/C5 Premium' system concepts, component arrangement, and subsystem function. Table 5-1...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 5-2Block Diagram 5-2-1 System Diagram Figure 5-50 LOGIQ C3/C5 Premium System Diagram 5 - 2 Chapter 5 Components and Functions (Theory)
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 5-3 Common Service Platform 5-3-1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. The Common Service Platform will increase service productivity and reduce training and service costs.
Page 115
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
Page 116
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 6-2 Monitor Adjustments 6-2-1 Adjustments Procedures To adjust the brightness: For LOGIQ C3/LC5 Premium: Adjust the LCD monitor’s button, located the right side of the LCD Monitor.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-2 Gathering Trouble Data 7-2-1 Overview There may be a time when it would be advantageous to capture trouble images and system data (logs) for acquisition to be sent back to the manufacturer for analysis. There are different options to acquire this data that would give different results.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-2-3 Collect a Trouble Image with Logs If the system should malfunction, press the Alt-D keys simultaneously. This will collect a screen capture of the image monitor, system presets and the following logs:...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-3 USB Quick Save 7-3-1 Overview There may be times when the customer or field engineer will want to directly save images into USB memory. This is accomplished by saving individual Cine clips (moving images- avi format) or still images (jpg format) directly to a USB memory disk by pressing Print or Store Key.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-3-3 Setting the Print or Store Key to USB Quick Save If the Print or Store Key is not set to screen capture: 1.) While on the Connect screen, with the Buttons tab displayed, go to the Destinations list.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-4 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print or Store Key.
Page 123
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Clipboard Thumbnai Menu Figure 7-56 Select Image to Capture 3.) Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-4-3 Capturing a Screen (cont’d) Figure 7-58 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk or CD-R.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-5 Common Diagnostics 7-5-1 Utilities Provides two selections: 7-5-1-1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting. 7-5-1-2 System Shutdown Allows for system shutdown from the diagnostic menu. Select to Restart System or Shutdown System.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-5-2 PC Diagnostics (Non-Interactive Tests) 7-5-2-1 Essential Tests 7-5-2-2 System Board 7-5-2-3 Hard Disk Surface Scan 7-5-2-4 Hard Disk Quick Test 7-5-2-5 PCIBus 7-5-2-6 Video Card...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-5-3 PC Diagnostics (Interactive Tests) 7-5-3-1 Keyboard Test 7-5-3-2 Mouse Test 7-5-4 Restart LOGIQ C3/C5 Premium After Diagnostics Always shutdown the system and reboot after a diagnostics session.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-6 Network Configuration 7-6-1 Network Configuration 7-6-1-1 Wire-LAN Network 1.) Connect system with network. 2.) Press Utility -> Connectivity -> TCPIP, in IP Settings window check the Enable DHCP box. In Network Speed box, choose the proper speed available.
Page 129
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-6-1-1 Wire-LAN Network (cont’d) NOTE: If user want to setup static IP address, uncheck Enable DHCP option, input static address in IP- Address box, Subnet Mask and Default Gateway box. In Network Speed box, choose the proper speed available.
Page 130
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 7 - 14 Section 7-6 - Network Configuration...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-1 Contents in Chapter 8...
C5 Premium, GE will contact the customer and request for urgent collection of that patient information. GE will keep this patient information in a secure environment for a maximum period of 1 month. All patient information will be permanently deleted at that point.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 8-2-3 Air filter and Knob Screw 8-2-3-1 Tools • Common phillips screwdrivers 8-2-3-2 Preparations • Shut Down the System and disconnect the power cord. 8-2-3-3 Removal Procedure 1.) Unscrew the knob screw which fix air filter on the edge of power box, refer to 8-2-3 "Air filter and...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-3 Trackball Roller Cleaning Purpose: This is a description on how to remove the trackball and clean the roller. 8-3-0-1 Tools • No need. 8-3-0-2 Needed Manpower •...
Page 135
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-3 Trackball Roller Cleaning (cont’d) Figure 8-64 Trackball Roller Cleaning Section 8-3 - Trackball Roller Cleaning 8 - 5...
Page 136
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-4Loading Base Image Software NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-4 Loading Base Image Software (cont’d) 6.) Select one of the options for loading the system. Select choice 1] to load the complete disk. Figure 8-66 Selection for loading the system...
Page 138
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-4 Loading Base Image Software (cont’d) 8.) System CD will be loaded as shown in the screen below. Figure 8-68 Upgrading message 9.) System updating finished, refer to...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ C3/C5 Premium. Table 9-1 Contents in Chapter 9...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-3Renewal Parts Lists 9-3-1 Equipment Models Covered in this Chapter Table 9-2 Renewal Parts List for Customer Part Name Part Number Description Quantity AC Power Cord 5177123...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-5Accessories and kits Table 9-3 Accessories ,Cables and Kits Item Part Number Description Cable Kit for Upper Part (including the following cables) 5332380 Cable from KBDTRSF board to Main...
Page 143
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 9-3 Accessories ,Cables and Kits Item Part Number Description Service Tools for LOGIQ C (including the following tools) 5262716 Caster Spanner 5003 5328133 5324751 TBJMA Screw Driver FJ6-40...
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-6Manuals Item Part Number Description 6000 5341787-100 LOGIQ C3/C5 Premium Service Manual System User Manuals 6001 5341361 -100 User Manual, LOGIQ C3/C5 Premium, English 6002 5341361 -101...
Page 145
GE H VIVID FIVE EALTHCARE 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-7Probes Table 9-4 Probes on LOGIQ C3/C5 Premium Item Part Name Part Number Description Quantity 7000 4C-RC 5315581 Probe (Center Frequency: 3.75MHz) 7001 7.5C-RC 5315683 Probe (Center Frequency: 7.5MHz) 7002 6.5C-RC...
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
Page 149
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 10-3 Maintenance Task Schedule (cont’d) Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes •...
Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
Page 151
EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Part Number Comments...
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-3 Input Power 10-4-3-1 Mains Cable Inspection Table 10-6 Mains Cable Inspection Step Item Description Unplug Cord Disconnect the mains cable from the wall and system.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-6 Outlet Test -Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangment by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-7 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test. CAUTION Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-8 Chassis Leakage Current Test 10-4-8-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-9 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1. Select and test each of the five ECG lead positions (except ALL) on the LEAD selector, testing each to the power condition combinations found in the table.
Page 158
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-10-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate.
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-11 Probe Leakage Current Test 10-4-11-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
Page 160
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-11-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values. Record all data on the PM Inspection Certificate.
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 10-5 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
Page 162
Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________...
Page 163
GE H PROPRIETARY TO GE LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
Page 164
GE H LOGIQ C3/C5 P EALTHCARE REMIUM 5341787-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 10 - 18 Section 10-5 - When There's Too Much Leakage Current...
Page 165