Page 3
MyWorkshop/ePDM (GE electronic Product Data Management). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
Page 4
Important precautions Translation policy LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 5
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 6
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 7
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 8
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 9
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 10
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 11
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 12
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 13
LOGIQ V3/V5/V5 Expert – Basic Service Manual i-11 5726264-100 English Rev.8...
GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Wuxi National Hi-Tech Dev. Zone Jiangsu P.R.China 214028 TEL: +86 510 85225888; FAX: +86 510 85226688 GE employees should use TrackWise to report service documentation issues. These issues will then be in the internal problem reporting tool and communicated to the writer.
For a complete review of all safety requirements, refer to Chapter 1 in the Service Manual. i-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
If you are a GE employee or a customer who has entered into such a license agreement with GE to use this proprietary software, you are authorized to use this Material according to the conditions stated in your license agreement.
Page 18
LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
This chapter describes important issues related to safely servicing the Ultrasound system. The service provider must read and understand all the information presented here before installing or servicing the units. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Manual Overview 1-2 Manual Overview This manual provides installation and service information for the LOGIQ V3/V5/V5 Expert Ultrasound system. It is divided in ten chapters as shown below. 1-2-1 Contents in this manual The manual is divided into ten chapters.
Licensed Hospital’s Service Providers 1-2-3 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ V5/LOGIQ V3 and also kept near the unit for quick reference. NOTE: Probe information displayed on screen does not necessarily reflect the probes available on your ultrasound system.
Manual Overview 1-2-4 LOGIQ V3/V5/V5 Expert models covered by this manual Table 1-2: LOGIQ V3/V5/V5 Expert Model Designations Model Number Description System SW H48612AA LOGIQ V3 R2.x.x H48612AB LOGIQ V5 R2.x.x H48612AC LOGIQ V5 EXPERT R2.x.x NOTE: When not otherwise specified, the contents of this manual applies to all LOGIQ V3/V5/V5 Expert models.
Important conventions, used in this document, are described next. 1-3-1-1 Model designations This manual covers the LOGIQ V3/V5/V5 Expert Ultrasound systems listed in: 1-2-4 ‘LOGIQ V3/V5/V5 Expert models covered by this manual’ on page 1-5. 1-3-1-2 Icons Pictures, or icons, are used wherever they will reinforce the printed message.
Page 31
Notes are used to provide important information about an item or a procedure. NOTE: Be sure to read the notes; the information contained in a note can often save you time or effort. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Standard hazard icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH NOTE: Even if a symbol isn’t used on the product or in this manual, it may be included for your reference. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 33
Tag and Lock Out Wear Eye Protection Hand Protection Foot Protection Wear Eye Protection Be sure to read the notes; the information contained in a note can often save you time or effort. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
1-4 Product icons It is important to refer to the current revision of the Ultrasound system’s Basic User Manual for a full list of product labels prior to servicing the system. 1-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Basic User Manual for a full list of product labels prior to servicing the system. 1-5-1 Rating Plate Location LOGIQ V3/V5/V5 Expert labels are provided in English. The labels are at the bottom of the system. The label content may be different for different regions. Please refer to the labels on the system for the actual content.
Page 39
Rating plate Serial Number. NOTE: If the new label is needed during the service activities, please click “Ask an Expert“ to submit the case in the support central: http://supportcentral.ge.com/products/sup_products.asp?prod_ id=44177. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-15 5726264-100 English Rev.8...
Operating personnel must not remove the system covers. • Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ V3/V5/V5 Expert Training Seminar are authorized to service the equipment. 1-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual...
Page 41
Because of the limited access to cabinets and equipment in the WARNING field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
Page 42
WARNING sharp edges are encountered, the appropriate PPE should be used to reduce the risk of injury. Wear all PPE including gloves as indicated in the chemical WARNING MSDS. 1-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Take extra care when moving the system. CAUTION The LOGIQ V3/V5/V5 Expert weighs approximately 65 kg (144 lbs) or more, depending on installed peripherals, when ready for use. To avoid possible injury and equipment damage when transporting from one area of use to another: •...
Page 44
Do not move the Ultrasound system if the Operator Panel is in CAUTION unlocked position. Do not transport LOGIQ V3/V5/V5 Expert in a vehicle without CAUTION locking the casters (wheels) and securing it as described in chapter 4.
Page 45
Secure system with straps or as directed otherwise to prevent motion during transport. • Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-21 5726264-100 English Rev.8...
• Both the system power cable and the power connector must meet international electrical standards. Connecting a LOGIQ V3/V5/V5 Expert to the wrong voltage WARNING level will most likely destroy it. DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU WARNING UNIT LEVEL AT ANY CIRCUMSTANCES.
Page 47
The system has more than one type of probe port. Use the appropriate probe port designed for the probe you are connecting. 1-6-5-3 Peripherals Refer to the Patient Safety Environment section of the User’s Manual for peripheral isolation information. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-23 5726264-100 English Rev.8...
DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. If the covers are removed from an operating LOGIQ V3/V5/V5 WARNING Expert, some metal surfaces may be warm enough to pose a potential heat hazard if touched, even while in shutdown mode.
4. Apply Lockout/Tagout Devices. 5. Control all stored and residual energy. 6. Verify isolation. All potentially hazardous stored or residual energy is relieved. Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1.
It is always possible during system failure and repair to lose patient data. GE is not responsible for the loss of this data. If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer.
1-10-2 CE Compliance LOGIQ V3/V5/V5 Expert conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
Always use non-conductive handles designed for the removal and replacement of ESD sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before making contact. 1-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
5. Is the issue repeatable? 6. Imaging mode, probe, preset/application. 7. Media brand, speed, capacity, type. 8. Save secondary image capture, cine loop. NOTE: Restart the application before resuming clinical scanning. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-29 5726264-100 English Rev.8...
System manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE Medical Systems (China) Co., Ltd. +86 510 85225888 +86 510 85226688 No.19 Changjiang Road WuXi National Hi-Tech Development Zone Jiangsu P.R.China 214028 1-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
+49 761 45 43 -0 / Manual. +49 761 45 43 -233 Authorized EU Representative/European registered place of business: GE Medical Systems Information Technologies GmbH (GEMS IT GmbH) Munzinger Strasse 5, D-79111 Freiburg, GERMANY 1-11-5 Factory Site Table 1-9: Factory Site...
Page 56
Introduction 1-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
This chapter provides the information required to plan and prepare for the setup of an Ultrasound system. Included are descriptions of the facility and electrical needs to be met by the purchaser of the units. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Site Preparations 2-1 Overview 2-1-1 Contents in this chapter • 2-1 ‘Overview’ on page 2-2 • 2-2 ‘General Ultrasound system requirements’ on page 2-3 • 2-3 ‘Facility needs’ on page 2-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Table 2-1: System Acclimation Time Chart Degree C -4.5 42.5 47.5 Degree F 23.9 28.4 32.9 37.4 108.5 117.5 hours LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 60
C * 9/5) + 32. Ensure that the probe face temperature does not exceed the CAUTION normal operation temperature range. The LOGIQ V3/V5/V5 Expert system and probe connector is CAUTION not waterproof. Do not expose the device to water or any kind of liquid.
Page 61
(dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality. These controls should be selected to minimize possible interference. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Site Preparations 2-2-3 Electrical requirements 2-2-3-1 General requirements NOTE: GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system. Sites with a mains power system with defined Neutral and Live:...
Page 63
2-2-3-6 Unit power plug If the Ultrasound system arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. LOGIQ V3/V5/V5 Expert – Basic Service Manual...
Page 64
70%T (30 dip) for 25 70%T (30 dip) for 25 environment. cycles; cycles; < 5%T (>95% dip) for 5 < 5%T (>95% dip) for 5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
AC/DC adapters • portable phones The presence of a broadcast station or broadcast van may also cause interference. See: 2-2-4-1 ‘EMI prevention/abatement’ on page 2-10 for EMI prevention tips. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 66
Where applicable, do not allow cables to lie across the top of the Card Rack or peripheral cables hang out of the peripheral bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to the frame. 2-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
2-2-6 Time and manpower requirements Site preparation takes time. Begin site preparation checks as soon as possible, if possible, six weeks before delivery, to allow enough time to make any changes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-11 5726264-100 English Rev.8...
All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment. The desire to use a non–listed or customer provided product or...
Site Preparations 2-3-3 Required facility needs NOTE: GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system. Sites with a mains power system with defined Neutral and Live:...
Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-15 5726264-100 English Rev.8...
Dual Level Lighting (bright of Ultrasound system from inches) and dim) wall or objects 7. Emergency Oxygen 13. Stool Figure 2-2. A 14 by 17 foot recommended floor plan LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-17 5726264-100 English Rev.8...
With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost. 2-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 75
Ultrasound system for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the revision of the device, is also included. This information may be useful for error solving. LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-19 5726264-100 English Rev.8...
Page 76
Site Preparations 2-3-7-5 DICOM option setup requirements(continued) Figure 2-3. Worksheet for DICOM Network Information 2-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, and external peripherals for electrical safety are also included in this procedure. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
3-5 ‘Completing the setup’ on page 3-14 • 3-6 ‘System Configuration’ on page 3-21 • 3-7-3 ‘Peripherals Installation Instructions’ on page 3-29 • 3-9 ‘Paperwork after setup’ on page 3-47 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
3-2 Setup reminders 3-2-1 Average setup time • Unpacking the LOGIQ V3/V5/V5 Expert: 30 minutes • Set up LOGIQ V3/V5/V5 Expert options: 30 minutes • DICOM Network Configuration: 30 minutes or more, depending on the configuration • Install Insite: 0.5 hour The LOGIQ V3/V5/V5 Expert installation and functional checkout will take approximately 1 hour.
Page 80
The following table describes guidelines for reaching operational temperatures from storage or transport temperatures. Table 3-1: System Acclimation Time Chart Degree C Degree F hours LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 81
3-2-2 Setup warnings(continued) Operator Manual(s) CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ V3/V5/V5 Expert and kept near the Ultrasound system for quick reference. Acoustic Output Hazard CAUTION Although the ultrasound energy transmitted from the LOGIQ V3/V5/V5 Expert probe is within AIUM/NEMA standards, avoid unnecessary exposure.
System Setup 3-3 Receiving and unpacking the equipment 3-3-1 Purpose of this section This section describes how to receive and unpack LOGIQ V3/ V5/V5 Expert. 3-3-2 Contents in this section • 3-3-3 ‘Warnings for receiving and unpacking’ on page 3-6 •...
Receiving and unpacking the equipment 3-3-4 LOGIQ V3/V5/V5 Expert transportation box label The LOGIQ V3/V5/V5 Expert transportation box label is located at the front of the transportation box. Figure 3-1. LOGIQ V3/V5/V5 Expert transportation box label 1 Figure 3-2. LOGIQ V3/V5/V5 Expert transportation box label 2 LOGIQ V3/V5/V5 Expert –...
Chapter 1. Please carefully unpack the system, and do not dispose the CAUTION package of LOGIQ V3/V5/V5 Expert, so that it can be reused for service. Do not lift the unit by the Keyboard. Equipment damage may CAUTION result.
Page 85
Remove the top cover. Remove the three plastic locks. Note: Rotate the inside plastic lock counterclockwise to remove it and then remove the outside lock. Remove the outside shipping box. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 86
Ignore this step if there is no dust bag. Remove the clear plastic (wrapped around the LOGIQ V3/V5/V5 Expert) from the unit. Cut the packing straps around the four wheels. 3-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 87
3-3-5-1 Moving into Position Please refer to Basic User Manual on how to move the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-11 5726264-100 English Rev.8...
System Setup 3-3-6 Packing the Equipment Please pack LOGIQ V3/V5/V5 Expert in the reverse order of unpacking. 3-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
3-4-3 EMI protection The LOGIQ V3/V5/V5 Expert has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference.
Verify that the site meets the requirements listed in Chapter 2. (See: 2-3 ‘Facility needs’ on page 2-12.) • Verify that the specifications below don’t conflict with any on-site conditions. 3-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
3-5-4-1 Verification of the system’s voltage setting Verify that the mains voltage specified for the LOGIQ V3/V5/V5 Expert is available on-site. Refer to the latest revision of the Basic User Manual for a full list of product labels prior to serving the system.
3-5-5 Connecting Cables For software Option Installation, the operator must login as WARNING Administrator. Table 3-5: List of Connecting Cables. Name Part No. Figure NOTE For USB USB Cable Max.300VA Printer LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-17 5726264-100 English Rev.8...
System Setup 3-5-6 Peripherals/Accessories Connector Panel LOGIQ V5/LOGIQ V3 peripherals and accessories can be properly connected using the side connector panel. Figure 3-3. Rear Connector Panel NOTE: The AC printer can be connected to the isolated Printer USB port on the rear panel only.
Page 95
Pin Assignment of S-Video Pin No. Output Signal Pin No. Output Signal Table 3-10: Pin Assignment of Composite Video Out Output Pin No. Output Signal Pin No. Signal Composite Out LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-19 5726264-100 English Rev.8...
GE. 3-5-7-1 Connect Ethernet Connect the network cable to the Ethernet connector on the I/O Rear Panel. The connector is located on the rear side of LOGIQ V3/V5/V5 Expert. 3-5-7-2 Connect USB Flash Drive NOTE: USB Flash Drive approved for LOGIQ V3/V5/V5 Expert are verified for EMC performance according to EN55011 class B.
System Configuration 3-6 System Configuration 3-6-1 Purpose of this section This section describes how to configure the LOGIQ V3/V5/V5 Expert. 3-6-2 LOGIQ V3/V5/V5 Expert configuration For complete instructions, refer to the latest revision of the LOGIQ V3/V5/V5 Expert Basic User Manual, Chapter 16.
Page 98
Figure 3-5. Replace Key Cap There is a key reserved for CWD function, replace the blank key cap for CWD option if CWD option installed. Figure 3-6. CWD Key Cap 3-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 99
Configurable Key and assign the user defined key. Figure 3-7. Assign User Configurable Key 3-6-2-2 Multi-language Keyboard Film installation 1. Take out the keyboard film. Figure 3-8. Keyboard Film LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-23 5726264-100 English Rev.8...
Page 100
Figure 3-9. Peel Off the Paster 3. Make sure the keyboard is clean. Align the keyboard film on the keyboard and attach it. Figure 3-10. Attach The Film 3-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 101
System Configuration 3-6-2-2 Multi-language Keyboard Film installation(continued) 4. Peel off the other part of the paster and align the whole film on the keyboard. Figure 3-11. Peel Off the Paster LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-25 5726264-100 English Rev.8...
System Setup 3-7 Peripherals Installation 3-7-1 Overview This section describes how to install and configure the peripherals validated for the LOGIQ V3/V5/V5 Expert. About the operation check-out of peripherals, 4-3-34 ‘Peripheral checks’ on page 4-43 for more information. Table 3-11:...
Materials furnished with USB Stick Item Description Quantity Note Keeber USB Stick USB cable • USB HDD Table 3-15: Materials furnished with the USB HDD Item Description Quantity Note USB HDD USB Cable LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-27 5726264-100 English Rev.8...
Page 104
Materials furnished with the Footswitch Item Description Quantity Note 1 Wireless Adaptor • DVD RW Table 3-18: Materials furnished with the Footswitch Item Description Quantity Note 1 DVD RW 1 Y cable 3-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
2. Ensure no physical damage. 3-7-3-2 Installation Procedure NOTE: The printer driver is customized for the LOGIQ V3/V5/V5 Expert at the factory, you do not need to change the settings. 1. UP-D711MD Printer can be properly connected using USB Ports.
Page 106
Figure 3-13. Power and USB Ports for DC printer a. DC Power Port b. USB POrt 2. Press Utility on the Control Panel. 3. Select Connectivity-> Service, add Standard Print. Figure 3-14. Connectivity->Service 3-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 107
Print2...) from the Physical Print Buttons selection. Select the printer from MyComputer column and press >> to move it to the Printflow View column. PressSave. Figure 3-16. Select Button LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-31 5726264-100 English Rev.8...
Page 108
Click Print -> Printing Preferences at the menu of Properties Window. Figure 3-18. Properties • Select Paper Size. Press Apply. Press OK. Figure 3-19. Printing Preferences • Press Save, then Exit. 3-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 109
2. Connect the USB Cable on the Printer. 3. Connect the power cord with the AC output in the wall outlet, then turn on the printer. 4. Connect USB cable to LOGIQ V3/V5/V5 Expert isolated USB port. Figure 3-20. Color Printer connection 5.
Page 110
2. Connect the USB Cable on the Printer. 3. Connect the power cord with the AC output in the wall outlet, then turn on the printer. 4. Connect USB cable to LOGIQ V3/V5/V5 Expert via the USB port under the control panel. Figure 3-21. UP-D898 Connection 5.
Page 111
Figure 3-22. Screw Screws 2. Mount the printer shelf on the system and make sure the 2 screws are in the fixed position. Figure 3-23. Mount the Printer Shelf LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-35 5726264-100 English Rev.8...
Page 112
System Setup 3-7-3-5-4 Installation Procedure(continued) 3. Fix 2 screws on both left and right sides to fix the printer shelf. Figure 3-24. Fix the Printer Shelf 3-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 113
1. Unpack the Footswitch. 2. Ensure no physical damage. 3-7-3-6-4 Installation Procedure 1. Connect the Footswitch to the USB port on the LOGIQ V3/ V5/V5 Expert system. Figure 3-25. Connect Footswitch to the system LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-37 5726264-100 English Rev.8...
Page 114
Next Heartcycle, Prev Heartcycle, Print1, Print2, Update, Next Step (Scan Assistant), Previous Step (Scan Assistant), Scan Assistant pause/Resume. Enter Utility -> Application-> Settings to configure the Footswitch functions. Figure 3-26. Configuring Footswitch Functions 3-38 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 115
1. Unpack the USB HDD. 2. Ensure no physical damage. 3-7-3-7-4 Installation Procedure 1. Connect the USB HDD to the USB port on the LOGIQ V3/ V5/V5 Expert system. Figure 3-27. Connect HDD to the system LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-39 5726264-100 English Rev.8...
Page 116
1. Unpack the ECG. 2. Ensure no physical damage. 3-7-3-8-4 Installation Procedure 1. Connect the ECG to the USB port on the LOGIQ V3/V5/V5 Expert system and the cables can be hang in the cable hook. Figure 3-28. Connect ECG to the system 3-40 LOGIQ V3/V5/V5 Expert –...
Page 117
1. Take out the DVDRW and USB Y cable and connect the USB Y cable to the DVDRW. Figure 3-29. Connect Y cable 2. Insert the DVDRW in the printer shelf right below the printer. Figure 3-30. Insert the DVDRW LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-41 5726264-100 English Rev.8...
Page 118
System Setup 3-7-3-9-4 Installation Procedure(continued) 3. Connect the two connectors on the Y cable to the USB ports on the system. Figure 3-31. DVD RW Connector 3-42 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 119
1. Unpack the wireless adaptor 2. Ensure no physical damage. 3-7-3-10-4 Installation Procedure 1. Connect the wireless adaptor to the USB port on the LOGIQ V3/V5/V5 Expert system. Figure 3-32. Wireless Adaptor Connector LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-43 5726264-100 English Rev.8...
3-8-1 Software Option Installation Procedure NOTE: Not all features described in this section may be available or cleared for sale in all markets. Please contact with your local GE Ultrasound representative to get the latest information. 1. Power on the system.
Page 121
5. After the system is powered on, check the option status. Figure 3-35. Check Option Status The option status explanation: • Permanent: This option is enabled in the system. • Disabled: This option is disabled in the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-45 5726264-100 English Rev.8...
Page 122
Figure 3-36. Replace Caps 7. This completes the option installation of LOGIQ V3/V5/V5 Expert. NOTE: Installing corresponding hardware is a prerequisite of loading software option successfully.
During and after setup, the documentation (i.e. CDs with documentation, User Manuals, Installation Manuals, etc.) for the LOGIQ V3/V5/V5 Expert and the peripherals must be kept as part of the original Ultrasound system documentation. This ensures that all relevant safety and user information is available during the operation and service of the complete Ultrasound system.
Page 124
System Setup 3-48 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 4 General Procedures and Functional Checks This chapter provides procedures for quickly checking major functions of the LOGIQ V3/V5/V5 Expert and diagnostics instructions using the built-in service software. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
4-3 ‘Functional checks’ on page 4-21 4-1-3 Special Equipment required To perform these tests, you'll need any of the sector, linear, or convex probes. (Normally you should check all the probes used on the system). LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
The covers are required for safe operation, good Ultrasound system performance and cooling purposes. Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1.
Do not cycle the Circuit Breaker ON-OFF-ON in less than five (5) seconds. When turning OFF the Circuit Breaker, the Ultrasound system should de-energize completely before turning the circuit breaker ON. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 130
General Procedures and Functional Checks 4-2-2-2 Connect AC (mains) Power to LOGIQ V3/V5/V5 Expert Connecting AC Power to the LOGIQ V3/V5/V5 Expert ultrasound unit, involves preliminary checks of the power cord, voltage level and compliance with electrical safety requirements. 1. Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
Page 131
General procedures 4-2-2-3 Switch ON the AC Power to LOGIQ V3/V5/V5 Expert 1. Switch ON the Mains Power Circuit Breaker at the rear of the unit. You should hear a “click” from the relays in the AC Power and the unit is ready to boot.
It is recommended to perform a full shutdown at least once a week. If the Shutdown button is dimmed, use the key-combination <Ctrl+Alt+Delete> to shut down the unit. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 133
Relocating the scanner. DO NOT unplug and/or transport the unit until after the power CAUTION off sequence has been completed. Failure to do so may result in corrupted patient files. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
• Enter Utility -> System -> General -> Date/Time. • Reset the system date and time. • Select Apply and then select OK. • Select Save. 4-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Refer to the latest revision of the User Manual to perform the following tasks: • Using Removable Media • Labeling Removable Media • Formatting Removable Media • Verifying Removable Media LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-11 5726264-100 English Rev.8...
Page 136
Presets should NOT be saved on the same USB memory device (or DVD-R disk) as images. The Archive Menu lists the images but does NOT list the presets stored on a USB memory device (or DVD-R disk). 4-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 137
4-6, select OK to continue. Figure 4-6. Format Warning Pop-up Window 5. If desired, verify that the format was successful by returning to Utility-> Connectivity->Removable Media and selecting Verify. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-13 5726264-100 English Rev.8...
Page 138
1. Insert a formatted media into the drive. 2. Enter enter Utility-> System-> Backup/Restore. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 3. In the Backup list, select Patient Archive, Report Archive, User Defined Configuration and Service.
Page 139
4. In the Media field, select the Backup/Restore CD/DVD (or USB memory device). 5. Select Restore. The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore screen. Figure 4-8. Restore message LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-15 5726264-100 English Rev.8...
Page 140
(backed up earlier) maintains pointers to the location of the images on the archive media. Figure 4-9. Format Media Screen 5. Enter Utility-> System-> Backup/Restore. 6. Select “Move File Older Than in Days”. 4-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 141
(even if the exam was previously up), check this option. If you uncheck this option, the system only backs up exams which have not yet been backed up. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-17 5726264-100 English Rev.8...
Page 142
EZBack/EZMove. Figure 4-13. EZBackup Wizard 2 NOTE: This message “Please insert a blank media...” appears if you press Next without inserting the backup media. Insert the media and continue. 4-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 143
If you need to insert the next media, a message appears providing you with the media label. 11. When the backup is complete, the completion wizard page appears. Press Finish. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-19 5726264-100 English Rev.8...
Page 144
General Procedures and Functional Checks 4-2-7-4 Archiving Images(continued) Figure 4-16. EZBackup Completion Window All databases, presets and images should now be saved to removable media. 4-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Functional checks 4-3 Functional checks 4-3-1 Overview In this section, the functional checks for LOGIQ V3/V5/V5 Expert are described. Functional checks are used to verify that the product works as intended. Functional checks may also be used during troubleshooting. 4-3-2 Contents in this Section 4-3-3 Preparation Turn on power to LOGIQ V3/V5/V5 Expert.
Functional checks 4-3-6 LOGIQ V5/LOGIQ V3 SoftMenu Key Tour • Rotary key: Rotate to adjust the menu. • Page Up/Down key: To turn the menu page up and down. Figure 4-18. SoftMenu Key Tour LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-23 5726264-100 English Rev.8...
13. Current date and time, Caps Lock: (lit when on), 27. Body Pattern. network connection indicator (PC=connected, PC with X=not connected), system messages display, InSite status, InSite controls. Image Preview 4-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Functional checks 4-3-8 Performance Tests 4-3-8-1 Test Phantoms The use of test phantoms is only recommended if required by your facility's (customer’s) QA program. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-25 5726264-100 English Rev.8...
Density (move to higher line density) to increase resolution in image. • Use Frequency (move to lower frequency) to increase penetration. • Use the Rejection control to reduce noise in the image. 4-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Figure 4-20. Check “Enable IQC for Service” 2. Press Utility -> Scan Coach -> Imaging Preset Manager. Select IQC from the left column, and select move it to the Imaging Preset Selection. Then Select Exit. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-27 5726264-100 English Rev.8...
Page 152
4-3-13 Image Quality Check (IQC) preset for service(continued) Figure 4-21. Image Quality Check 3. Press Probe/Preset on the control panel. Select SVC, then select Figure 4-22. Image Quality Check 4-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Adjust Gain and TGC controls to adjust the range to be imaged. Adjust Compression and Edge Enhance to further optimize the display. Adjust Rejection to reduce noise while taking care not to eliminate significant low-level diagnostic information. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-29 5726264-100 English Rev.8...
2. Use the trackball to position the ROI frame over the area to be examined. 3. Press Scan Area. The instruction Size should be highlighted in the trackball status bar. Use the trackball to adjust the dimension of the ROI. 4-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Spatial Filter smooths the image by averaging collected data along the same horizontal line. An increase of the spatial filter will reduce noise, but this will also reduce the lateral resolution. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-31 5726264-100 English Rev.8...
2. Use the trackball to position the ROI frame over the area to be examined. 3. Press Scan Area. The instruction Size should be highlighted in the trackball status bar. Use the trackball to adjust the dimension of the ROI. 4-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 157
Use Threshold to emphasize either the color flow overlay, or the underlying grey scale tissue detail. Use Baseline to emphasize flow either toward or away from the probe. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-33 5726264-100 English Rev.8...
If needed, adjust the Display’s Brightness and Contrast setting. Press PW to start Pulsed Wave Doppler (PW). Use the trackball to select the Area of Interest (Sample Volume) in PW. 4-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Use Invert to reverse the vertical component of the spectral Doppler area of the display. Use Angle Correct to steer the ultrasound beam to the blood flow to be measured. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-35 5726264-100 English Rev.8...
If the trackball control pointer is selected, press trackball to be able to select between Position and Size controls. Use the trackball to adjust the dimension of the ROI. 4-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
However, from this window the beam cannot be aligned to the muscle for all the parts of the ventricle. NOTE: PW will be optimized for Tissue Velocities when activated from inside TVI. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-37 5726264-100 English Rev.8...
To rotate through and activate previously fixed calipers, rotate Cursor Select button. NOTE: After you complete the measurement, to erase all data that has been measured to this point, but not data entered onto worksheets, press Clear. 4-38 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Launch the application. To change application, The selected application starts. press Probe/Preset key on the Control Panel. Verify there’s no EMI/RFI or artifacts specific to the No EMI/RFI or artifacts. probe. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-39 5726264-100 English Rev.8...
Page 164
5. Be sure that the probe head is clean before placing the probe in its storage box. Take the following precautions with the probe cables: Do not WARNING bend, be sure to keep probe cables free from the wheels. 4-40 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Fasten the three ECG Pads to your body and When the connection is completed, a typical connect the three leads to respective ECG clean ECG signal is displayed. Pad. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-41 5726264-100 English Rev.8...
If all the previous tests have been passed successfully, the Back End Processor is most likely OK. If the system seems to be operating erratically, please refer ‘Diagnostics/Troubleshooting’ on page 7-1. 4-42 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Functional checks 4-3-33 Operator Panel Test • The Operator Panel is tested when the LOGIQ V3/V5/V5 Expert is powered up as part of the start-up scripts, run at every start-up. For more info, please refer to ‘Diagnostics/Troubleshooting’ on page 7-1.
Page 168
4-3-36-2 Casters (Wheels) and Brakes Checks Examine the wheels frequently for defects to avoid breaking or jamming. Table 4-4: Wheel Characteristics Wheel Characteristics Front and Rear Swivel and Brake 4-44 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Verify that Location (Hospital Name) is correct. Verify Language settings are correct. Verify assignment of Print Keys. Verify all of the customer’s options are set up Demo Option strings turn on correctly. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-45 5726264-100 English Rev.8...
General Procedures and Functional Checks 4-5 Site Log Table 4-6: Site Log DATE SRVICE PERSON PROBLEM COMMENTS 4-46 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
2. Press defined key to access My Trainer. There are four sections in My Trainer. The four sections are displayed on the left side of the My Trainer interface. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-47 5726264-100 English Rev.8...
Page 172
3. Peripheral Connection: Back panel ports, Network Connectivity, Connect to a printer, DICOM setting, Back-up/ restore the patient information, software version 4. Maintenance: Trackball, Contact GE service, Insite, Log export, Air Filter Replacement 4-48 LOGIQ V3/V5/V5 Expert – Basic Service Manual...
Page 173
6. The left number shows the current page of the subsection. The right number shows the total pages of the subsection. 7. Illustration with graphic. Table 4-7: Page turning Select the arrow to go to the previ- ous page LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-49 5726264-100 English Rev.8...
Page 174
Select the arrow to move to the next page. Select the arrow to go to the home page Select the arrow to go to the end page. 9. Text instruction about how to operate. 4-50 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 5 Components and Functions (Theory) This chapter explains LOGIQ V3/V5/V5 Expert’s system concepts, component arrangement, and subsystem functions. It also describes the power distribution and the Common Service Desktop interface. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Components and Functions (Theory) Overview 5-1-1 Contents in this chapter • 5-1 ‘Overview’ on page 5-2 • 5-2 ‘Block Diagram and Theory’ on page 5-3 • 5-3 ‘Common Service Platform’ on page 5-5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Block Diagram and Theory Block Diagram and Theory 5-2-1 Block Diagram 5-2-1-1 System Diagram Figure 5-1. LOGIQ V3/V5/V5 Expert System Block Diagram LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
ACDC and provides all voltage needed to WMST board and SMIO board. SFAN PWA is a system FAN interface board, it provides power and control to system FAN (quantity is 4 internal the box). LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 179
The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. The Common Service Platform will increase service productivity and reduce training and service costs. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 180
Components and Functions (Theory) LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 6 Service Adjustments This chapter describes how to test and make adjustments to the LOGIQ V3/V5/V5 Expert. You can use these to test the system for errors. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Service Adjustments 6-1 Overview 6-1-1 Contents in this chapter • 6-1 ‘Overview’ on page 6-2 • 6-2 ‘LCD Monitor Adjustments’ on page 6-3 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
In rare cases, LCD monitor might have defect pixels appearing. This section describes how to discriminate the defect pixels with the number of defect pixels allowed for the LOGIQ V3/V5/V5 Expert to support LCD monitor replacement decision. Monitor Adjustments...
Page 184
To adjust the Contrast: 1. Press Middle key twice, the contrast indicator displays; Figure 6-4. Contrast setting menu 2. Press Left key or to decrease contrast or press right key to increase contrast. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 185
Keep pressing the middle key for 3 to 5 seconds, the Brightness and Contrast are reset. Figure 6-5. Reset setting menu Press the Right key once to activate main menu to have more settings. Figure 6-6. Main Menu LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 186
• To adjust the color temperature: press Standard, Brown, or Blue to choose the color temperature. The chosen color temperature displays in the Active area. Figure 6-8. Active Color Temperature LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 187
Move the adjustment button to the left to decrease the contrast or move the adjustment button to the right to increase the contrast. Figure 6-10. Contrast LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 188
3. Press Reset to return to the default setting; Press Exit to exit the setting menu. Figure 6-12. Reset NOTE: Press Reset button and the three color temperature settings will return to the default settings. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 7 Diagnostics/Troubleshooting This chapter describes LOGIQ V3/V5/V5 Expert Series how to setup and run the tools and software that help maintain image quality and system operation. Very basic host, system and board levels are run whenever power is applied. Some Service Tools may be run at the application level.
7-2 ‘Gathering Trouble Data’ on page 7-3 • 7-3 ‘Screen Capture’ on page 7-5 • 7-4 ‘Global Service User Interface (GSUI)’ on page 7-12 • 7-5 ‘Network and Insite II Configuration’ on page 7-29 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Product Name = LOGIQ V3/V5/V5 Expert From the Utility>System>General>About screen: Applications Software • Software Version • Software Part Number System Image Software • Image Date • Image Part Number Additional Information LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Options enables creation of a log file based on a selected bookmark or for a user configurable time frame. Different type of information can be selected to be part of the log file. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Store Key. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
1. Navigate to and display the image/screen to be captured. 2. Press Store. This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display. Figure 7-2. Select Image to Capture LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 195
USB Drive or CD/DVD. Figure 7-4. Save Dialog Box NOTE: Click Save -> Transfer To CD/DVD, if the archive location on the DVD/CD; Click Save, if the archive location is USB. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 196
Diagnostics/Troubleshooting 7-3-1 Capturing a Screen(continued) NOTE: It is better to save the image in Jpeg format. Image of this format can be easily reviewed in the computer. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
3. Press Ctrl+Alt+Delete, the Windows Security dialog is displayed. Then select Task Manager. Figure 7-5. Windows Security NOTE: It is also available to Exit to Windows and save the image on the system or removal media. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 198
7-3-2 Capturing a Screen with Service Dongle(continued) 4. In Windows Task Manager, select File-> New Task (Run...). Figure 7-6. Windows Task Manager 5. Type in the program name: mspaint. Figure 7-7. Create New Task 7-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 199
7-3-2 Capturing a Screen with Service Dongle(continued) 6. The paint screen is displayed, press Ctrl+V to paste the captured screen. And then save the image in the appropriate destination. Figure 7-8. Save Image LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-11 5726264-100 English Rev.8...
7-4-1 Enter global service user interface 1. Select Utility -> Service, then the service browser will start. Figure 7-9. 2. Select GE Service in option of User Level of GEMS Service Home page and input correct password, press OK button. 7-12 LOGIQ V3/V5/V5 Expert –...
Page 201
7-4-1 Enter global service user interface(continued) Figure 7-10. Service Login Page 3. Choose Diagnostic in Global Service Interface to active diagnostic functions, choose options to activate various functions correspondingly. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-13 5726264-100 English Rev.8...
Page 202
Diagnostics/Troubleshooting 7-4-1 Enter global service user interface(continued) Figure 7-11. Active Diagnostic Function 7-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
• Fail = Red • Pass = Green • Neither pass nor fail = Set back to Gray (for example, final code status is Aborted). LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-15 5726264-100 English Rev.8...
1. In the server side, select the system which the OLE would remotely connect to, and select Connect to device. Figure 7-12. Connect to device 2. Select CSD to Connect. Then the OLE is remotely connected to the system. 7-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 205
Global Service User Interface (GSUI) 7-4-3-1-27-4-3-1-1 Remote access(continued) Figure 7-13. Remotely Connect 3. Enter Utility-> Common Utilities-> Disruptive Mode, then select Yes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-17 5726264-100 English Rev.8...
Page 206
Virtual Console Observation and Start it. Then connect using VNC. Figure 7-15. Virtual Console Observation 5. Select the VNC to Connect. Then the OLE is remotely connected to the system. 7-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 207
Global Service User Interface (GSUI) 7-4-3-1-47-4-3-1-1 Remote access(continued) Figure 7-16. Virtual Console Observation Connect 7-4-3-2 Service Diagnostics BF, COMEXP and GFS modules are on the LOGIQ V3/V5/V5 Expert WMST PWA. CPS is WPWR/DCDC PWA. IOS is SMIO PWA. PID is SLRY 4-Probe PWA.
Page 208
Front End Interface FPGA Test reads the version of the GFE FPGA. • GFS Fixed Demo Test: GFS Fixed Demodulator signal path test. • DSP Master Clock Test tests DSP Master Clock. • GFS Swept Demodulator Long Test. 7-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 209
AFE IF Test • Complex Mixer Test • CE Decoder Test 7-4-3-2-7 ECG Tests • ECG Tests 7-4-3-2-8 Doppler Tests • Doppler Audio Test runs the tests for Doppler. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-21 5726264-100 English Rev.8...
Page 210
• Note: This diagnostic is intended to verify keyboard keys are in good working order. It is not intended to verify that keyboards produce desired characters. 7-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 211
Click the appropriate button to fill the screen with the associated test pattern. You can return to this dialog by clicking the mouse button or pressing any key. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-23 5726264-100 English Rev.8...
Page 212
Double-click on the picture of the monitor. If successful, the picture will change. • Sound Test generates sounds for testing the speakers. • USB Ports Test lists USB Devices. 7-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 213
7. Press Probe button on the control panel and select the first probe from the probe indicators. 8. Enter global service user interface as GE service. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-25 5726264-100 English Rev.8...
Page 214
Utility-> Probe Assessment to check the probe. If there is no broken element and the probe is acceptable for diagnostic, the following result window displays. Figure 7-18. Probe Assessment Pass 7-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 215
If there are broken elements and the probe is not acceptable for diagnostic. The assessment is fail and the failed elements are shown in the graphic. Figure 7-19. Probe Assessment Fail Reboot the system after the probe assessment. CAUTION LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-27 5726264-100 English Rev.8...
Page 216
Max 2 weak elements on the scan line: -3dB<elements<-2dB • Any one of the elements in one scan line: element<-3dB 7-4-3-3 Restart the system after diagnostics Always shutdown the system and reboot after a diagnostics session. 7-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
1. Connect system with network. 2. Enter Utility-> Connectivity-> TCP/IP, in IP settings window, check Enable DHCP, and select the proper network speed in Network Speed. Figure 7-20. Enable DHCP LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-29 5726264-100 English Rev.8...
Page 218
Figure 7-22. System Restart inquiry dialog 4. After the system restarts, the network icon at the left bottom of screen shows that the network is available. 7-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 219
Network and Insite II Configuration 7-5-1 Network Configuration(continued) Figure 7-23. Network icon LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-31 5726264-100 English Rev.8...
Page 220
2. Enter Utility-> Connectivity-> TCP/IP, press Configuration. Figure 7-24. Network Settings 3. Double click Wireless network. you chosen, or clink the Connect at the left bottom of screen. Figure 7-25. Wireless Network Conection 7-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 221
4. Input Network Key and Confirm Network Key, then click Connect. Figure 7-26. Connection Status 5. The network icon at the left bottom of screen shows that the wireless network is available. Figure 7-27. Wireless-LAN Connection LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-33 5726264-100 English Rev.8...
Page 222
Diagnostics/Troubleshooting 7-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 8 Replacement Procedures This chapter describes how to remove and install, or replace, modules and subsystems in the LOGIQ V3/V5/ V5 Expert. It also includes instructions for installing and re-installing the software. LOGIQ V3/V5/V5 Expert – Basic Service Manual...
8-1 Overview 8-1-1 Contents in this chapter • 8-1 ‘Overview’ on page 8-2 • 8-2 ‘Warnings and important information’ on page 8-3 • 8-3 ‘Loading the software’ on page 8-5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Because of the limited access to cabinets and equipment in the WARNING field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
Equipment being returned must be clean and free of blood and other infectious substances. GE policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment.
Loading the software 8-3 Loading the software 8-3-1 Purpose of this section This section describes how to reinstall software on LOGIQ V3/ V5/V5 Expert. 8-3-2 Customer provided prerequisite • Formatted and labelled media for Images storage. • Formatted and labelled media for Patient Archive and Presets (User Defined Settings).
Database, as needed after a hard disk replacement, or if all the content on the hard disk has been erased, the images must be moved away from LOGIQ V3/V5/V5 Expert before doing backup of the Patient Database. Depending on the location set-up, either move the images to a remote server or to removable media like DVD or CD discs.
CAUTION It is considered to be best practice to always keep a record on paper of the settings for the LOGIQ V3/V5/V5 Expert. Verify if it is current before you start to load software! Always keep a record of the settings for the LOGIQ V3/V5/V5 Expert on paper.
If the system will not shutdown normally, hold down the Power On/Off Switch until the light turns off. 3. Power on the system. The system will detect the USB memory stick automatically. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 231
(Customer Presets) will be lost. • To select [4], Screen size. This option is to adjust the size of screen display. NOTE: All the patient data cannot be recovered after wipe tool. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 232
5. System USB memory stick will be loading. Wait for the software installation to complete. (Typical installation time: 10-15 minutes). Status bar on the screen indicates progress. Do not interrupt the software loading at any time. WARNING 8-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 233
If you do not remove the USB memory stick, the software system loading process repeats when the system boots up.Do not remove the USB stick after the software reload if you want to adjust the screen size. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-11 5726264-100 English Rev.8...
Page 234
Power On/Off Switch. In System-Exit window, select Shutdown to shutdown the system. NOTE: If the system will not shutdown normally, hold down the Power On/Off Switch until the light turns off. 8-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 235
Figure 8-4. Upgrade message 5. The system indicates all data will be erased, select Yes to continue. Figure 8-5. Confirm Dialog NOTE: All the patient data cannot be recovered after wipe tool. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-13 5726264-100 English Rev.8...
Page 236
8-3-7-2 Wipe Tool(continued) 6. The process may take twenty minutes or more. Figure 8-6. Wipe Tool Process 7. Select OK when the process is complete, and the system will shutdown automatically. 8-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 237
MUST be the same as the ones before installation. Please record the serial number and option key before installation. 1. Insert the USB memory stick labeled “System & Application Software” to the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-15 5726264-100 English Rev.8...
Page 238
If you select [3], ALL existing software and data will be erased. CAUTION If backup has not been performed, all data like Patient Database, System Configuration and User Configurations (Customer Presets) will be lost. Figure 8-8. System Software load instruction 8-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 239
GEWuXiChina and it is case sensitive. Figure 8-9. System upgrade complete 6. Click Next to continue this operation. Figure 8-10. System upgrade complete 7. Select Yes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-17 5726264-100 English Rev.8...
Page 240
8. System USB memory stick will be loading. Wait for the software installation to complete. NOTE: Do not interrupt the software loading at any time. Figure 8-12. System upgrade complete 8-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 241
NOTE: The serial number and option key MUST be the same as the ones before installation. Please record the serial number and option key before installation. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-19 5726264-100 English Rev.8...
Page 242
2. Power on the system. The system will detect the USB memory stick automatically. 3. Select choice [4] -> Next. Figure 8-14. Screen size 4. Enter the ultrasound maintenance access password.The password is GEWuXiChina and it is case sensitive. 8-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 243
NOTE: If you do not remove the USB memory stick, the software system loading process repeats when the system boots up. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-21 5726264-100 English Rev.8...
1. If the serial number on VPD and HDD does not match, the following dialog will display when the system is booting up.The password is the same as that of service dongle. Figure 8-18. Input Password 8-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 245
The serial number is case sensitivity. Figure 8-20. Input Serial Number NOTE: The serial number can be found on the rating plate on the rear side of the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-23 5726264-100 English Rev.8...
Page 246
8-3-8 Rewrite the Serial Number(continued) 3. The serial number is reset, reboot the system. Figure 8-21. Reboot the System NOTE: Make sure the system is shut down after the serial number is reset. 8-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 247
Loading the software 8-3-8-1 Software Version check out 8-3-8-1-1 Functional Check-out 1. Power on LOGIQ V3/V5/V5 Expert system and wait until system booting to scanning screen. NOTE: If selected to load C Disk only when loading the system software, the system will display a screen to restore Computer name before entering the scanning screen.
Page 248
2. Press Utility on the Control Pane, then select System -> General. 3. In the Location portion, select the right region in the field of Preset Region. Figure 8-23. Preset Region 4. Reboot the system. 8-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 249
The status “disabled” means the option keys are not activated and not working. Check if the option is installed and if the serial number and option key are correct. Figure 8-24. Software Option LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-27 5726264-100 English Rev.8...
Page 250
Plug in the probe. In scanning mode, the probe information is displayed on the right top location of the screen. About the probe specification for intended use on LOGIQ V3/V5/V5 Expert. Plug in at least one of each type of the probes and check if each of the probes is recognized and the probe information is displayed correctly.
Page 251
The instruction about installing DICOM devices is not incorporated in this manual. To access the instruction about installing DICOM devices please refer to another manual Basic User Manual. Please use the latest revision of this document. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-29 5726264-100 English Rev.8...
Page 252
Replacement Procedures 8-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Chapter 9 Renewal Parts This chapter lists the renewal parts available for the LOGIQ V3/V5/V5 Expert. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Renewal Parts 9-1 Overview 9-1-1 Contents in this chapter • List of Abbreviations • Renewal Parts Lists LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
ASSEMBLY FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufacturer (lead time involved) Keyboard Liquid Crystal Display Brightness and Volume WMST Master Board LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
9-3-1 Operator Console Assy The following figure illustrates what is the front side (1), left side (2), back/rear side (3), right side (4) of the system. Figure 9-1. Operator Console Assy LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
ECG Detachable cable AHA Type USA 5146739 ECG detachable cable IEC type EURO and ASIA 5146055 USB Cable for ECG 5129487 ECG module from Norav Isral (not applies to Muslim country and China) LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 258
RAB2-6-RS Biopsy Kit E8385MJ E8C-RS Biopsy Kit USB HDD 5434317-4 1TB mobile USB HDD DVD RW 5729697 LV R2 DVD RW Kit with SW 5653589 LITEON eBAU108 DVD Writer Kit LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 259
Ayan R2 R2.0.7 System and Application Soft- ware (for Service) Others H48702AS VGA to Svideo adapter H48822AC LF/LV AN Key film H48612AS LV R2 CHN KBD Film H48832AC Wireless Adaptor H48552LJ Printer Power Cable for Brazil LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
5177154-2 AC Power Cord Switzerland 5177187-3 AC Power Cord Australia 5177146-2 AC Power Cord USA 5400868-2 AC Power Cord Brazil 5176753-2 AC Power Cord Israel 5177126-2 AC Power Cord Japan LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
9-3-5 Manuals Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1000 5729745 -100 LOGIQ V3/V5/V5 Expert Quick Card for USA 1001 5726265-200 LOGIQ V3/V5/V5 Expert Manual CD 1002 5719190-100 LOGIQ V3/V5/V5 Expert Advanced Reference Manual 1003 5726264-100...
Page 262
Renewal Parts Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1028 5719188-167 LOGIQ V3/V5/V5 Expert User Guide Romanian version 1029 5719188-168 LOGIQ V3/V5/V5 Expert User Guide Croatian version 1030 5719188-174 LOGIQ V3/V5/V5 Expert User Guide Lithuanian version 1031...
Page 263
Renewal Parts Lists Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1057 5726260-155 LOGIQ V3/V5/V5 Expert Release Notes Czech version 1058 5726260-159 LOGIQ V3/V5/V5 Expert Release Notes Turkish version 1059 5726260-160 LOGIQ V3/V5/V5 Expert Release Notes Danish version 1060...
Page 264
Renewal Parts 9-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Ultrasound system and peripherals. These procedures are intended to maintain the quality of the Ultrasound system’s performance. Read this chapter completely and familiarize yourself with the procedures before performing a task. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-1 5726264-100 English Rev.8...
Do not pull out or insert circuit boards while power is on. CAUTION Do not operate this Ultrasound system unless all board covers CAUTION and frame panels are securely in place. System performance and cooling require this. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-3 5726264-100 English Rev.8...
Care and Maintenance 10-3 Why do maintenance 10-3-1 Periodic maintenance inspections It has been determined by engineering that your LOGIQ V3/V5/ V5 Expert does not have any high wear components that fail with use, therefore no Periodic Maintenance inspections are mandatory.
Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Contact GE for coverage and/or price for service.
V5/V5 Expert should be serviced and outlines items requiring special attention. NOTE: It is the customer’s responsibility to ensure the LOGIQ V3/V5/V5 Expert care and maintenance is performed as scheduled in order to retain its high level of safety, dependability and performance.
Page 271
See Notes also after corrective maintenance NOTE: The maintenance may require specialized equipment to complete. NOTE: The periodic maintenances are not mandatory. The table above is for reference only. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-7 5726264-100 English Rev.8...
For a list of required tools for servicing the LOGIQ V3/V5/V5 Expert, refer to chapter 8. 10-5-1 Standard GE tool kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required. Table 10-2:...
Page 273
Weller - Soldering Iron Tip KH175 Hemostat 5 inch curved 9-WEWDP12 Weller - Desoldering Pump 9-Z9480121 Alignment tool (red) 93383 Flashlight Mini-Mag Lite (AAA Bat.) 9-GH408 Tweezers 21576 Brush - Bristle LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-9 5726264-100 English Rev.8...
Screwdriver 5/16 inch × 8 inch #36 Drill 9-41586 Screwdriver, Stubby 5/16 inch #29 Drill 9-GH19512 Countersink 1/2 inch 9-44046 3/8 inch Socket Set 9-44741 12 PC Combination Wrench 10-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Quality Assurance Program tests. B/W Printer Cleaning Sheet See printer user manual for requirements Color Printer Cleaning Sheet See printer user manual for requirements Disposable Gloves LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-11 5726264-100 English Rev.8...
• Verify that InSite is functioning properly. • Ensure two-way remote communications. Diagnostics Optional. Presets Backup all Customer Presets to an appropriate media. Image Archive Back up the Image Archive onto appropriate media. 10-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Verify the distance by graduate rule. Distance Accuracy should be within ±5%. (Name result from result window Result A, result from graduate rule Result B; Distance Accuracy = (Result B-Result A)/Result A) LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-13 5726264-100 English Rev.8...
Page 278
Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no strands may cause a short circuit. Verify Inlet connector retainer is functional. 10-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
• can be adjusted in the up/down positions. Note: LCD Arm movement may vary and is not applicable to all Ultrasound systems. External I/O Check all connectors for damage. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-15 5726264-100 English Rev.8...
1. Review the system error log for any problems. 2. Check the temperature log to see if there are any trends that could cause problems in the future. 3. Check the Configuration Log; update if needed. 10-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
Page 282
User Manual. Doing so could result in irreparable damage to the probe. Follow care instructions that came with the probe. Disinfect a defective probe before you return it. Be sure to tag CAUTION the probe as being disinfected. 10-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Possible risk of infection. Do not handle soiled or contaminated CAUTION probes and other components that have been in patient contact. Follow appropriate cleaning and disinfecting procedures before handling the equipment. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-19 5726264-100 English Rev.8...
Page 284
Check probe leakage current: • Based on your facilities QA program for surface probes. • Based on your facilities QA program for endocavitary probes. • whenever probe damage is suspected. 10-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Electrical safety tests 10-7-2 Leakage current limits Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1. Follow LOCK OUT/TAG OUT procedures.
Page 286
'heart in box' CF symbol. Sink The current resulting from the application of mains Leakage voltage to the applied part. This test is required test for Type CF applied parts. 10-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
(grounded supply) conductor and the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-23 5726264-100 English Rev.8...
The maximum allowable limit for chassis source leakage is shown in Table 10-12 on page 10-21. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-25 5726264-100 English Rev.8...
Page 290
Optional Optional Power Switch Switch Cover Parts Test 3 Test 4 Enter Name of tested peripheral here: NORM OPEN NORM CLOSED OPEN CLOSED NORM OPEN NORM CLOSED OPEN CLOSED 10-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
If no leakage current is detected, check the configuration of the test equipment. 10-7-6-2 Tools For needed tools, see: 10-5 ‘Tools required’ on page 10-8. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-27 5726264-100 English Rev.8...
Page 292
Follow the Safety Analyzer tool instruction to test each transducer for leakage current. The electrical Safety Analyzer tool should be calibrated and compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551. 10-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 293
Typical Data Sheet For Transducer Source Leakage Current Transducer Tested: Tester Power Polarity Tester GROUND or Unit Power Switch NUETRAL Switch Measurement NORM OPEN NORM CLOSED OPEN CLOSED NORM OPEN NORM CLOSED OPEN CLOSED LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-29 5726264-100 English Rev.8...
If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. 10-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Ultrasound system and if situation can not be corrected, submit a Safety Failure Report to document the Ultrasound system problem. Remove Ultrasound system from operation. 10-8-7 ECG Fails Inspect cables for damage or poor connections. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-31 5726264-100 English Rev.8...
* Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other Figure 10-5. Ultrasound Inspection Certificate * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other 10-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 297
Comments Probe 1: Probe 2: Probe 3: Final Check. All system covers are in place. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Figure 10-7. Electrical Safety LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-33 5726264-100 English Rev.8...
Patient lead source leakage (lead to ground) Patient lead source leakage (lead to lead) Patient lead source leakage (isolation) Peripheral 1 leakage current Peripheral 1 ground continuity Peripheral 2 leakage current 10-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 299
Peripheral 2 ground continuity Peripheral 3 leakage current Peripheral 3 ground continuity Table 10-19: Electrical safety tests (probes) log Max value Comments value measured allowed Probe LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-35 5726264-100 English Rev.8...
Page 300
Care and Maintenance 10-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 302
9-3 probes environmental requirements, 2-11 loading software, 8-5 operational and storage temperature for probes, lockout/tagout (LOTO) requirements, 1-25 2-11 LOTO product icons, 1-10 lockout/tagout requirements, 1-25 Index-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
Page 303
10-12 system manufacturer, 1-30 system requirements verification, 3-14 system setup, 3-1 time and manpower requirements site preparations, 2-11 tools required, 10-8 special tools, supplies and equipment, 10-11 LOGIQ V3/V5/V5 Expert – Basic Service Manual Index-3 5726264-100 English Rev.8...
Page 304
Index-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...