Technical Publication
LOGIQ V3/V5/V5 Expert
Basic Service Manual
Direction Number: 5726264-100 English
Rev.8
All Material Copyright © 2016-2017 by General Electric Company Inc.
All Rights Reserved.

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ V3

  • Page 1 Technical Publication LOGIQ V3/V5/V5 Expert Basic Service Manual Direction Number: 5726264-100 English Rev.8 All Material Copyright © 2016-2017 by General Electric Company Inc. All Rights Reserved.
  • Page 2 Product Information This Manual covers the software version of R2.x.x for LOGIQ V3, LOGIQ V5, LOGIQ V5 Expert ultrasound system. Copyright © 2016-2017, General Electric Company. GE Medical Systems, a General Electric Company, going to market as GE Manufacturer: GE Medical Systems (China) Co., Ltd.
  • Page 3 MyWorkshop/ePDM (GE electronic Product Data Management). If you need to know the latest revision, contact your distributor, local GE Sales Representative or in the USA call the GE Ultrasound Clinical Answer Center at 1 800 682 5327 or 1 262 524 5698.
  • Page 4 Important precautions Translation policy LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 6 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 7 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 8 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 9 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 11 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 13 LOGIQ V3/V5/V5 Expert – Basic Service Manual i-11 5726264-100 English Rev.8...
  • Page 14: Damage In Transportation

    GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 15: Omission And Errors

    Wuxi National Hi-Tech Dev. Zone Jiangsu P.R.China 214028 TEL: +86 510 85225888; FAX: +86 510 85226688 GE employees should use TrackWise to report service documentation issues. These issues will then be in the internal problem reporting tool and communicated to the writer.
  • Page 16: Service Safety Considerations

    For a complete review of all safety requirements, refer to Chapter 1 in the Service Manual. i-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 17: Legal Notes

    If you are a GE employee or a customer who has entered into such a license agreement with GE to use this proprietary software, you are authorized to use this Material according to the conditions stated in your license agreement.
  • Page 18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 19: Table Of Contents

    Proprietary to GE - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -...
  • Page 20 LOGIQ V3/V5/V5 Expert transportation box label- - - - - - - - - - - - - - - - - 3-7 Unpacking the LOGIQ V3/V5/V5 Expert - - - - - - - - - - - - - - - - - - - - - - - 3-8...
  • Page 21 Purpose of this section - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 3-21 LOGIQ V3/V5/V5 Expert configuration - - - - - - - - - - - - - - - - - - - - - - - 3-21...
  • Page 22 Peripheral checks - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 4-43 Turn OFF Power to LOGIQ V3/V5/V5 Expert- - - - - - - - - - - - - - - - - - - 4-43...
  • Page 23 Standard GE tool kit - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -...
  • Page 24 Ultrasound Inspection Forms - - - - - - - - - - - - - - - - - - - - - - - - - - - - - 10-32 Electrical Safety Tests Log Index i-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 25: Chapter 1 - Introduction

    This chapter describes important issues related to safely servicing the Ultrasound system. The service provider must read and understand all the information presented here before installing or servicing the units. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 26: Overview

    1-8 ‘Lockout/Tagout (LOTO) requirements’ on page 1-25 • 1-9 ‘Returning probes and repair parts’ on page 1-26 • 1-10 ‘EMC, EMI and ESD’ on page 1-27 • 1-11 ‘Customer assistance’ on page 1-29 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 27: Manual Overview

    Manual Overview 1-2 Manual Overview This manual provides installation and service information for the LOGIQ V3/V5/V5 Expert Ultrasound system. It is divided in ten chapters as shown below. 1-2-1 Contents in this manual The manual is divided into ten chapters.
  • Page 28: Typical Users Of The Proprietary Service Manual

    Licensed Hospital’s Service Providers 1-2-3 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ V5/LOGIQ V3 and also kept near the unit for quick reference. NOTE: Probe information displayed on screen does not necessarily reflect the probes available on your ultrasound system.
  • Page 29: Logiq V3/V5/V5 Expert Models Covered By This Manual

    Manual Overview 1-2-4 LOGIQ V3/V5/V5 Expert models covered by this manual Table 1-2: LOGIQ V3/V5/V5 Expert Model Designations Model Number Description System SW H48612AA LOGIQ V3 R2.x.x H48612AB LOGIQ V5 R2.x.x H48612AC LOGIQ V5 EXPERT R2.x.x NOTE: When not otherwise specified, the contents of this manual applies to all LOGIQ V3/V5/V5 Expert models.
  • Page 30: Important Conventions

    Important conventions, used in this document, are described next. 1-3-1-1 Model designations This manual covers the LOGIQ V3/V5/V5 Expert Ultrasound systems listed in: 1-2-4 ‘LOGIQ V3/V5/V5 Expert models covered by this manual’ on page 1-5. 1-3-1-2 Icons Pictures, or icons, are used wherever they will reinforce the printed message.
  • Page 31 Notes are used to provide important information about an item or a procedure. NOTE: Be sure to read the notes; the information contained in a note can often save you time or effort. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 32: Standard Hazard Icons

    Standard hazard icons ELECTRICAL MECHANICAL RADIATION LASER HEAT PINCH NOTE: Even if a symbol isn’t used on the product or in this manual, it may be included for your reference. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 33 Tag and Lock Out Wear Eye Protection Hand Protection Foot Protection Wear Eye Protection Be sure to read the notes; the information contained in a note can often save you time or effort. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 34: Product Icons

    1-4 Product icons It is important to refer to the current revision of the Ultrasound system’s Basic User Manual for a full list of product labels prior to servicing the system. 1-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 35: Labels Locations

    Basic User Manual for a full list of product labels prior to servicing the system. 1-5-1 Rating Plate Location LOGIQ V3/V5/V5 Expert labels are provided in English. The labels are at the bottom of the system. The label content may be different for different regions. Please refer to the labels on the system for the actual content.
  • Page 36 Introduction 1-5-1 Rating Plate Location(continued) Figure 1-1. LOGIQ V5 Expert Label Location 1-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 37 Labels locations 1-5-1 Rating Plate Location(continued) Figure 1-2. LOGIQ V5 Label Location LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-13 5726264-100 English Rev.8...
  • Page 38 Introduction 1-5-1 Rating Plate Location(continued) Figure 1-3. LOGIQ V3 Label Location 1-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 39 Rating plate Serial Number. NOTE: If the new label is needed during the service activities, please click “Ask an Expert“ to submit the case in the support central: http://supportcentral.ge.com/products/sup_products.asp?prod_ id=44177. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-15 5726264-100 English Rev.8...
  • Page 40: Safety Considerations

    Operating personnel must not remove the system covers. • Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ V3/V5/V5 Expert Training Seminar are authorized to service the equipment. 1-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual...
  • Page 41 Because of the limited access to cabinets and equipment in the WARNING field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
  • Page 42 WARNING sharp edges are encountered, the appropriate PPE should be used to reduce the risk of injury. Wear all PPE including gloves as indicated in the chemical WARNING MSDS. 1-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 43: Mechanical Safety

    Take extra care when moving the system. CAUTION The LOGIQ V3/V5/V5 Expert weighs approximately 65 kg (144 lbs) or more, depending on installed peripherals, when ready for use. To avoid possible injury and equipment damage when transporting from one area of use to another: •...
  • Page 44 Do not move the Ultrasound system if the Operator Panel is in CAUTION unlocked position. Do not transport LOGIQ V3/V5/V5 Expert in a vehicle without CAUTION locking the casters (wheels) and securing it as described in chapter 4.
  • Page 45 Secure system with straps or as directed otherwise to prevent motion during transport. • Prevent vibration damage by driving cautiously. Avoid unpaved roads, excessive speeds, and erratic stops or starts. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-21 5726264-100 English Rev.8...
  • Page 46: Electrical Safety

    • Both the system power cable and the power connector must meet international electrical standards. Connecting a LOGIQ V3/V5/V5 Expert to the wrong voltage WARNING level will most likely destroy it. DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU WARNING UNIT LEVEL AT ANY CIRCUMSTANCES.
  • Page 47 The system has more than one type of probe port. Use the appropriate probe port designed for the probe you are connecting. 1-6-5-3 Peripherals Refer to the Patient Safety Environment section of the User’s Manual for peripheral isolation information. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-23 5726264-100 English Rev.8...
  • Page 48: Dangerous Procedure Warnings

    DEATH, ARE PRESENT IN THIS EQUIPMENT. USE EXTREME CAUTION WHEN HANDLING, TESTING AND ADJUSTING. If the covers are removed from an operating LOGIQ V3/V5/V5 WARNING Expert, some metal surfaces may be warm enough to pose a potential heat hazard if touched, even while in shutdown mode.
  • Page 49: Lockout/Tagout (Loto) Requirements

    4. Apply Lockout/Tagout Devices. 5. Control all stored and residual energy. 6. Verify isolation. All potentially hazardous stored or residual energy is relieved. Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1.
  • Page 50: Returning Probes And Repair Parts

    It is always possible during system failure and repair to lose patient data. GE is not responsible for the loss of this data. If PHI (Patient Healthcare Information) data needs to be sent to GE employees for service purposes, GE will ascertain agreement from the customer.
  • Page 51: Emc, Emi And Esd

    1-10-2 CE Compliance LOGIQ V3/V5/V5 Expert conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
  • Page 52: Electrostatic Discharge (Esd) Prevention

    Always use non-conductive handles designed for the removal and replacement of ESD sensitive parts. All parts that have the potential for storing energy must be discharged or isolated before making contact. 1-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 53: Customer Assistance

    5. Is the issue repeatable? 6. Imaging mode, probe, preset/application. 7. Media brand, speed, capacity, type. 8. Save secondary image capture, cine loop. NOTE: Restart the application before resuming clinical scanning. LOGIQ V3/V5/V5 Expert – Basic Service Manual 1-29 5726264-100 English Rev.8...
  • Page 54: Phone Numbers For Customer Assistance

    System manufacturer MANUFACTURER PHONE NUMBER FAX NUMBER GE Medical Systems (China) Co., Ltd. +86 510 85225888 +86 510 85226688 No.19 Changjiang Road WuXi National Hi-Tech Development Zone Jiangsu P.R.China 214028 1-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 55: Authorized Representative

    +49 761 45 43 -0 / Manual. +49 761 45 43 -233 Authorized EU Representative/European registered place of business: GE Medical Systems Information Technologies GmbH (GEMS IT GmbH) Munzinger Strasse 5, D-79111 Freiburg, GERMANY 1-11-5 Factory Site Table 1-9: Factory Site...
  • Page 56 Introduction 1-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 57: Chapter 2 - Site Preparations

    This chapter provides the information required to plan and prepare for the setup of an Ultrasound system. Included are descriptions of the facility and electrical needs to be met by the purchaser of the units. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 58: Overview

    Site Preparations 2-1 Overview 2-1-1 Contents in this chapter • 2-1 ‘Overview’ on page 2-2 • 2-2 ‘General Ultrasound system requirements’ on page 2-3 • 2-3 ‘Facility needs’ on page 2-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 59: General Ultrasound System Requirements

    Table 2-1: System Acclimation Time Chart Degree C -4.5 42.5 47.5 Degree F 23.9 28.4 32.9 37.4 108.5 117.5 hours LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 60 C * 9/5) + 32. Ensure that the probe face temperature does not exceed the CAUTION normal operation temperature range. The LOGIQ V3/V5/V5 Expert system and probe connector is CAUTION not waterproof. Do not expose the device to water or any kind of liquid.
  • Page 61 (dim/bright) is recommended. Keep in mind that lighting controls and dimmers can be a source of EMI which could degrade image quality. These controls should be selected to minimize possible interference. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 62: Electrical Requirements

    Site Preparations 2-2-3 Electrical requirements 2-2-3-1 General requirements NOTE: GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system. Sites with a mains power system with defined Neutral and Live:...
  • Page 63 2-2-3-6 Unit power plug If the Ultrasound system arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. LOGIQ V3/V5/V5 Expert – Basic Service Manual...
  • Page 64 70%T (30 dip) for 25 70%T (30 dip) for 25 environment. cycles; cycles; < 5%T (>95% dip) for 5 < 5%T (>95% dip) for 5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 65: Emi Limitations

    AC/DC adapters • portable phones The presence of a broadcast station or broadcast van may also cause interference. See: 2-2-4-1 ‘EMI prevention/abatement’ on page 2-10 for EMI prevention tips. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 66 Where applicable, do not allow cables to lie across the top of the Card Rack or peripheral cables hang out of the peripheral bays. Loop the excess length for peripheral cables inside the peripheral bays. Attach the monitor cables to the frame. 2-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 67: Probes Environmental Requirements

    2-2-6 Time and manpower requirements Site preparation takes time. Begin site preparation checks as soon as possible, if possible, six weeks before delivery, to allow enough time to make any changes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-11 5726264-100 English Rev.8...
  • Page 68: Facility Needs

    2-3-4 ‘Desirable features’ on page 2-15 • 2-3-5 ‘Minimal floor plan suggestion’ on page 2-16 • 2-3-6 ‘Recommended floor plan suggestion’ on page 2-17 • 2-3-7 ‘Networking setup requirements’ on page 2-18 2-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 69: Purchaser Responsibilities

    All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment. The desire to use a non–listed or customer provided product or...
  • Page 70: Required Facility Needs

    Site Preparations 2-3-3 Required facility needs NOTE: GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the Ultrasound system. Sites with a mains power system with defined Neutral and Live:...
  • Page 71: Desirable Features

    Storage for linens and equipment • Nearby waiting room, lavatory, and dressing room • Dual level lighting (bright and dim) • Lockable cabinet ordered by GE for its software and proprietary manuals LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-15 5726264-100 English Rev.8...
  • Page 72: Minimal Floor Plan Suggestion

    4. Linen Supply 11. Dedicated Power Outlet - 5. Probes/Supplies Circuit Breaker protected and easily accessible Figure 2-1. Minimal floor plan, 2.5 m x 3 m (8 by 10 foot) 2-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 73: Recommended Floor Plan Suggestion

    Dual Level Lighting (bright of Ultrasound system from inches) and dim) wall or objects 7. Emergency Oxygen 13. Stool Figure 2-2. A 14 by 17 foot recommended floor plan LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-17 5726264-100 English Rev.8...
  • Page 74: Networking Setup Requirements

    With DICOM, images can be archived, stored, and retrieved faster, easier, and at a lower cost. 2-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 75 Ultrasound system for DICOM APPLICATION INFORMATION. A field for the make (manufacturer) and the revision of the device, is also included. This information may be useful for error solving. LOGIQ V3/V5/V5 Expert – Basic Service Manual 2-19 5726264-100 English Rev.8...
  • Page 76 Site Preparations 2-3-7-5 DICOM option setup requirements(continued) Figure 2-3. Worksheet for DICOM Network Information 2-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 77: Chapter 3 - System Setup

    How to prepare the facility and unit of the actual installation, and how to check and test the unit, probes, and external peripherals for electrical safety are also included in this procedure. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 78: Overview

    3-5 ‘Completing the setup’ on page 3-14 • 3-6 ‘System Configuration’ on page 3-21 • 3-7-3 ‘Peripherals Installation Instructions’ on page 3-29 • 3-9 ‘Paperwork after setup’ on page 3-47 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 79: Setup Reminders

    3-2 Setup reminders 3-2-1 Average setup time • Unpacking the LOGIQ V3/V5/V5 Expert: 30 minutes • Set up LOGIQ V3/V5/V5 Expert options: 30 minutes • DICOM Network Configuration: 30 minutes or more, depending on the configuration • Install Insite: 0.5 hour The LOGIQ V3/V5/V5 Expert installation and functional checkout will take approximately 1 hour.
  • Page 80 The following table describes guidelines for reaching operational temperatures from storage or transport temperatures. Table 3-1: System Acclimation Time Chart Degree C Degree F hours LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 81 3-2-2 Setup warnings(continued) Operator Manual(s) CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ V3/V5/V5 Expert and kept near the Ultrasound system for quick reference. Acoustic Output Hazard CAUTION Although the ultrasound energy transmitted from the LOGIQ V3/V5/V5 Expert probe is within AIUM/NEMA standards, avoid unnecessary exposure.
  • Page 82: Receiving And Unpacking The Equipment

    System Setup 3-3 Receiving and unpacking the equipment 3-3-1 Purpose of this section This section describes how to receive and unpack LOGIQ V3/ V5/V5 Expert. 3-3-2 Contents in this section • 3-3-3 ‘Warnings for receiving and unpacking’ on page 3-6 •...
  • Page 83: Logiq V3/V5/V5 Expert Transportation Box Label

    Receiving and unpacking the equipment 3-3-4 LOGIQ V3/V5/V5 Expert transportation box label The LOGIQ V3/V5/V5 Expert transportation box label is located at the front of the transportation box. Figure 3-1. LOGIQ V3/V5/V5 Expert transportation box label 1 Figure 3-2. LOGIQ V3/V5/V5 Expert transportation box label 2 LOGIQ V3/V5/V5 Expert –...
  • Page 84: Unpacking The Logiq V3/V5/V5 Expert

    Chapter 1. Please carefully unpack the system, and do not dispose the CAUTION package of LOGIQ V3/V5/V5 Expert, so that it can be reused for service. Do not lift the unit by the Keyboard. Equipment damage may CAUTION result.
  • Page 85 Remove the top cover. Remove the three plastic locks. Note: Rotate the inside plastic lock counterclockwise to remove it and then remove the outside lock. Remove the outside shipping box. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 86 Ignore this step if there is no dust bag. Remove the clear plastic (wrapped around the LOGIQ V3/V5/V5 Expert) from the unit. Cut the packing straps around the four wheels. 3-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 87 3-3-5-1 Moving into Position Please refer to Basic User Manual on how to move the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-11 5726264-100 English Rev.8...
  • Page 88: Packing The Equipment

    System Setup 3-3-6 Packing the Equipment Please pack LOGIQ V3/V5/V5 Expert in the reverse order of unpacking. 3-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 89: Preparing For Setup

    3-4-3 EMI protection The LOGIQ V3/V5/V5 Expert has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference.
  • Page 90: Completing The Setup

    Verify that the site meets the requirements listed in Chapter 2. (See: 2-3 ‘Facility needs’ on page 2-12.) • Verify that the specifications below don’t conflict with any on-site conditions. 3-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 91: Completing The Setup

    Weight (system with 15" LCD Monitor): approx. 43 kg (95 lbs) • Weight (system with 17" LCD Monitor): approx. 45 kg (99 lbs) • Weight (system with 19" LCD Monitor): approx. 48 kg (106 lbs) LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-15 5726264-100 English Rev.8...
  • Page 92: Electrical Specifications

    3-5-4-1 Verification of the system’s voltage setting Verify that the mains voltage specified for the LOGIQ V3/V5/V5 Expert is available on-site. Refer to the latest revision of the Basic User Manual for a full list of product labels prior to serving the system.
  • Page 93: Connecting Cables

    3-5-5 Connecting Cables For software Option Installation, the operator must login as WARNING Administrator. Table 3-5: List of Connecting Cables. Name Part No. Figure NOTE For USB USB Cable Max.300VA Printer LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-17 5726264-100 English Rev.8...
  • Page 94: Peripherals/Accessories Connector Panel

    System Setup 3-5-6 Peripherals/Accessories Connector Panel LOGIQ V5/LOGIQ V3 peripherals and accessories can be properly connected using the side connector panel. Figure 3-3. Rear Connector Panel NOTE: The AC printer can be connected to the isolated Printer USB port on the rear panel only.
  • Page 95 Pin Assignment of S-Video Pin No. Output Signal Pin No. Output Signal Table 3-10: Pin Assignment of Composite Video Out Output Pin No. Output Signal Pin No. Signal Composite Out LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-19 5726264-100 English Rev.8...
  • Page 96: Connections On The I/O Rear Panel

    GE. 3-5-7-1 Connect Ethernet Connect the network cable to the Ethernet connector on the I/O Rear Panel. The connector is located on the rear side of LOGIQ V3/V5/V5 Expert. 3-5-7-2 Connect USB Flash Drive NOTE: USB Flash Drive approved for LOGIQ V3/V5/V5 Expert are verified for EMC performance according to EN55011 class B.
  • Page 97: System Configuration

    System Configuration 3-6 System Configuration 3-6-1 Purpose of this section This section describes how to configure the LOGIQ V3/V5/V5 Expert. 3-6-2 LOGIQ V3/V5/V5 Expert configuration For complete instructions, refer to the latest revision of the LOGIQ V3/V5/V5 Expert Basic User Manual, Chapter 16.
  • Page 98 Figure 3-5. Replace Key Cap There is a key reserved for CWD function, replace the blank key cap for CWD option if CWD option installed. Figure 3-6. CWD Key Cap 3-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 99 Configurable Key and assign the user defined key. Figure 3-7. Assign User Configurable Key 3-6-2-2 Multi-language Keyboard Film installation 1. Take out the keyboard film. Figure 3-8. Keyboard Film LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-23 5726264-100 English Rev.8...
  • Page 100 Figure 3-9. Peel Off the Paster 3. Make sure the keyboard is clean. Align the keyboard film on the keyboard and attach it. Figure 3-10. Attach The Film 3-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 101 System Configuration 3-6-2-2 Multi-language Keyboard Film installation(continued) 4. Peel off the other part of the paster and align the whole film on the keyboard. Figure 3-11. Peel Off the Paster LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-25 5726264-100 English Rev.8...
  • Page 102: Peripherals Installation

    System Setup 3-7 Peripherals Installation 3-7-1 Overview This section describes how to install and configure the peripherals validated for the LOGIQ V3/V5/V5 Expert. About the operation check-out of peripherals, 4-3-34 ‘Peripheral checks’ on page 4-43 for more information. Table 3-11:...
  • Page 103: Furnished Materials

    Materials furnished with USB Stick Item Description Quantity Note Keeber USB Stick USB cable • USB HDD Table 3-15: Materials furnished with the USB HDD Item Description Quantity Note USB HDD USB Cable LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-27 5726264-100 English Rev.8...
  • Page 104 Materials furnished with the Footswitch Item Description Quantity Note 1 Wireless Adaptor • DVD RW Table 3-18: Materials furnished with the Footswitch Item Description Quantity Note 1 DVD RW 1 Y cable 3-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 105: Peripherals Installation Instructions

    2. Ensure no physical damage. 3-7-3-2 Installation Procedure NOTE: The printer driver is customized for the LOGIQ V3/V5/V5 Expert at the factory, you do not need to change the settings. 1. UP-D711MD Printer can be properly connected using USB Ports.
  • Page 106 Figure 3-13. Power and USB Ports for DC printer a. DC Power Port b. USB POrt 2. Press Utility on the Control Panel. 3. Select Connectivity-> Service, add Standard Print. Figure 3-14. Connectivity->Service 3-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 107 Print2...) from the Physical Print Buttons selection. Select the printer from MyComputer column and press >> to move it to the Printflow View column. PressSave. Figure 3-16. Select Button LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-31 5726264-100 English Rev.8...
  • Page 108 Click Print -> Printing Preferences at the menu of Properties Window. Figure 3-18. Properties • Select Paper Size. Press Apply. Press OK. Figure 3-19. Printing Preferences • Press Save, then Exit. 3-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 109 2. Connect the USB Cable on the Printer. 3. Connect the power cord with the AC output in the wall outlet, then turn on the printer. 4. Connect USB cable to LOGIQ V3/V5/V5 Expert isolated USB port. Figure 3-20. Color Printer connection 5.
  • Page 110 2. Connect the USB Cable on the Printer. 3. Connect the power cord with the AC output in the wall outlet, then turn on the printer. 4. Connect USB cable to LOGIQ V3/V5/V5 Expert via the USB port under the control panel. Figure 3-21. UP-D898 Connection 5.
  • Page 111 Figure 3-22. Screw Screws 2. Mount the printer shelf on the system and make sure the 2 screws are in the fixed position. Figure 3-23. Mount the Printer Shelf LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-35 5726264-100 English Rev.8...
  • Page 112 System Setup 3-7-3-5-4 Installation Procedure(continued) 3. Fix 2 screws on both left and right sides to fix the printer shelf. Figure 3-24. Fix the Printer Shelf 3-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 113 1. Unpack the Footswitch. 2. Ensure no physical damage. 3-7-3-6-4 Installation Procedure 1. Connect the Footswitch to the USB port on the LOGIQ V3/ V5/V5 Expert system. Figure 3-25. Connect Footswitch to the system LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-37 5726264-100 English Rev.8...
  • Page 114 Next Heartcycle, Prev Heartcycle, Print1, Print2, Update, Next Step (Scan Assistant), Previous Step (Scan Assistant), Scan Assistant pause/Resume. Enter Utility -> Application-> Settings to configure the Footswitch functions. Figure 3-26. Configuring Footswitch Functions 3-38 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 115 1. Unpack the USB HDD. 2. Ensure no physical damage. 3-7-3-7-4 Installation Procedure 1. Connect the USB HDD to the USB port on the LOGIQ V3/ V5/V5 Expert system. Figure 3-27. Connect HDD to the system LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-39 5726264-100 English Rev.8...
  • Page 116 1. Unpack the ECG. 2. Ensure no physical damage. 3-7-3-8-4 Installation Procedure 1. Connect the ECG to the USB port on the LOGIQ V3/V5/V5 Expert system and the cables can be hang in the cable hook. Figure 3-28. Connect ECG to the system 3-40 LOGIQ V3/V5/V5 Expert –...
  • Page 117 1. Take out the DVDRW and USB Y cable and connect the USB Y cable to the DVDRW. Figure 3-29. Connect Y cable 2. Insert the DVDRW in the printer shelf right below the printer. Figure 3-30. Insert the DVDRW LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-41 5726264-100 English Rev.8...
  • Page 118 System Setup 3-7-3-9-4 Installation Procedure(continued) 3. Connect the two connectors on the Y cable to the USB ports on the system. Figure 3-31. DVD RW Connector 3-42 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 119 1. Unpack the wireless adaptor 2. Ensure no physical damage. 3-7-3-10-4 Installation Procedure 1. Connect the wireless adaptor to the USB port on the LOGIQ V3/V5/V5 Expert system. Figure 3-32. Wireless Adaptor Connector LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-43 5726264-100 English Rev.8...
  • Page 120: Option Setup

    3-8-1 Software Option Installation Procedure NOTE: Not all features described in this section may be available or cleared for sale in all markets. Please contact with your local GE Ultrasound representative to get the latest information. 1. Power on the system.
  • Page 121 5. After the system is powered on, check the option status. Figure 3-35. Check Option Status The option status explanation: • Permanent: This option is enabled in the system. • Disabled: This option is disabled in the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 3-45 5726264-100 English Rev.8...
  • Page 122 Figure 3-36. Replace Caps 7. This completes the option installation of LOGIQ V3/V5/V5 Expert. NOTE: Installing corresponding hardware is a prerequisite of loading software option successfully.
  • Page 123: Paperwork After Setup

    During and after setup, the documentation (i.e. CDs with documentation, User Manuals, Installation Manuals, etc.) for the LOGIQ V3/V5/V5 Expert and the peripherals must be kept as part of the original Ultrasound system documentation. This ensures that all relevant safety and user information is available during the operation and service of the complete Ultrasound system.
  • Page 124 System Setup 3-48 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 125: Chapter 4 - General Procedures And Functional Checks

    Chapter 4 General Procedures and Functional Checks This chapter provides procedures for quickly checking major functions of the LOGIQ V3/V5/V5 Expert and diagnostics instructions using the built-in service software. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 126: Overview

    4-3 ‘Functional checks’ on page 4-21 4-1-3 Special Equipment required To perform these tests, you'll need any of the sector, linear, or convex probes. (Normally you should check all the probes used on the system). LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 127: General Procedures

    The covers are required for safe operation, good Ultrasound system performance and cooling purposes. Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1.
  • Page 128: Overview

    4-2-2 ‘Power ON/Boot Up’ on page 4-5 • 4-2-3 ‘Power off’ on page 4-8 • 4-2-6 ‘Removable media’ on page 4-11 • 4-2-7 ‘Backup and Restore Database, Preset Configurations and Images’ on page 4-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 129: Power On/Boot Up

    Do not cycle the Circuit Breaker ON-OFF-ON in less than five (5) seconds. When turning OFF the Circuit Breaker, the Ultrasound system should de-energize completely before turning the circuit breaker ON. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 130 General Procedures and Functional Checks 4-2-2-2 Connect AC (mains) Power to LOGIQ V3/V5/V5 Expert Connecting AC Power to the LOGIQ V3/V5/V5 Expert ultrasound unit, involves preliminary checks of the power cord, voltage level and compliance with electrical safety requirements. 1. Ensure that the wall outlet is of appropriate type, and that the Circuit Breaker is turned off.
  • Page 131 General procedures 4-2-2-3 Switch ON the AC Power to LOGIQ V3/V5/V5 Expert 1. Switch ON the Mains Power Circuit Breaker at the rear of the unit. You should hear a “click” from the relays in the AC Power and the unit is ready to boot.
  • Page 132: Power Off

    It is recommended to perform a full shutdown at least once a week. If the Shutdown button is dimmed, use the key-combination <Ctrl+Alt+Delete> to shut down the unit. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 133 Relocating the scanner. DO NOT unplug and/or transport the unit until after the power CAUTION off sequence has been completed. Failure to do so may result in corrupted patient files. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 134: Check System Date And Time

    • Enter Utility -> System -> General -> Date/Time. • Reset the system date and time. • Select Apply and then select OK. • Select Save. 4-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 135: Manufacturer

    Refer to the latest revision of the User Manual to perform the following tasks: • Using Removable Media • Labeling Removable Media • Formatting Removable Media • Verifying Removable Media LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-11 5726264-100 English Rev.8...
  • Page 136 Presets should NOT be saved on the same USB memory device (or DVD-R disk) as images. The Archive Menu lists the images but does NOT list the presets stored on a USB memory device (or DVD-R disk). 4-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 137 4-6, select OK to continue. Figure 4-6. Format Warning Pop-up Window 5. If desired, verify that the format was successful by returning to Utility-> Connectivity->Removable Media and selecting Verify. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-13 5726264-100 English Rev.8...
  • Page 138 1. Insert a formatted media into the drive. 2. Enter enter Utility-> System-> Backup/Restore. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 3. In the Backup list, select Patient Archive, Report Archive, User Defined Configuration and Service.
  • Page 139 4. In the Media field, select the Backup/Restore CD/DVD (or USB memory device). 5. Select Restore. The system performs the restore. As it proceeds, status information is displayed on the Backup/Restore screen. Figure 4-8. Restore message LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-15 5726264-100 English Rev.8...
  • Page 140 (backed up earlier) maintains pointers to the location of the images on the archive media. Figure 4-9. Format Media Screen 5. Enter Utility-> System-> Backup/Restore. 6. Select “Move File Older Than in Days”. 4-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 141 (even if the exam was previously up), check this option. If you uncheck this option, the system only backs up exams which have not yet been backed up. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-17 5726264-100 English Rev.8...
  • Page 142 EZBack/EZMove. Figure 4-13. EZBackup Wizard 2 NOTE: This message “Please insert a blank media...” appears if you press Next without inserting the backup media. Insert the media and continue. 4-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 143 If you need to insert the next media, a message appears providing you with the media label. 11. When the backup is complete, the completion wizard page appears. Press Finish. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-19 5726264-100 English Rev.8...
  • Page 144 General Procedures and Functional Checks 4-2-7-4 Archiving Images(continued) Figure 4-16. EZBackup Completion Window All databases, presets and images should now be saved to removable media. 4-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 145: Functional Checks

    Functional checks 4-3 Functional checks 4-3-1 Overview In this section, the functional checks for LOGIQ V3/V5/V5 Expert are described. Functional checks are used to verify that the product works as intended. Functional checks may also be used during troubleshooting. 4-3-2 Contents in this Section 4-3-3 Preparation Turn on power to LOGIQ V3/V5/V5 Expert.
  • Page 146: Basic Controls

    21. Set/B Pause key 33. M Mode/M Mode Gain 11. End Exam key 22. Freeze key 34. CW (Option) 12. Alphanumeric Keyboard 23. Print key 35. PDI 36. CF Mode/CF Gain/PDI Gain 4-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 147: Logiq V5/Logiq V3 Softmenu Key Tour

    Functional checks 4-3-6 LOGIQ V5/LOGIQ V3 SoftMenu Key Tour • Rotary key: Rotate to adjust the menu. • Page Up/Down key: To turn the menu page up and down. Figure 4-18. SoftMenu Key Tour LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-23 5726264-100 English Rev.8...
  • Page 148: Control Panel

    13. Current date and time, Caps Lock: (lit when on), 27. Body Pattern. network connection indicator (PC=connected, PC with X=not connected), system messages display, InSite status, InSite controls. Image Preview 4-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 149: Performance Tests

    Functional checks 4-3-8 Performance Tests 4-3-8-1 Test Phantoms The use of test phantoms is only recommended if required by your facility's (customer’s) QA program. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-25 5726264-100 English Rev.8...
  • Page 150: Mode (B Mode) Checks

    Density (move to higher line density) to increase resolution in image. • Use Frequency (move to lower frequency) to increase penetration. • Use the Rejection control to reduce noise in the image. 4-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 151: Image Quality Check (Iqc) Preset For Service

    Figure 4-20. Check “Enable IQC for Service” 2. Press Utility -> Scan Coach -> Imaging Preset Manager. Select IQC from the left column, and select move it to the Imaging Preset Selection. Then Select Exit. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-27 5726264-100 English Rev.8...
  • Page 152 4-3-13 Image Quality Check (IQC) preset for service(continued) Figure 4-21. Image Quality Check 3. Press Probe/Preset on the control panel. Select SVC, then select Figure 4-22. Image Quality Check 4-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 153: M Mode Checks

    Adjust Gain and TGC controls to adjust the range to be imaged. Adjust Compression and Edge Enhance to further optimize the display. Adjust Rejection to reduce noise while taking care not to eliminate significant low-level diagnostic information. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-29 5726264-100 English Rev.8...
  • Page 154: Color Mode Checks

    2. Use the trackball to position the ROI frame over the area to be examined. 3. Press Scan Area. The instruction Size should be highlighted in the trackball status bar. Use the trackball to adjust the dimension of the ROI. 4-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 155: Adjust The Color Mode Controls

    Spatial Filter smooths the image by averaging collected data along the same horizontal line. An increase of the spatial filter will reduce noise, but this will also reduce the lateral resolution. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-31 5726264-100 English Rev.8...
  • Page 156: Select Color M Mode

    2. Use the trackball to position the ROI frame over the area to be examined. 3. Press Scan Area. The instruction Size should be highlighted in the trackball status bar. Use the trackball to adjust the dimension of the ROI. 4-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 157 Use Threshold to emphasize either the color flow overlay, or the underlying grey scale tissue detail. Use Baseline to emphasize flow either toward or away from the probe. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-33 5726264-100 English Rev.8...
  • Page 158: Pw Doppler Mode Checks

    If needed, adjust the Display’s Brightness and Contrast setting. Press PW to start Pulsed Wave Doppler (PW). Use the trackball to select the Area of Interest (Sample Volume) in PW. 4-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 159: Adjust The Pw Doppler Mode Controls

    Use Invert to reverse the vertical component of the spectral Doppler area of the display. Use Angle Correct to steer the ultrasound beam to the blood flow to be measured. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-35 5726264-100 English Rev.8...
  • Page 160: Tissue Velocity Imaging (Tvi) Checks

    If the trackball control pointer is selected, press trackball to be able to select between Position and Size controls. Use the trackball to adjust the dimension of the ROI. 4-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 161: Adjust The Tvi Controls

    However, from this window the beam cannot be aligned to the muscle for all the parts of the ventricle. NOTE: PW will be optimized for Tissue Velocities when activated from inside TVI. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-37 5726264-100 English Rev.8...
  • Page 162: Basic Measurements

    To rotate through and activate previously fixed calipers, rotate Cursor Select button. NOTE: After you complete the measurement, to erase all data that has been measured to this point, but not data entered onto worksheets, press Clear. 4-38 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 163: Probe/Connectors Check

    Launch the application. To change application, The selected application starts. press Probe/Preset key on the Control Panel. Verify there’s no EMI/RFI or artifacts specific to the No EMI/RFI or artifacts. probe. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-39 5726264-100 English Rev.8...
  • Page 164 5. Be sure that the probe head is clean before placing the probe in its storage box. Take the following precautions with the probe cables: Do not WARNING bend, be sure to keep probe cables free from the wheels. 4-40 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 165: Ecg Check

    Fasten the three ECG Pads to your body and When the connection is completed, a typical connect the three leads to respective ECG clean ECG signal is displayed. Pad. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-41 5726264-100 English Rev.8...
  • Page 166: Cineloop Check

    If all the previous tests have been passed successfully, the Back End Processor is most likely OK. If the system seems to be operating erratically, please refer ‘Diagnostics/Troubleshooting’ on page 7-1. 4-42 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 167: Operator Panel Test

    Functional checks 4-3-33 Operator Panel Test • The Operator Panel is tested when the LOGIQ V3/V5/V5 Expert is powered up as part of the start-up scripts, run at every start-up. For more info, please refer to ‘Diagnostics/Troubleshooting’ on page 7-1.
  • Page 168 4-3-36-2 Casters (Wheels) and Brakes Checks Examine the wheels frequently for defects to avoid breaking or jamming. Table 4-4: Wheel Characteristics Wheel Characteristics Front and Rear Swivel and Brake 4-44 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 169: Application Turnover Check List

    Verify that Location (Hospital Name) is correct. Verify Language settings are correct. Verify assignment of Print Keys. Verify all of the customer’s options are set up Demo Option strings turn on correctly. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-45 5726264-100 English Rev.8...
  • Page 170: Site Log

    General Procedures and Functional Checks 4-5 Site Log Table 4-6: Site Log DATE SRVICE PERSON PROBLEM COMMENTS 4-46 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 171: My Trainer

    2. Press defined key to access My Trainer. There are four sections in My Trainer. The four sections are displayed on the left side of the My Trainer interface. LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-47 5726264-100 English Rev.8...
  • Page 172 3. Peripheral Connection: Back panel ports, Network Connectivity, Connect to a printer, DICOM setting, Back-up/ restore the patient information, software version 4. Maintenance: Trackball, Contact GE service, Insite, Log export, Air Filter Replacement 4-48 LOGIQ V3/V5/V5 Expert – Basic Service Manual...
  • Page 173 6. The left number shows the current page of the subsection. The right number shows the total pages of the subsection. 7. Illustration with graphic. Table 4-7: Page turning Select the arrow to go to the previ- ous page LOGIQ V3/V5/V5 Expert – Basic Service Manual 4-49 5726264-100 English Rev.8...
  • Page 174 Select the arrow to move to the next page. Select the arrow to go to the home page Select the arrow to go to the end page. 9. Text instruction about how to operate. 4-50 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 175: Chapter 5 - Components And Functions (Theory)

    Chapter 5 Components and Functions (Theory) This chapter explains LOGIQ V3/V5/V5 Expert’s system concepts, component arrangement, and subsystem functions. It also describes the power distribution and the Common Service Desktop interface. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 176: Overview

    Components and Functions (Theory) Overview 5-1-1 Contents in this chapter • 5-1 ‘Overview’ on page 5-2 • 5-2 ‘Block Diagram and Theory’ on page 5-3 • 5-3 ‘Common Service Platform’ on page 5-5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 177: Block Diagram And Theory

    Block Diagram and Theory Block Diagram and Theory 5-2-1 Block Diagram 5-2-1-1 System Diagram Figure 5-1. LOGIQ V3/V5/V5 Expert System Block Diagram LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 178: Introduction

    ACDC and provides all voltage needed to WMST board and SMIO board. SFAN PWA is a system FAN interface board, it provides power and control to system FAN (quantity is 4 internal the box). LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 179 The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. The Common Service Platform will increase service productivity and reduce training and service costs. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 180 Components and Functions (Theory) LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 181: Chapter 6 - Service Adjustments

    Chapter 6 Service Adjustments This chapter describes how to test and make adjustments to the LOGIQ V3/V5/V5 Expert. You can use these to test the system for errors. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 182: Overview

    Service Adjustments 6-1 Overview 6-1-1 Contents in this chapter • 6-1 ‘Overview’ on page 6-2 • 6-2 ‘LCD Monitor Adjustments’ on page 6-3 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 183: Lcd Monitor Adjustments

    In rare cases, LCD monitor might have defect pixels appearing. This section describes how to discriminate the defect pixels with the number of defect pixels allowed for the LOGIQ V3/V5/V5 Expert to support LCD monitor replacement decision. Monitor Adjustments...
  • Page 184 To adjust the Contrast: 1. Press Middle key twice, the contrast indicator displays; Figure 6-4. Contrast setting menu 2. Press Left key or to decrease contrast or press right key to increase contrast. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 185 Keep pressing the middle key for 3 to 5 seconds, the Brightness and Contrast are reset. Figure 6-5. Reset setting menu Press the Right key once to activate main menu to have more settings. Figure 6-6. Main Menu LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 186 • To adjust the color temperature: press Standard, Brown, or Blue to choose the color temperature. The chosen color temperature displays in the Active area. Figure 6-8. Active Color Temperature LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 187 Move the adjustment button to the left to decrease the contrast or move the adjustment button to the right to increase the contrast. Figure 6-10. Contrast LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 188 3. Press Reset to return to the default setting; Press Exit to exit the setting menu. Figure 6-12. Reset NOTE: Press Reset button and the three color temperature settings will return to the default settings. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 189: Chapter 7 - Diagnostics/Troubleshooting

    Chapter 7 Diagnostics/Troubleshooting This chapter describes LOGIQ V3/V5/V5 Expert Series how to setup and run the tools and software that help maintain image quality and system operation. Very basic host, system and board levels are run whenever power is applied. Some Service Tools may be run at the application level.
  • Page 190: Overview

    7-2 ‘Gathering Trouble Data’ on page 7-3 • 7-3 ‘Screen Capture’ on page 7-5 • 7-4 ‘Global Service User Interface (GSUI)’ on page 7-12 • 7-5 ‘Network and Insite II Configuration’ on page 7-29 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 191: Gathering Trouble Data

    Product Name = LOGIQ V3/V5/V5 Expert From the Utility>System>General>About screen: Applications Software • Software Version • Software Part Number System Image Software • Image Date • Image Part Number Additional Information LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 192: Collect A Trouble Image With Logs

    Options enables creation of a log file based on a selected bookmark or for a user configurable time frame. Different type of information can be selected to be part of the log file. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 193: Screen Capture

    There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Store Key. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 194: Capturing A Screen

    1. Navigate to and display the image/screen to be captured. 2. Press Store. This will place a snapshot of the screen on the “clipboard” displayed at the bottom of the scan image display. Figure 7-2. Select Image to Capture LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 195 USB Drive or CD/DVD. Figure 7-4. Save Dialog Box NOTE: Click Save -> Transfer To CD/DVD, if the archive location on the DVD/CD; Click Save, if the archive location is USB. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 196 Diagnostics/Troubleshooting 7-3-1 Capturing a Screen(continued) NOTE: It is better to save the image in Jpeg format. Image of this format can be easily reviewed in the computer. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 197: Capturing A Screen With Service Dongle

    3. Press Ctrl+Alt+Delete, the Windows Security dialog is displayed. Then select Task Manager. Figure 7-5. Windows Security NOTE: It is also available to Exit to Windows and save the image on the system or removal media. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 198 7-3-2 Capturing a Screen with Service Dongle(continued) 4. In Windows Task Manager, select File-> New Task (Run...). Figure 7-6. Windows Task Manager 5. Type in the program name: mspaint. Figure 7-7. Create New Task 7-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 199 7-3-2 Capturing a Screen with Service Dongle(continued) 6. The paint screen is displayed, press Ctrl+V to paste the captured screen. And then save the image in the appropriate destination. Figure 7-8. Save Image LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-11 5726264-100 English Rev.8...
  • Page 200: Global Service User Interface (Gsui)

    7-4-1 Enter global service user interface 1. Select Utility -> Service, then the service browser will start. Figure 7-9. 2. Select GE Service in option of User Level of GEMS Service Home page and input correct password, press OK button. 7-12 LOGIQ V3/V5/V5 Expert –...
  • Page 201 7-4-1 Enter global service user interface(continued) Figure 7-10. Service Login Page 3. Choose Diagnostic in Global Service Interface to active diagnostic functions, choose options to activate various functions correspondingly. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-13 5726264-100 English Rev.8...
  • Page 202 Diagnostics/Troubleshooting 7-4-1 Enter global service user interface(continued) Figure 7-11. Active Diagnostic Function 7-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 203: Control Frame

    • Fail = Red • Pass = Green • Neither pass nor fail = Set back to Gray (for example, final code status is Aborted). LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-15 5726264-100 English Rev.8...
  • Page 204: Common Diagnostics

    1. In the server side, select the system which the OLE would remotely connect to, and select Connect to device. Figure 7-12. Connect to device 2. Select CSD to Connect. Then the OLE is remotely connected to the system. 7-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 205 Global Service User Interface (GSUI) 7-4-3-1-27-4-3-1-1 Remote access(continued) Figure 7-13. Remotely Connect 3. Enter Utility-> Common Utilities-> Disruptive Mode, then select Yes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-17 5726264-100 English Rev.8...
  • Page 206 Virtual Console Observation and Start it. Then connect using VNC. Figure 7-15. Virtual Console Observation 5. Select the VNC to Connect. Then the OLE is remotely connected to the system. 7-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 207 Global Service User Interface (GSUI) 7-4-3-1-47-4-3-1-1 Remote access(continued) Figure 7-16. Virtual Console Observation Connect 7-4-3-2 Service Diagnostics BF, COMEXP and GFS modules are on the LOGIQ V3/V5/V5 Expert WMST PWA. CPS is WPWR/DCDC PWA. IOS is SMIO PWA. PID is SLRY 4-Probe PWA.
  • Page 208 Front End Interface FPGA Test reads the version of the GFE FPGA. • GFS Fixed Demo Test: GFS Fixed Demodulator signal path test. • DSP Master Clock Test tests DSP Master Clock. • GFS Swept Demodulator Long Test. 7-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 209 AFE IF Test • Complex Mixer Test • CE Decoder Test 7-4-3-2-7 ECG Tests • ECG Tests 7-4-3-2-8 Doppler Tests • Doppler Audio Test runs the tests for Doppler. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-21 5726264-100 English Rev.8...
  • Page 210 • Note: This diagnostic is intended to verify keyboard keys are in good working order. It is not intended to verify that keyboards produce desired characters. 7-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 211 Click the appropriate button to fill the screen with the associated test pattern. You can return to this dialog by clicking the mouse button or pressing any key. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-23 5726264-100 English Rev.8...
  • Page 212 Double-click on the picture of the monitor. If successful, the picture will change. • Sound Test generates sounds for testing the speakers. • USB Ports Test lists USB Devices. 7-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 213 7. Press Probe button on the control panel and select the first probe from the probe indicators. 8. Enter global service user interface as GE service. LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-25 5726264-100 English Rev.8...
  • Page 214 Utility-> Probe Assessment to check the probe. If there is no broken element and the probe is acceptable for diagnostic, the following result window displays. Figure 7-18. Probe Assessment Pass 7-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 215 If there are broken elements and the probe is not acceptable for diagnostic. The assessment is fail and the failed elements are shown in the graphic. Figure 7-19. Probe Assessment Fail Reboot the system after the probe assessment. CAUTION LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-27 5726264-100 English Rev.8...
  • Page 216 Max 2 weak elements on the scan line: -3dB<elements<-2dB • Any one of the elements in one scan line: element<-3dB 7-4-3-3 Restart the system after diagnostics Always shutdown the system and reboot after a diagnostics session. 7-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 217: Network And Insite Ii Configuration

    1. Connect system with network. 2. Enter Utility-> Connectivity-> TCP/IP, in IP settings window, check Enable DHCP, and select the proper network speed in Network Speed. Figure 7-20. Enable DHCP LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-29 5726264-100 English Rev.8...
  • Page 218 Figure 7-22. System Restart inquiry dialog 4. After the system restarts, the network icon at the left bottom of screen shows that the network is available. 7-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 219 Network and Insite II Configuration 7-5-1 Network Configuration(continued) Figure 7-23. Network icon LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-31 5726264-100 English Rev.8...
  • Page 220 2. Enter Utility-> Connectivity-> TCP/IP, press Configuration. Figure 7-24. Network Settings 3. Double click Wireless network. you chosen, or clink the Connect at the left bottom of screen. Figure 7-25. Wireless Network Conection 7-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 221 4. Input Network Key and Confirm Network Key, then click Connect. Figure 7-26. Connection Status 5. The network icon at the left bottom of screen shows that the wireless network is available. Figure 7-27. Wireless-LAN Connection LOGIQ V3/V5/V5 Expert – Basic Service Manual 7-33 5726264-100 English Rev.8...
  • Page 222 Diagnostics/Troubleshooting 7-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 223: Chapter 8 - Replacement Procedures

    Chapter 8 Replacement Procedures This chapter describes how to remove and install, or replace, modules and subsystems in the LOGIQ V3/V5/ V5 Expert. It also includes instructions for installing and re-installing the software. LOGIQ V3/V5/V5 Expert – Basic Service Manual...
  • Page 224: Overview

    8-1 Overview 8-1-1 Contents in this chapter • 8-1 ‘Overview’ on page 8-2 • 8-2 ‘Warnings and important information’ on page 8-3 • 8-3 ‘Loading the software’ on page 8-5 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 225: Warnings And Important Information

    Because of the limited access to cabinets and equipment in the WARNING field, placing people in awkward positions, GE has limited the lifting weight for one person in the field to 16 KG (35 LBS). Anything over 16 KG (35 LBS) requires 2 people.
  • Page 226: Returning/Shipping Probes And Repair Parts

    Equipment being returned must be clean and free of blood and other infectious substances. GE policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE employees, as well as customers, are responsible for ensuring that parts/equipment have been properly decontaminated prior to shipment.
  • Page 227: Loading The Software

    Loading the software 8-3 Loading the software 8-3-1 Purpose of this section This section describes how to reinstall software on LOGIQ V3/ V5/V5 Expert. 8-3-2 Customer provided prerequisite • Formatted and labelled media for Images storage. • Formatted and labelled media for Patient Archive and Presets (User Defined Settings).
  • Page 228: Backing Up The Patient Archive And System Configurations

    Database, as needed after a hard disk replacement, or if all the content on the hard disk has been erased, the images must be moved away from LOGIQ V3/V5/V5 Expert before doing backup of the Patient Database. Depending on the location set-up, either move the images to a remote server or to removable media like DVD or CD discs.
  • Page 229: Recording Important Settings And Parameters

    CAUTION It is considered to be best practice to always keep a record on paper of the settings for the LOGIQ V3/V5/V5 Expert. Verify if it is current before you start to load software! Always keep a record of the settings for the LOGIQ V3/V5/V5 Expert on paper.
  • Page 230: Loading The System Software

    If the system will not shutdown normally, hold down the Power On/Off Switch until the light turns off. 3. Power on the system. The system will detect the USB memory stick automatically. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 231 (Customer Presets) will be lost. • To select [4], Screen size. This option is to adjust the size of screen display. NOTE: All the patient data cannot be recovered after wipe tool. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 232 5. System USB memory stick will be loading. Wait for the software installation to complete. (Typical installation time: 10-15 minutes). Status bar on the screen indicates progress. Do not interrupt the software loading at any time. WARNING 8-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 233 If you do not remove the USB memory stick, the software system loading process repeats when the system boots up.Do not remove the USB stick after the software reload if you want to adjust the screen size. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-11 5726264-100 English Rev.8...
  • Page 234 Power On/Off Switch. In System-Exit window, select Shutdown to shutdown the system. NOTE: If the system will not shutdown normally, hold down the Power On/Off Switch until the light turns off. 8-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 235 Figure 8-4. Upgrade message 5. The system indicates all data will be erased, select Yes to continue. Figure 8-5. Confirm Dialog NOTE: All the patient data cannot be recovered after wipe tool. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-13 5726264-100 English Rev.8...
  • Page 236 8-3-7-2 Wipe Tool(continued) 6. The process may take twenty minutes or more. Figure 8-6. Wipe Tool Process 7. Select OK when the process is complete, and the system will shutdown automatically. 8-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 237 MUST be the same as the ones before installation. Please record the serial number and option key before installation. 1. Insert the USB memory stick labeled “System & Application Software” to the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-15 5726264-100 English Rev.8...
  • Page 238 If you select [3], ALL existing software and data will be erased. CAUTION If backup has not been performed, all data like Patient Database, System Configuration and User Configurations (Customer Presets) will be lost. Figure 8-8. System Software load instruction 8-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 239 GEWuXiChina and it is case sensitive. Figure 8-9. System upgrade complete 6. Click Next to continue this operation. Figure 8-10. System upgrade complete 7. Select Yes. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-17 5726264-100 English Rev.8...
  • Page 240 8. System USB memory stick will be loading. Wait for the software installation to complete. NOTE: Do not interrupt the software loading at any time. Figure 8-12. System upgrade complete 8-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 241 NOTE: The serial number and option key MUST be the same as the ones before installation. Please record the serial number and option key before installation. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-19 5726264-100 English Rev.8...
  • Page 242 2. Power on the system. The system will detect the USB memory stick automatically. 3. Select choice [4] -> Next. Figure 8-14. Screen size 4. Enter the ultrasound maintenance access password.The password is GEWuXiChina and it is case sensitive. 8-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 243 NOTE: If you do not remove the USB memory stick, the software system loading process repeats when the system boots up. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-21 5726264-100 English Rev.8...
  • Page 244: Rewrite The Serial Number

    1. If the serial number on VPD and HDD does not match, the following dialog will display when the system is booting up.The password is the same as that of service dongle. Figure 8-18. Input Password 8-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 245 The serial number is case sensitivity. Figure 8-20. Input Serial Number NOTE: The serial number can be found on the rating plate on the rear side of the system. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-23 5726264-100 English Rev.8...
  • Page 246 8-3-8 Rewrite the Serial Number(continued) 3. The serial number is reset, reboot the system. Figure 8-21. Reboot the System NOTE: Make sure the system is shut down after the serial number is reset. 8-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 247 Loading the software 8-3-8-1 Software Version check out 8-3-8-1-1 Functional Check-out 1. Power on LOGIQ V3/V5/V5 Expert system and wait until system booting to scanning screen. NOTE: If selected to load C Disk only when loading the system software, the system will display a screen to restore Computer name before entering the scanning screen.
  • Page 248 2. Press Utility on the Control Pane, then select System -> General. 3. In the Location portion, select the right region in the field of Preset Region. Figure 8-23. Preset Region 4. Reboot the system. 8-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 249 The status “disabled” means the option keys are not activated and not working. Check if the option is installed and if the serial number and option key are correct. Figure 8-24. Software Option LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-27 5726264-100 English Rev.8...
  • Page 250 Plug in the probe. In scanning mode, the probe information is displayed on the right top location of the screen. About the probe specification for intended use on LOGIQ V3/V5/V5 Expert. Plug in at least one of each type of the probes and check if each of the probes is recognized and the probe information is displayed correctly.
  • Page 251 The instruction about installing DICOM devices is not incorporated in this manual. To access the instruction about installing DICOM devices please refer to another manual Basic User Manual. Please use the latest revision of this document. LOGIQ V3/V5/V5 Expert – Basic Service Manual 8-29 5726264-100 English Rev.8...
  • Page 252 Replacement Procedures 8-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 253: Chapter 9 - Renewal Parts

    Chapter 9 Renewal Parts This chapter lists the renewal parts available for the LOGIQ V3/V5/V5 Expert. LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 254: Overview

    Renewal Parts 9-1 Overview 9-1-1 Contents in this chapter • List of Abbreviations • Renewal Parts Lists LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 255: List Of Abbreviations

    ASSEMBLY FRU 1 Replacement part available in part hub FRU 2 Replacement part available from the manufacturer (lead time involved) Keyboard Liquid Crystal Display Brightness and Volume WMST Master Board LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 256: Renewal Parts Lists

    9-3-1 Operator Console Assy The following figure illustrates what is the front side (1), left side (2), back/rear side (3), right side (4) of the system. Figure 9-1. Operator Console Assy LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 257: Probe

    ECG Detachable cable AHA Type USA 5146739 ECG detachable cable IEC type EURO and ASIA 5146055 USB Cable for ECG 5129487 ECG module from Norav Isral (not applies to Muslim country and China) LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 258 RAB2-6-RS Biopsy Kit E8385MJ E8C-RS Biopsy Kit USB HDD 5434317-4 1TB mobile USB HDD DVD RW 5729697 LV R2 DVD RW Kit with SW 5653589 LITEON eBAU108 DVD Writer Kit LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 259 Ayan R2 R2.0.7 System and Application Soft- ware (for Service) Others H48702AS VGA to Svideo adapter H48822AC LF/LV AN Key film H48612AS LV R2 CHN KBD Film H48832AC Wireless Adaptor H48552LJ Printer Power Cable for Brazil LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 260: Power Cord

    5177154-2 AC Power Cord Switzerland 5177187-3 AC Power Cord Australia 5177146-2 AC Power Cord USA 5400868-2 AC Power Cord Brazil 5176753-2 AC Power Cord Israel 5177126-2 AC Power Cord Japan LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 261: Manuals

    9-3-5 Manuals Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1000 5729745 -100 LOGIQ V3/V5/V5 Expert Quick Card for USA 1001 5726265-200 LOGIQ V3/V5/V5 Expert Manual CD 1002 5719190-100 LOGIQ V3/V5/V5 Expert Advanced Reference Manual 1003 5726264-100...
  • Page 262 Renewal Parts Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1028 5719188-167 LOGIQ V3/V5/V5 Expert User Guide Romanian version 1029 5719188-168 LOGIQ V3/V5/V5 Expert User Guide Croatian version 1030 5719188-174 LOGIQ V3/V5/V5 Expert User Guide Lithuanian version 1031...
  • Page 263 Renewal Parts Lists Table 9-5: Manuals for LOGIQ V3/V5/V5 Expert Item Part Number Description 1057 5726260-155 LOGIQ V3/V5/V5 Expert Release Notes Czech version 1058 5726260-159 LOGIQ V3/V5/V5 Expert Release Notes Turkish version 1059 5726260-160 LOGIQ V3/V5/V5 Expert Release Notes Danish version 1060...
  • Page 264 Renewal Parts 9-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 265: Chapter 10 - Care And Maintenance

    Ultrasound system and peripherals. These procedures are intended to maintain the quality of the Ultrasound system’s performance. Read this chapter completely and familiarize yourself with the procedures before performing a task. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-1 5726264-100 English Rev.8...
  • Page 266: Overview

    10-7 ‘Electrical safety tests’ on page 10-19 • 10-8 ‘When there's too much leakage current …’ on page 10-30 • 10-9 ‘Inspection Paperwork’ on page 10-32 • 10-10 ‘Electrical Safety Tests Log’ on page 10-34 10-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 267: Warnings

    Do not pull out or insert circuit boards while power is on. CAUTION Do not operate this Ultrasound system unless all board covers CAUTION and frame panels are securely in place. System performance and cooling require this. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-3 5726264-100 English Rev.8...
  • Page 268: Why Do Maintenance

    Care and Maintenance 10-3 Why do maintenance 10-3-1 Periodic maintenance inspections It has been determined by engineering that your LOGIQ V3/V5/ V5 Expert does not have any high wear components that fail with use, therefore no Periodic Maintenance inspections are mandatory.
  • Page 269: Quality Assurance

    Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Contact GE for coverage and/or price for service.
  • Page 270: Maintenance Task Schedule

    V5/V5 Expert should be serviced and outlines items requiring special attention. NOTE: It is the customer’s responsibility to ensure the LOGIQ V3/V5/V5 Expert care and maintenance is performed as scheduled in order to retain its high level of safety, dependability and performance.
  • Page 271 See Notes also after corrective maintenance NOTE: The maintenance may require specialized equipment to complete. NOTE: The periodic maintenances are not mandatory. The table above is for reference only. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-7 5726264-100 English Rev.8...
  • Page 272: Tools Required

    For a list of required tools for servicing the LOGIQ V3/V5/V5 Expert, refer to chapter 8. 10-5-1 Standard GE tool kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required. Table 10-2:...
  • Page 273 Weller - Soldering Iron Tip KH175 Hemostat 5 inch curved 9-WEWDP12 Weller - Desoldering Pump 9-Z9480121 Alignment tool (red) 93383 Flashlight Mini-Mag Lite (AAA Bat.) 9-GH408 Tweezers 21576 Brush - Bristle LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-9 5726264-100 English Rev.8...
  • Page 274: Tool Kit

    Screwdriver 5/16 inch × 8 inch #36 Drill 9-41586 Screwdriver, Stubby 5/16 inch #29 Drill 9-GH19512 Countersink 1/2 inch 9-44046 3/8 inch Socket Set 9-44741 12 PC Combination Wrench 10-10 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 275: Special Tools, Supplies And Equipment Used For Maintenance

    Quality Assurance Program tests. B/W Printer Cleaning Sheet See printer user manual for requirements Color Printer Cleaning Sheet See printer user manual for requirements Disposable Gloves LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-11 5726264-100 English Rev.8...
  • Page 276: System Maintenance

    • Verify that InSite is functioning properly. • Ensure two-way remote communications. Diagnostics Optional. Presets Backup all Customer Presets to an appropriate media. Image Archive Back up the Image Archive onto appropriate media. 10-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 277: Functional Checks

    Verify the distance by graduate rule. Distance Accuracy should be within ±5%. (Name result from result window Result A, result from graduate rule Result B; Distance Accuracy = (Result B-Result A)/Result A) LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-13 5726264-100 English Rev.8...
  • Page 278 Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no strands may cause a short circuit. Verify Inlet connector retainer is functional. 10-14 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 279: Physical Inspection

    • can be adjusted in the up/down positions. Note: LCD Arm movement may vary and is not applicable to all Ultrasound systems. External I/O Check all connectors for damage. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-15 5726264-100 English Rev.8...
  • Page 280: Optional Diagnostic Checks

    1. Review the system error log for any problems. 2. Check the temperature log to see if there are any trends that could cause problems in the future. 3. Check the Configuration Log; update if needed. 10-16 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 281: Probe Maintenance

    Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
  • Page 282 User Manual. Doing so could result in irreparable damage to the probe. Follow care instructions that came with the probe. Disinfect a defective probe before you return it. Be sure to tag CAUTION the probe as being disinfected. 10-18 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 283: Electrical Safety Tests

    Possible risk of infection. Do not handle soiled or contaminated CAUTION probes and other components that have been in patient contact. Follow appropriate cleaning and disinfecting procedures before handling the equipment. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-19 5726264-100 English Rev.8...
  • Page 284 Check probe leakage current: • Based on your facilities QA program for surface probes. • Based on your facilities QA program for endocavitary probes. • whenever probe damage is suspected. 10-20 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 285: Leakage Current Limits

    Electrical safety tests 10-7-2 Leakage current limits Energy Control and Power Lockout for LOGIQ V3/V5/V5 WARNING Expert. When servicing parts of the Ultrasound system where there is exposure to voltage greater than 30 volts: 1. Follow LOCK OUT/TAG OUT procedures.
  • Page 286 'heart in box' CF symbol. Sink The current resulting from the application of mains Leakage voltage to the applied part. This test is required test for Type CF applied parts. 10-22 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 287: Outlet Test - Wiring Arrangement

    (grounded supply) conductor and the Grounding (protective earth) conductor are reversed. If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-23 5726264-100 English Rev.8...
  • Page 288: Grounding Continuity

    1. GROUND PIN 2. OHMMETER 3. LOGIQ V3/V5/V5 Expert 4. ACCESSIBLE METAL PART: • MONITOR HOUSING • PEAR PANEL CONNECTOR • ANY CASTER/WHEEL SUPPORT Figure 10-2. Ground continuity test 10-24 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 289: Chassis Leakage Current Test

    The maximum allowable limit for chassis source leakage is shown in Table 10-12 on page 10-21. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-25 5726264-100 English Rev.8...
  • Page 290 Optional Optional Power Switch Switch Cover Parts Test 3 Test 4 Enter Name of tested peripheral here: NORM OPEN NORM CLOSED OPEN CLOSED NORM OPEN NORM CLOSED OPEN CLOSED 10-26 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 291: Probe Leakage Current Test

    If no leakage current is detected, check the configuration of the test equipment. 10-7-6-2 Tools For needed tools, see: 10-5 ‘Tools required’ on page 10-8. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-27 5726264-100 English Rev.8...
  • Page 292 Follow the Safety Analyzer tool instruction to test each transducer for leakage current. The electrical Safety Analyzer tool should be calibrated and compliant with AAMI/ESI 1993 or IEC 60601 or AS/NZS 3551. 10-28 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 293 Typical Data Sheet For Transducer Source Leakage Current Transducer Tested: Tester Power Polarity Tester GROUND or Unit Power Switch NUETRAL Switch Measurement NORM OPEN NORM CLOSED OPEN CLOSED NORM OPEN NORM CLOSED OPEN CLOSED LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-29 5726264-100 English Rev.8...
  • Page 294: When There's Too Much Leakage Current

    If later tests indicate high leakage currents, this should be suspected as a possible cause and the outlet wiring should be visually inspected. 10-30 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 295: Probe Fails

    Ultrasound system and if situation can not be corrected, submit a Safety Failure Report to document the Ultrasound system problem. Remove Ultrasound system from operation. 10-8-7 ECG Fails Inspect cables for damage or poor connections. LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-31 5726264-100 English Rev.8...
  • Page 296: Inspection Paperwork

    * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other Figure 10-5. Ultrasound Inspection Certificate * Scan Format: Phased Array, Linear Array, Curved Array, Mechanical Array or Other 10-32 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 297 Comments Probe 1: Probe 2: Probe 3: Final Check. All system covers are in place. System scans with all probes as expected. Accepted by: ______________________________________________________________________ Figure 10-7. Electrical Safety LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-33 5726264-100 English Rev.8...
  • Page 298: Electrical Safety Tests Log

    Patient lead source leakage (lead to ground) Patient lead source leakage (lead to lead) Patient lead source leakage (isolation) Peripheral 1 leakage current Peripheral 1 ground continuity Peripheral 2 leakage current 10-34 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 299 Peripheral 2 ground continuity Peripheral 3 leakage current Peripheral 3 ground continuity Table 10-19: Electrical safety tests (probes) log Max value Comments value measured allowed Probe LOGIQ V3/V5/V5 Expert – Basic Service Manual 10-35 5726264-100 English Rev.8...
  • Page 300 Care and Maintenance 10-36 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 301 1-6 ethernet cooling, 2-5 connect, 3-20 copyrights, i-15 external I/O customer assistance, 1-29 connections, 3-20 phone numbers, 1-30 customer order verification, 3-13 customer provided prerequisite, 8-5 facility needs, 2-12 LOGIQ V3/V5/V5 Expert – Basic Service Manual Index-1 5726264-100 English Rev.8...
  • Page 302 9-3 probes environmental requirements, 2-11 loading software, 8-5 operational and storage temperature for probes, lockout/tagout (LOTO) requirements, 1-25 2-11 LOTO product icons, 1-10 lockout/tagout requirements, 1-25 Index-2 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...
  • Page 303 10-12 system manufacturer, 1-30 system requirements verification, 3-14 system setup, 3-1 time and manpower requirements site preparations, 2-11 tools required, 10-8 special tools, supplies and equipment, 10-11 LOGIQ V3/V5/V5 Expert – Basic Service Manual Index-3 5726264-100 English Rev.8...
  • Page 304 Index-4 LOGIQ V3/V5/V5 Expert – Basic Service Manual 5726264-100 English Rev.8...

This manual is also suitable for:

Logiq v5Logiq v5 expertH48612aaH48612abH48612ac

Table of Contents