hit counter script
GE LOGIQ F Series Basic Service Manual
Hide thumbs Also See for LOGIQ F Series:
Table of Contents

Advertisement

Quick Links

GE
LOGIQ F Series
Basic Service Manual
Part Number: 5446617-100
Revision: 10

Advertisement

Table of Contents
loading

Summary of Contents for GE LOGIQ F Series

  • Page 1 LOGIQ F Series Basic Service Manual Part Number: 5446617-100 Revision: 10...
  • Page 3 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Important Precautions...
  • Page 4 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 5 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 6 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 7 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 8 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 9 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 10 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL viii...
  • Page 11 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 12 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 13 GE personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 14 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL...
  • Page 15 If you are a GE employee or a customer who has entered into such a license agreement with GE to use this proprietary software, you are authorized to use this Material according to the conditions stated in your license agreement.
  • Page 16 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Revision History Revision Date Reason for change 2013/01/10 Initial Release. 2013/06/15 Update for CRU part list update. 2013/10/09 Update for CRU part list update. 2014/05/20 Update for AC power cord update. 2014/10/29 Update for printer list update 2015/09/16...
  • Page 17: Table Of Contents

    Purpose of Operator Manual(s) ........1 - 2 LOGIQ F Series Models Covered by this Manual ....1 - 2 Peripheral List .
  • Page 18 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 2 Site Preparations Overview ............2 - 1 Purpose of chapter 2 .
  • Page 19 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 3 System Setup Overview ............3 - 1 Purpose of Chapter 3 .
  • Page 20 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose for Chapter 4 .
  • Page 21 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
  • Page 22 GE H LOGIQ F S EALTHCARE ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of this chapter 6 .
  • Page 23 PC Diagnostics (Interactive Tests) ....... 7 - 8 Restart LOGIQ F Series After Diagnostics ......7 - 8 Network Configuration .
  • Page 24 GE H LOGIQ F S EALTHCARE ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview ............8 - 1 Purpose of Chapter 8 .
  • Page 25 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
  • Page 26 Standard GE Tool Kit ........
  • Page 27: Overview

    Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ F Series. The service provider must read and understand all the information presented in this manual before installing or servicing a unit. Table 1-1...
  • Page 28: Typical Users Of The Basic Service Manual

    Contractors (Some parts of Chapter 2 - Site Preparation) 1-1-4 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ F Series and also kept near the unit for quick reference. NOTE: Probe information displayed on screen does not necessarily reflect the probes available on your ultrasound system.
  • Page 29: Peripheral List

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-1-6 Peripheral List Table 1-6 LOGIQ F Series Peripheral List Item Part Name Part Number Replaced By Printers 8000 Sony UP-D897 Chinese kit 5151262 8000A Sony UP-D897 USA kit...
  • Page 30 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 1-6 LOGIQ F Series Peripheral List Item Part Name Part Number Replaced By 8014 USB ECG Module (IEC) 5173040 Biopsy Kit 8015 4C-RS biopsy kit 5160703 8016 E8C-RS biopsy kit...
  • Page 31 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-1-6 Peripheral List (cont’d) Table 1-7 LOGIQ F Series System and Application Software List Item Part Name Part Number Replaced By LOGIQ F Series R1.0.0 System and 8022 5485882-S 5485882-2S Application Software USB LOGIQ F Series R1.0.1 System and...
  • Page 32: Important Conventions

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 1-2 Important Conventions 1-2-1 Conventions Used in Book Icons Pictures, or icons, are used wherever they reinforce the printed message. The icons, labels and conventions used on the product and in the service information are described in this chapter. Safety Precaution Messages Various levels of safety precaution messages may be found on the equipment and in the service information.
  • Page 33: Standard Hazard Icons

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 34: Safety Considerations

    1-3-2 Human Safety Servicing should be performed by authorized personnel only. Only personnel who have participated in a LOGIQ F Series Training are authorized to service the equipment. 1-3-3 Mechanical Safety...
  • Page 35: Electrical Safety

    Both the system power cable and the power connector meet international electrical standards. DO NOT SERVICE OR DISASSEMBLE PARTS UNDER FRU UNIT LEVEL AT ANY WARNING CIRCUMSTANCES. 1-3-5 Label Locations Refer to LOGIQ F Series User Guide Section Warning Label Location in Safety Chapter. Section 1-3 - Safety Considerations 1 - 9...
  • Page 36: Dangerous Procedure Warnings

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 1-3-6 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT DANGER IN THIS EQUIPMENT.
  • Page 37: Emc, Emi, And Esd

    1-4-2 CE Compliance The LOGIQ F Series unit conforms to all applicable conducted and radiated emission limits and to immunity from electrostatic discharge, radiated and conducted RF fields, magnetic fields and power line transient requirements.
  • Page 38: Lockout/Tagout (Loto) Requirements

    Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
  • Page 39: Customer Assistance

    Service 1-800-321-7937 Latin America Application Support 1-262-524-5698 Europe (OLC- EMEA) OLC - EMEA GE Ultraschall Deutschland GmbH Phone: +49 (0)212 2802 - 652 Beethovenstraße 239 +33 1 3083 1300 Postfach 11 05 60, D-42655 Solingen Fax: +49 (0) 212 2802 - 431...
  • Page 40: System Manufacturer

    ANUAL 1-6-2 System Manufacturer Table 1-11 System Manufacturer Manufacturer FAX Number GE Medical Systems (China) Co., Ltd. TEL: +86 510-85225888 No.19, Changjiang Road, Wuxi National Hi-Tech Development Zone, FAX: +86 510-85226688 Jiangsu, P.R. China 214028 1-6-3 Factory Site Table 1-12...
  • Page 41: Site Preparations

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 2 Site Preparations Section 2-1 Overview 2-1-1 Purpose of chapter 2 This chapter provides the information required to plan and prepare for the installation of LOGIQ F Series.Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
  • Page 42: General Console Requirements

    Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 43 2-2-2-5 Unit Power Plug If the unit arrives without the power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-6 Power Stability Requirements Voltage drop-out Max 10 ms.
  • Page 44: Emi Limitations

    Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transient in the air wiring. They also generate EMI. The LOGIQ F Series complies with limits as stated on the EMC label. However there is no guarantee that interface will not occur in a particular installation.
  • Page 45: Scan Probe Environmental Requirements

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-2-4 Scan Probe Environmental Requirements Operation: 3° to 40° C Storage: -5° to 50° C NOTE: Temperature in degrees C. Conversion to Degrees F = (Degrees C * (9/5) + 32). NOTICE SYSTEMS AND ELECTRONIC PROBES ARE DESIGNED FOR STORAGE TEMPERATURES OF -10 TO + 60 degrees C.
  • Page 46: Facility Needs

    All electrical work on these product must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 47: Required Features

    2-3-2 Required Features NOTE: GE Medical Systems requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. Sites with a mains power system with defined Neutral and Live: The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size ground wire from the distribution panel to the Ultrasound outlet.
  • Page 48: Recommended And Alternate Ultrasound Room Layout

    EXTERNAL PERIPHERALS DEDICATED POWER OUTLETS FOOT LOGIQ 5 CONSOLE DEDICATED ANALOG TELEPHONE STOOL LINE FOR CONNECTION TO INSITE EXAMINARION GE CABINET TABLE 24 IN. FOT SOFTWARE 76 IN. (61 CM) AND MANUALS (193 CM) DOOR 30 IN. (76 CM) An 8 by 10 foot Minimal Floor Plan...
  • Page 49: Networking Pre-Installation Requirements

    Information must include: • A host name, local port number, AE Title, IP address and Net Mask for the LOGIQ F Series. • The IP addresses for the default gateway and other routers at the site for ROUTING INFORMATION.
  • Page 50 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 2-3-4-4 DICOM Option Pre-installation Requirements (cont’d) LOGIQ F IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses Destination IP Addresses Default ROUTER1 ROUTER2 ROUTER3 DICOM APPLICATION INFORMATION NAME...
  • Page 51: System Setup

    How to prepare the facility and unit of the actual installation, and how to check and test the unit and external peripherals for electrical safety are included in this procedure. Also included LOGIQ F Series in this section are guidelines for transporting the unit to a new site.
  • Page 52: Setup Reminders

    DICOM Option Dependant on the amount of configuration The LOGIQ F Series has been designed to be installed and checked out by an experienced service technician in approximately 1 hour. LOGIQ F Series consoles with optional equipment may take slightly longer.
  • Page 53: Installation Warnings

    3-2-2 Installation Warnings 1.) Since the LOGIQ F Series weighs approximately 65 kg without options, preferably two people should unpack it. Two people are also preferable for installing any additional bulky items. 2.) There are no operator serviceable components. To prevent shock, do not remove any covers or panels.
  • Page 54: Safety Reminders

    CAUTION System performance and cooling require this. OPERATOR MANUAL(S) CAUTION The User Manual(s) should be fully read and understood before operating the LOGIQ F Series and kept near the unit for quick reference. Figure 3-1 Environmental Labels 3 - 4...
  • Page 55: Receiving And Unpacking The Equipment

    CAUTION Do not lift the unit by the Keyboard. Equipment damage may result. CAUTION The crate with the LOGIQ F Series weighs approximately 75kg. Be prepared for a sudden shift of weight as the unit is removed from its base (pallet) Unpacking the the equipment 1.) Tear off the "stop open"...
  • Page 56 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Figure 3-2 Open the box 3 - 6 Section 3-3 - Receiving and Unpacking the Equipment...
  • Page 57: Moving Into Position

    To avoid injury by tipping over. Set the monitor to the lowest position before moving. In general, a single adult can move the LOGIQ F Series along an even surface with no steep grades. At least two people should move the machine when large humps, grooves, or grades will be encountered.
  • Page 58: Preparing For Installation

    System Voltage Settings Verify that Docking Cart is set to the correct voltage. The Voltage settings for the LOGIQ F Series is found on a label to the right of the Power switch and External I/O, on the rear of the system.
  • Page 59: Completing The Installation

    LOGIQ F SERIES 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-5 Completing the Installation 3-5-1 Power On / Boot Up NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
  • Page 60: Transducer Connection

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-5-2-1 Back-end Processor Power Down (cont’d) 2.) The System-Exit window is displayed. Figure 3-5 System Exit Window 3.) Using the Trackball or Select key, select Shutdown. 4.) The shutdown process takes a few seconds and is complete when the power status LED is turned blue.
  • Page 61: System Configuration

    System Configuration 3-6-1 System Specifications 3-6-1-1 Physical Dimensions The physical dimensions of the LOGIQ F Series unit are summarized in Table 3-4 on page 3-11 The Size of LOGIQ F Series. Table 3-4 Physical Dimensions of LOGIQ F Series Height...
  • Page 62: Electrical Specifications

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-1-2 Weight Table 3-5 Weight of LOGIQ F Series With Monitor and Without Other Peripherals Model Weight [kg] Weight [lb] LOGIQ F Series Approximately 65 Approximately 144 3-6-2 Electrical Specifications Table 3-6...
  • Page 63: Peripherals/Accessories Connector Panel

    IRECTION EVISION ASIC ERVICE ANUAL 3-6-5 Peripherals/Accessories Connector Panel LOGIQ F Series peripherals and accessories can be properly connected using the side connector panel. 3-6-5-1 Rear Panel Connector Figure 3-6 Rear Connector Panel Table 3-8 Peripheral/Accessory Connector Panel Isolated USB port...
  • Page 64 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-2 Pin assignment for each connector Table 3-9 Pin Assignments of External VGA Pin No. Signal Pin No. Signal GREEN BLUE Table 3-10 Pin Assignments of USB Pin No. Signal Pin No.
  • Page 65 UP-D711MD Printer can be properly connected using USB Ports. Figure 3-7 Connect B/W printer to the system Sony UP-D711MD Printer is connected to the LOGIQ F Series system via the USB port on the USB hub under the control panel.
  • Page 66 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-3 Connect Peripherals (cont’d) B.) Connect UP-D25MD color printer to the system. UP-D25MD Color Printer can be properly connected using the isolated USB Port. And connect the power cable of UP-D25MD to the wall outlet.
  • Page 67 LOGIQ F SERIES 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-3 Connect Peripherals (cont’d) D.) Connect Foot Switch to the system. Foot Switch can be properly connected using USB Ports. 1-Pedal Footswitch 3-Pedal Footswitch Figure 3-11 Connect Foot Switch to the system E.) Connect the USB Memory to the system.
  • Page 68 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-3 Connect Peripherals (cont’d) F.) ECG can be properly connected using USB ports. Figure 3-13 ECG Connection G.) Connect HP Officejet 100 printer to the system. HP Officejet 100 Printer can be properly connected using the isolated USB Port.
  • Page 69 LOGIQ F SERIES 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL 3-6-5-4 Digital Printer Setup There are two steps to do when setting up a digital printer: 1. follow the procedure below for each printer, then 2. set up specific properties for each printer if you need.
  • Page 70 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Digital Printer 3-6-5-5 Instructions Follow these steps to set up the paper size of the printer, take Sony UP-D897 as an example. 1.) Press Utility-->System-->Peripherals. Select the UP-D897 from the pull-down menu under Standard Printer Properties.
  • Page 71: Available Probes

    EVISION ASIC ERVICE ANUAL 3-6-6 Available Probes See in specification in the LOGIQ F Series User Reference Manual for Probes and intended use. Table 3-14 List of Probes on LOGIQ F Series Probe Name Material of Headshell Area of Using...
  • Page 72: Connectivity Installation Worksheet

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-8 Connectivity Installation Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ SN: Type: REV: CONTACT INFORMATION Name Title Phone E-Mail Address TCP/IP Settings Name - AE Title:...
  • Page 73: Loading Base Image Software

    LOGIQ F SERIES 5273008-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 3-9 Loading Base Image Software Refer to Section 8-5 "Loading Base Image Software" on page 8-7. Chapter 3 System Setup 3 - 23...
  • Page 74: Software Version Check Out

    Section 3-10Software Version check out 3-10-1 Functional Check-out 1.) Power on LOGIQ F Series scanner and wait until system booting to main screen. 2.) Select Utility on control panel or touch panel. 3.) Select About on screen or touch panel.
  • Page 75: Paperwork

    User Manual(s) User Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. Chapter 3 System Setup...
  • Page 76 LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 3 - 26 Section 3-11 - Paperwork...
  • Page 77: Overview

    Section 4-1 Overview 4-1-1 Purpose for Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ F Series console, diagnostics by using the built-in service software, and power supply. Table 4-1 Contents in Chapter 4 Section...
  • Page 78: General Procedure

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 4-3 General Procedure CAUTION SYSTEM REQUIRES ALL COVERS Operate this unit only when all board covers and frame panels are securely in place. The covers are required for safe operation, good system performance and cooling purposes. 4-3-1 Power On/Boot Up After connect the system to the electrical supply, the power is applied to the scanner.
  • Page 79: Power Off/ Shutdown

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-2 Power Off/ Shutdown NOTE: After turning off a system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly. 4-3-2-1 Back-end Processor Power Down To power down the system:...
  • Page 80 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-2-3 Check System Date and Time A warning message “Please check the system date and time are correct” appears on the screen when the system is powered on. This warning message appears for the possible reasons: •...
  • Page 81: Archiving And Loading Presets

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-3 Archiving and Loading Presets NOTE: Always save presets before any software reload. This ensures the presets loaded after the software reload are as up–to–date as possible. All user presets except changes to Summary, Anatomy, and Biometry pages, can be saved on an CD- R disk (or USB memory device) for reloading on the system.
  • Page 82: Adjusting The Display Monitor

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-3-2 Loading Presets from an CD-R disk (or USB memory device) 1.) Insert the CD-R disk with the archived Presets into the DVD-RW. 2.) Access to the Utility Menu, and select System. The Restore sheet will be shown on the LCD display. 3.) Select the item to restore either from Resource Files.
  • Page 83: System Features

    CWD, PDI, 4D, CF and touch panel are not supported on LOGIQ F3. 4-3-5-2 LOGIQ F Series SoftMenu Key Tour 1.) Page Up/Down key: To turn the menu page up and down. 2.) Rotary key: Rotate to adjust the menuMonitor Display...
  • Page 84 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-5-3 Touch Panel (Option) NOTE: Touch panel is an option, and it is not supported on LOGIQ F3. Touch Panel contains exam function and mode/function specific controls. a.) Exam Function Controls Figure 4-7 Exam Function Controls 1.) Patient: Enters Patient screen 2.) Scan: Enters scanning mode screen...
  • Page 85 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-5-3 Touch Panel (Option) (cont’d) b.) Mode/Function Specific Controls In general, the key name is indicated at the top of the key. There are different types of Touch panel keys as illustrated below: Figure 4-8 Mode/Function Specific Controls 1.) Press to toggle control on/off.
  • Page 86 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-5-4 Monitor Display Figure 4-9 Monitor Display Tour 1. Institution/Hospital Name, Date, Time, Operator Identification. 14. Image Preview. 2. Patient Name, Patient Identification. 15. Measurement Summary Window. 3. Power Output Readout. 16.
  • Page 87: B Mode Checks

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-6 B Mode Checks 4-3-6-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-21, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
  • Page 88 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-6-2 B Mode OP Panel Controls Table 4-2 B Mode Controls Description/Benefit Control Possible Bioeffect Depth Depth controls the distance over which the B-Mode images anatomy. To visualize deeper structures, increase the depth.
  • Page 89 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-2 B Mode Controls Description/Benefit Control Possible Bioeffect Width You can widen or narrow the size of the sector angle to maximize the image's region of interest (ROI). Tilt You can steer the sector angle to get more information without moving the probe while in B-Mode, M-Mode,...
  • Page 90 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-2 B Mode Controls Description/Benefit Control Possible Bioeffect SRI-HD SRI-HD (Speckle Reduction Imaging High Definition) is an adaptive algorithm to reduce the unwanted effects of speckle in the ultrasound image. Image speckle usually appears as a grainy texture in otherwise uniform areas of tissue.
  • Page 91: M Mode Controls

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-7 M Mode Controls 4-3-7-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-21, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
  • Page 92 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-7-2 M Mode Controls Table 4-3 M Mode Controls Control Possible Bioeffect Description/Benefit Sweep Changes the speed at which the timeline is swept. Speed Available in M-Mode, Doppler Mode and M Color Flow Mode.
  • Page 93: Doppler Mode Checks

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-8 Doppler Mode Checks 4-3-8-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-21, in Chapter 3 Installation to the System probe connector. 2.) Turn ON the scanner (if it isn’t turned on already).
  • Page 94 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-8-2 Doppler Mode Controls Table 4-4 Doppler Mode Controls Control Possible Bioeffect Description/Benefit Auto Spectral Auto in Doppler Mode optimizes the spectral data. Auto Optimize [ASO] adjusts the Velocity Scale/PRF (on live images only), (Auto) baseline shift, Dynamic Range and invert (if preset).
  • Page 95 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-4 Doppler Mode Controls Control Possible Bioeffect Description/Benefit Display Format Changes the horizontal/vertical layout between B-Mode and M-Mode, or timeline only. Modify Auto Calcs Activates the menu to select which calculations are automatically calculated.
  • Page 96: Color Flow Mode Checks

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-9 Color Flow Mode Checks NOTE: Color Flow Mode only support by LOGIQ C5 Premium 4-3-9-1 Preparations 1.) Connect one of the probes listed in 3-6-6 "Available Probes" on page 3-21, in Chapter 3 Installation to the System probe connector.
  • Page 97 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-9-2 Color Flow Mode Controls Table 4-5 Color Flow Mode Controls Control Possible Bioeffect Description/Benefit Flow Selection In the Lower Extremity Vein (LEV) and Abdominal applications, you can quickly select the flow state via a shortcut on the Color Flow Mode menu.
  • Page 98 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 4-5 Color Flow Mode Controls Control Possible Bioeffect Description/Benefit CF/PDI Auto Set the default value at Utility -> Imaging -> CF Mode. Sample Volume CF/PDI Focus Depth CF/PDI Frequency CF/PDI Auto Frequency CF/PDI Center...
  • Page 99: Basic Measurements

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-10 Basic Measurements NOTE: The following instructions assume that you first scan the patient and then press Freeze. 4-3-10-1 Distance and Tissue Depth Measurements 1.) Press Measure once, an active caliper displays. 2.) To position the active caliper at the start point (distance) or the most anterior point (tissue depth), move the Trackball.
  • Page 100: Probe/Connectors Usage

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-10-3 Worksheets Measurement/Calculation worksheets are available to display and edit measurements and calculations. There are generic worksheets as well as Application specific worksheets. The worksheets are selected from the Measurement Touch Panel. 4-3-10-4 Report Pages Measurements/Calculations that are included on the worksheet can also be displayed on Report Pages.
  • Page 101: Using Cine

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-12 Using Cine 4-3-12-1 Activating CINE Press Freeze, then roll the Trackball to activate CINE. To start CINE Loop playback, press Run/Stop. To stop CINE Loop playback. press Run/Stop. 4-3-12-2 Quickly Move to Start/End Frame Press First to move to the first CINE frame;...
  • Page 102: Backup And Restore Database, Preset Configurations And Images

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13 Backup and Restore Database, Preset Configurations and Images 4-3-13-1 Formatting Media 1.) To format the backup media, enter Utility-> Connectivity->Removable Media. 2.) Select the media type from the drop down menu. 3.) Enter the label for the media as shown in Figure 4-18.
  • Page 103 2.) Enter enter Utility-> System-> Backup/Restore. NOTE: If you are not logged in as GE Service or with administrator privileges, the Operator Login window is displayed. Log on with administrator privileges. 3.) In the Backup list, select Patient Archive, Report Archive, User Defined Configuration and Service.
  • Page 104 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13-3 Restore System Presets and Configurations CAUTION The restore procedure overwrites the existing database on the local hard drive. Make sure to insert the correct CD (or USB memory device). 1.) Insert the Backup/Restore CD/DVD (or USB memory device) into the drive.
  • Page 105 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13-4 Archiving Images 1.) Insert the archive media. 2.) To format the archive media, enter Utility-> Connectivity-> Removable Media. 3.) Format the CD. Verify the format if desired. 4.) Images will be moved from the hard drive by date. Therefore, the best way is to label media by date.
  • Page 106 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13-4 Archiving Images (cont’d) 6.) Go to Archive menu and select EZBackup/EZMove, the EZBackup/EZMove wizard starts. Figure 4-24 Archive Screen 7.) Verify the information on the first page of EZBack/EZMove wizard, then select Next. If you want to backup all of the exams in the range (even if the exam was previously backed up), check this option.
  • Page 107 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13-4 Archiving Images (cont’d) 8.) It indicates the size of the data and the storage. Select Next to continue. NOTE: The calculation for the number of backup CD is only an estimate. Allow for one additional CD when performing an EZBack/EZMove.
  • Page 108 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 4-3-13-4 Archiving Images (cont’d) 9.) The status menu appears. When the backup/move has been complete, press Next. Figure 4-28 EZBackup Wizard 3 NOTE: If you need to insert the next media, a message appears providing you with the media label. Label the media, then insert the next media and press OK.
  • Page 109: Software Configuration Checks

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 4-4 Software Configuration Checks Table 4-6 Software Configuration Checks Step Task to do Expected Result(s) Check Date and Time setting Date and Time are correct Check that Location (Hospital Name) is correct Location Name is correct Check Language settings Desired Language is displayed...
  • Page 110 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 4 - 34 Section 4-5 - Peripheral Checks...
  • Page 111: Components And Functions (Theory)

    ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ F SERIES' system concepts, component arrangement, and subsystem function. Table 5-1 Contents in Chapter 5 Section Description Page Number Overview...
  • Page 112: System Diagram

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 5-2Block Diagram 5-2-1 System Diagram Figure 5-30 LOGIQ F Series System Diagram 5 - 2 Chapter 5 Components and Functions (Theory)
  • Page 113: Software Diagram

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 5-2-2 Software Diagram Figure 5-31 LOGIQ F Series Software Diagram Chapter 5 Components and Functions (Theory) 5 - 3...
  • Page 114: Common Service Platform

    LOGIQ F SERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 5-3 Common Service Platform 5-3-1 Introduction The Service Platform contains a set of software modules that are common to all PC backend ultrasound and cardiology systems. The Common Service Platform will increase service productivity and reduce training and service costs.
  • Page 115: Service Adjustments

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of this chapter 6 This section describes how to test and adjust the scanner. These tests are optional. You may use them to check the system for errors.
  • Page 116: Monitor Adjustments

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 6-2 Monitor Adjustments 6-2-1 Adjustments Procedures To adjust the brightness: For LOGIQ C3/LC5 Premium: Adjust the LCD monitor’s button, located the right side of the LCD Monitor. For LOGIQ C3/LC5 Premium: Adjust the LCD monitor’s button, located the bottom of the LCD Monitor. Figure 6-32 LCD Monitor To adjust the light:...
  • Page 117: Diagnostics/Troubleshooting

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1 Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
  • Page 118: Gathering Trouble Data

    Collect Vital System Information The following information is necessary in order to properly analyze data or images being reported as a malfunction or being returned to the manufacturer: Product Name = LOGIQ F Series From the Utility>System>General>About screen: Applications Software...
  • Page 119: Collect A Trouble Image With Logs

    Power Supply Temperature NOTE: Power Supply and Temperature logs are not currently being updated by the LOGIQ F Series. This Alt-D function is available at all times. Figure 7-1 ALT-D Dialog Box When Alt-D is pressed, a menu box appears that allows for:...
  • Page 120: Screen Captures

    GE H LOGIQ F S EALTHCARE ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-3 Screen Captures There may be times when the customer or field engineer will want to capture a presentation on the screen. This is accomplished by first saving the image(s) to the clipboard using a Print Key.
  • Page 121 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-3-3 Capturing a Screen (cont’d) Figure 7-3 Select Image to Capture 3.) Click FREEZE to unfreeze the image to view the image screen and the snapshots displayed on the bottom. 4.) Highlight the snapshot to be stored to RD (Removable Disk) or CD-R.
  • Page 122: Reset The Store Key To Customer's Functionality

    GE H LOGIQ F S EALTHCARE ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 7-3-3 Capturing a Screen (cont’d) Figure 7-4 Save Dialog Box 6.) A Save dialog box will be opened. Choose d:\export folder as the archive location to save the image on the hard disk or CD-R.
  • Page 123: Common Diagnostics

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-4 Common Diagnostics 7-4-1 Utilities Provides two selections: 7-4-1-1 Disruptive Mode Allows you to enable or disable disruptive mode troubleshooting. 7-4-1-2 System Shutdown Allows for system shutdown from the diagnostic menu. Select to Restart System or Shutdown System. Also, select to retain Disruptive Mode or Not.
  • Page 124: Pc Diagnostics (Non-Interactive Tests)

    PC Diagnostics (Interactive Tests) 7-4-3-1 AVI Playback 7-4-3-2 Keyboard 7-4-3-3 Microphone 7-4-3-4 Monitor 7-4-3-5 Sound 7-4-3-6 USB Ports 7-4-4 Restart LOGIQ F Series After Diagnostics Always shutdown the system and reboot after a diagnostics session. 7 - 8 Section 7-4 - Common Diagnostics...
  • Page 125: Network Configuration

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 7-5 Network Configuration 7-5-1 Network Configuration 1.) Connect system with network. 2.) Enter Utility-> Connectivity-> TCP/IP, in IP settings window, check Enable DHCP, and select the proper network speed in Network Speed. Figure 7-5 Enable DHCP NOTE: If user wants to setup static IP address, uncheck Enable DHCP option, input static address in...
  • Page 126 GE H LOGIQ F S EALTHCARE ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Connect system with network. (cont’d) 3.) Select Save, and a popup window displays. Select OK to restart the system and activate the changes. Figure 7-7 System Restart inquiry dialog 4.) After the system restarts, the network icon at the left bottom of screen displays as connected.
  • Page 127: Replacement Procedures

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter describes replacement procedures for the following modules and subsystems. Table 8-1 Contents in Chapter 8 Section Description Page Number...
  • Page 128: Disassembly/Re-Assembly

    If the LOGIQ F Series needs to be sent for repair, ensure that any patient information is erased from the Harddisk/Storage Device. In case that any patient information is still residing on the LOGIQ F Series, GE will contact the customer and request for urgent collection of that patient information.
  • Page 129: Air Filter

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 8-2-3 Air filter 8-2-3-1 Tools • Common phillips screwdrivers 8-2-3-2 Preparations • Shut Down the System and disconnect the power cord. 8-2-3-3 Removal Procedure 1.) Pull out the air filter from the rear panel. 2.) Dust the filter with a vacuum cleaner and/or wash it with a mild soapy solution.
  • Page 130: Trackball Roller Cleaning

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-3 Trackball Roller Cleaning Purpose: This is description on how to remove the trackball and clean the roller. 8-3-0-1 Tools • No need. 8-3-0-2 Needed Manpower • 1 person, 3 minutes + travel 8-3-0-3 Preparations •...
  • Page 131: Attaching Keyboard Film

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-4 Attaching Keyboard Film 8-4-0-1 Tools • 8-4-0-2 Preparations • 8-4-0-3 Time Needed • 1 person, 1minutes+travel 8-4-0-4 Installation Procedure Attach the keyboard film to the keyboard. 1.) Use a cotton swab to clean around the keys. NOTE: Be careful and keep the keyboard film clean when you take it out.
  • Page 132 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-4 Attaching Keyboard Film (cont’d) Fit the film on the keyboard well and then remove the remaining release paper. Figure 8-13 Fit the keyboard film 3.) The keyboard film is attached successfully. Figure 8-14 Attached Successfully 8 - 6 Section 8-4 - Attaching Keyboard Film...
  • Page 133 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5Loading Base Image Software 8-5-0-1 Loading Base Image Software for Software version R1.0.8 and previous NOTE: While it is believed to be unnecessary, It would not hurt to disconnect the system from the network and remove all transducers.
  • Page 134: Loading Base Image Software

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) 6.) Select one of the options for loading the system. Select choice 1] to load the complete disk. Figure 8-16 Selection for loading the system While the software install procedure is designed to preserve data, you should select WARNING choice [2] to format disk C only.
  • Page 135 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) 8.) System will be loaded as shown in the screen below. Figure 8-18 System Load 9.) System updating finished, refer to Figure 8-19 on page 8-9 Figure 8-19 System upgrade complete 10.)Remove the USB stick, then press power key to restart the system.
  • Page 136 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) 8-5-0-2 Loading Base Image Software for Software version R1.0.9 and above 1.) Insert the USB disk labeled “System & Application Software” in the USB port of the system. 2.) Properly turn off the system by momentarily pressing the Power On/Off Switch.
  • Page 137 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) 6.) Press any key to continue the upgrade process, refer to Figure 8-22 on page 8-11 Or it is available to remove the USB moemory stick and shut down the system by pressing the Power On/Off key.
  • Page 138 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) While the software installation procedure is designed to preserve data, select choice WARNING [2] to format disk C only. If selecting [1], please make sure the patient data and images have been backed up WARNING first, then upgrade the system.
  • Page 139 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 8-5 Loading Base Image Software (cont’d) 10.)The upgrade process is complete, the system will shut down automatically. Refer to Figure 8-26 on page 8-13 Figure 8-26 System upgrade complete Section 8-5 - Loading Base Image Software 8 - 13...
  • Page 140: Direction 5446617-100, Revision 10 Basic Service Manual

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 8 - 14 Section 8-5 - Loading Base Image Software...
  • Page 141: Renewal Parts

    ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Renewal Parts for LOGIQ F Series. Table 9-1 Contents in Chapter 9 Section Description Page Number Overview List of Abbreviations...
  • Page 142: Renewal Parts Lists

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-3Renewal Parts Lists 9-3-1 Equipment Models Covered in this Chapter Table 9-2 Renewal Parts List for Customer Part Name Part Number Description Quantity AC Power Cord 5177123-2 Power cord Europe Class AC Power Cord 5176304-2 Power cord China Class...
  • Page 143 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-4Operator Console Assy Figure 9-27 Operator Console Assy Chapter 9 Renewal Parts 9 - 3...
  • Page 144 Part Number Description 6000 5446617-100 LOGIQ F Series Service Manual System User Manuals 6001 5454368-100 LOGIQ F Series Basic User Manual, English 6002 5453550-100 LOGIQ F Series User Guide, English 6003 5453550-101 LOGIQ F Series User Guide, French 6004 5453550-106...
  • Page 145: Direction 5446617-100, Revision 10 Basic Service Manual

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Item Part Number Description 6033 5454731-141 LOGIQ F3 User Guide, Chinese 6034 5457182-141 LOGIQ F5/F6 User Guide, Chinese 6035 5444538-141 LOGIQ F8 User Guide, Chinese Chapter 9 Renewal Parts 9 - 5...
  • Page 146 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 9-6Probes Table 9-3 Probes on LOGIQ F Series Replaced Item Part Name Part Number Description Quantity 7000 4C-RS 5451471 Probe (Center Frequency: 3.10MHz) 5488477 7000A 4C-RS 5488477 Probe (Center Frequency: 3.10MHz)
  • Page 147: Care & Maintenance

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 148: Why Do Maintenance

    Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
  • Page 149 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 10-3 Maintenance Task Schedule (cont’d) Table 10-2 Customer Care Schedule Per Facilities Service at Indicated Time Daily Weekly Monthly QA Program Notes • Clean Probe Holders • more frequently depending on Clean Air Filter your environment •...
  • Page 150: Tools Required

    Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs. Table 10-3 Overview of GE-1 Tool Kit Contents Tool ID...
  • Page 151 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) Tool ID Description Tool ID Description 9-65283 Case 8.5 in. x 4.5 in. x 2 in. Deep 9-45072 Pliers 6 inch Diagonal...
  • Page 152: Special Tools, Supplies And Equipment

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Care & Maintenance Table 10-5 Overview of Requirements for Care & Maintenance Tool Comments Digital Volt Meter (DVM) Leakage Current Ultrasound Kit For 120V and 220V Units Kit includes anti–static mat, wrist strap and cables for 200 to 240 V system 3M #2204 Large adjustable wrist strap...
  • Page 153: Input Power

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-3 Input Power 10-4-3-1 Mains Cable Inspection Table 10-6 Mains Cable Inspection Step Item Description Unplug Cord Disconnect the mains cable from the wall and system. Inspect Inspect it and its connectors for damage of any kind. Verify that the LINE, NEUTRAL and GROUND wires are properly attached to the terminals, and that no Verify strands may cause a short circuit.
  • Page 154: Outlet Test -Wiring Arrangement - Usa & Canada

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-6 Outlet Test -Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangment by plugging in the neon outlet tester and noting the combination of lights that are illuminated. Any problems found should be reported to the hospital immediately and the receptacle should not be used.
  • Page 155: Grounding Continuity

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-7 Grounding Continuity CAUTION Electric Shock Hazard. The patient must not be contacted to the equipment during this test. Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case. The ground wire resistance should be less than 0.2 ohms.
  • Page 156: Chassis Leakage Current Test

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-8 Chassis Leakage Current Test 10-4-8-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
  • Page 157: Isolated Patient Lead (Source) Leakage-Lead To Lead

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-9 Isolated Patient Lead (Source) Leakage–Lead to Lead Reference the procedure in the IEC 60601-1. Select and test each of the five ECG lead positions (except ALL) on the LEAD selector, testing each to the power condition combinations found in the table. Record the highest leakage current measured.
  • Page 158 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-10-1 Data Sheet for ECG Leakage Current The test passes when all readings measure less than the value shown in the table below. Record all data on the PM Inspection Certificate. Table 10-9 Maximum Allowance Limit for ECG Leakage Current Maximum Allowance Limit...
  • Page 159: Probe Leakage Current Test

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-11 Probe Leakage Current Test 10-4-11-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-4-11-2 Generic Procedure Measurements should be made with the ground open and closed, with power line polarity normal and...
  • Page 160 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL 10-4-11-5 Data Sheet for Transducer Source Leakage Current The test passes when all readings measure less than the values. Record all data on the PM Inspection Certificate. CAUTION Equipment damage possibility. Never switch the Polarity and the status of Neutral when the unit is powered ON.
  • Page 161: When There's Too Much Leakage Current

    LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Section 10-5 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part. Tighten all grounds.
  • Page 162 Air Filter Applicable Software Options Probe Holders Applicable Hardware Options External I/O Control Panel Wheels, Brakes & Swivel Locks Monitor Cables and Connectors GE Approved Peripherals (VCR, CD-RW, MOD, Touch Panel Printers) Measurement Accuracy GE Approved Peripherals COMMENTS: ______________________________________________________________________________ ______________________________________________________________________________ ______________________________________________________________________________...
  • Page 163 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Chassis Source Leakage Current - Caster Chassis Source Leakage Current - CRT Patient Lead Source Leakage (Lead to Ground)
  • Page 164 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL This page was intentionally left blank. 10 - 18 Section 10-5 - When There's Too Much Leakage Current...
  • Page 165 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL INDEX Abbreviations, 9-1 Hazard Icons, 1-7 Archiving Images Human Safety, 1-8 Move, 4-29 Backup Patient Database, 4-26 Preset Configurations, 4-26 LOTO, 1-12 Basic Measurements Functional Checks, 4-23 Boot Up, 3-9 Mechanical Safety, 1-8 Models Covered, 1-2 Monitor Display...
  • Page 166 LOGIQ F S ERIES 5446617-100, R IRECTION EVISION ASIC ERVICE ANUAL Touch Panel Functional Checks, 4-7 Trouble Image with Logs, 7-3 Troubleshooting Gathering Trouble Data, 7-2 Screen Captures, 7-4 Trouble Image with Logs, 7-3 Vital System Information, 7-2 Ultrasound Room Layout, 2-8 Voltage Settings, 3-8 Warnings and Cautions, 1-8 Index...
  • Page 167 2013-2017, General Electric Company. © GE-GEMedical Systems Ultrasound. 9900 Innovation Drive Wawautosa, Wisconsin 53226...

Table of Contents

Save PDF