Patient Safety - GE LOGIQ 7 Basic User Manual

Hide thumbs Also See for LOGIQ 7:
Table of Contents

Advertisement

Patient Safety

Related Hazards
WARNING
Patient
identification
Diagnostic
information
CAUTION
LOGIQ 7 Basic User Manual
Direction 2392206-100 Rev. 1
The concerns listed can seriously affect the safety of patients
undergoing a diagnostic ultrasound examination.
Always include proper identification with all patient data and
verify the accuracy of the patient's name and ID numbers when
entering such data. Make sure correct patient ID is provided on
all recorded data and hard copy prints. Identification errors could
result in an incorrect diagnosis.
Equipment malfunction or incorrect settings can result in
measurement errors or failure to detect details within the image.
The equipment user must become thoroughly familiar with the
equipment operation in order to optimize its performance and
recognize possible malfunctions. Applications training is
available through the local GE representative. Added
confidence in the equipment operation can be gained by
establishing a quality assurance program.
Allowing the machine to transmit acoustic output with the probe
not in use (or in its holder) can cause the transducer to build up
heat. Always turn off acoustic output or freeze the image when
not in use. The system's acoustic output remains transmitting
when the user controls are being used.
Safety Precautions