Page 2
GE H EALTHCARE 5305880-100 , R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Your Notes: Page i-ii Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 3 o f 1 9 7...
Page 3
• WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLÄGE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN. Page i-i Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 4 o f 1 9 7...
Page 4
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL • ESTE MANUAL DE SERVICIO SÓLO EXISTE EN INGLÉS. AVISO • SI ALGÚN PROVEEDOR DE SERVICIOS AJENO A GEMS SOLICITA UN IDIOMA QUE NO SEA EL INGLÉS, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIÓN.
Page 5
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page i-iii Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 6 o f 1 9 7...
Page 6
GE Medical Systems personnel. In performing all electrical work on these products, GE will use its own specially trained filed engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
Page 7
The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Medical Systems. GE Medical Systems may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
Page 8
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page i-vi Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 o f 1 9 7...
Page 9
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change Preliminary Release. JAN 26 2008 Changes incorporated for correcting the SPR’s raised against DOC0421436. APRIL 24 2008 Changes incorporated on contact details and added additional information based on service August 4 2008 teams requirements.
Page 10
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page i-vi Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 1 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Table Of Contents Chapter Table Of Contents Chapter 1 Introduction Section 1-1Overview ..................1-1 1-1-1 Overview ........................1-1 1-1-2 Purpose of Chapter 1..................... 1-1 1-1-3 Purpose of Service Manual.................... 1-1 1-1-4 Typical Users of the Basic Service Manual ...............
Page 12
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Lighting 2 2-2-2 Electrical Requirements ....................2-2 2-2-2-1 LOGIQ™ 100 PRO Power Requirements ............2-2 2-2-2-2 Site Circuit Breaker ................... 2-2 2-2-2-3 Site Power Outlets..................... 2-2 2-2-2-4 Unit Power Plug....................2-2 2-2-2-5 Power Stability Requirements ................
Page 13
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 3-7Paperwork................... 3-28 3-7-1 User Manual(s) ......................3-29 3-7-2 Product Locator Card....................3-29 Chapter 4 ..................Functional Checks Section 4-1 ..................Overview 4-1 4-1-1 Purpose of Chapter 4..................... 4-1 Section 4-2 General Procedures ..............
Page 14
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1Overview..................7-1 7-1-1 Purpose of Chapter 7 ..................... 7-1 Section 7-2 Power On Diagnostics ..............7-3 7-2-1 DAVINCI Board Tests 7-4 7-2-2 FEB Test for LOGIQ™ 100 PRO .................. 7-7...
Page 15
Section 10-3 Periodic Maintenance Schedule ..........10-2 10-3-1 How often should PMs be performed? ................ 10-2 Section 10-4 Tools Required ............... 10-4 10-4-1 Standard GE Tool Kit....................10-4 10-4-2 Special Tools, Supplies and Equipment ..............10-6 10-4-2-1Specific Requirements for Periodic Maintenance .......... 10-6 Section 10-5 System Periodic Maintenance ..........
Page 16
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-6 Probe Leakage Current Test 10-19 10-6-6-1Definition....................... 10-19 10-6-6-2Generic Procedure for Leakage current ............10-19 10-6-6-3Meter Procedure Using Dale Meter to measure leakage current ....10-19 Meter Procedure using Probe Adapter to measure Probe isolation(Sink) Current 21 Section 10-7 When There's Too Much Leakage Current......
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1Overview 1-1-1 Overview The LOGIQ™ 100 PRO is a compact portable ultrasound scanner which is designed for OB/GYN, Abdomen, Urology and small part Scans using the convex, linear and micro-convex (Transvaginal) probes.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9.) Chapter 9, Replacement Parts: Contains a complete list of replacement parts for the LOGIQ™ 100 PRO. 10.)Chapter 10, Quality Assurance: Provides periodic maintenance procedures for the LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 1-2 Safety 1-2-1 Warnings CAREFULLY READ ALL THE WARNINGS BELOW WARNING 1.) The operator manual should be read and understood before operating the LOGIQ™ 100 PRO and kept nearby for quick reference.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 1-3 Important Conventions 1-3-1 Conventions Used in Book Model Designations. This manual covers the LOGIQ™ 100 PRO scanners. (See “LOGIQ™ 100 PRO Models Covered by this Manual” on page 1-2.) Icons.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment. LABEL/SYMBOL PURPOSE/MEANING LOCATION Identification and Rating Plate...
Page 23
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL LABEL/SYMBOL PURPOSE/MEANING LOCATION "CAUTION" The equilateral Various triangle is usually used in combination with other symbols to advise or warn the user. ATTENTION - Consult Various accompanying documents " is...
Page 24
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL LABEL/SYMBOL PURPOSE/MEANING LOCATION "Equipotentiality" Indicates Rear of console the terminal to be used for connecting equipotential conductors when interconnecting (grounding) with other equipment. "Non-Ionizing Radiation" Rear of console near power...
Page 25
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL LABEL/SYMBOL PURPOSE/MEANING LOCATION No hazardous substance, Rear Panel. above the maximum concentration value, are present. Maximum concentration values for electronic information products, as set by the People’s Republic of China...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 1-4Safety Considerations 1-4-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment. Failure to comply with these precautions or with specific warnings elsewhere in this manual, violates safety standards of design, manufacture and intended use of the equipment.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-4-3 Mechanical Safety ULTRASOUND PROBES ARE HIGHLY SENSITIVE MEDICAL INSTRUMENTS THAT CAN EASILY BE DAMAGED BY IMPROPER HANDLING. USE CARE WHEN HANDLING AND PROTECT FROM DAMAGE WHEN NOT IN USE. DO NOT USE A DAMAGED OR WARNING DEFECTIVE PROBE.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-4-5 Labels Locations GHOST Label Figure 1-1 Warning Lables Page 1-12 Section 1-1 - Overview Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 2 9 o f 1 9 7...
Page 29
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-2 Warning Lables Chapter 1 - Introduction Page 1-13 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 3 0 o f 1 9 7...
Page 30
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-3 Warning Labels Page 1-14 Section 1-1 - Overview Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 3 1 o f 1 9 7...
Page 31
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-4 Warning Labels Chapter 1 - Introduction Page 1-15 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 3 2 o f 1 9 7...
Page 32
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) CUIDADO ATTENZIONE CAUTION Não utilize os dispositivos seguintes per- Non usare i seguenti apparecchi in pros- to deste equipamento: simità di questa apparecchiatura:...
Page 33
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-3-5 Labels Locations (cont’d) Figure 1-6 Warning Lables Chapter 1 - Introduction Page 1-17 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 3 4 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-4-6 Dangerous Procedure Warnings Warnings, such as the example below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed. DANGEROUS VOLTAGES, CAPABLE OF CAUSING DEATH, ARE PRESENT IN THIS EQUIPMENT.
3.) Keep the equipment as far as possible from other electronic equipment. 4.) Be sure to use the only cables provided by GEMS or ones approved by Wipro GE. Install the unit/peripherals as per the installation procedure provided in Chapter 3 - INSTALLATION. The layout of the LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1-5-5 General Notice 1.) Designation of Peripheral Equipment Connectable to this Product. The peripheral which conforms to EN60601-1-2 (IEC601-1-2), can be connected up to the LOGIQ™ 100 PRO without compromising its EMC performance.
Operator's Manual, please contact your service support center. If the service engineer needs additional information to repair this equip ment, please contact any relevant addresses given below: 1-6-1 System Manufacturer WIPRO GE HEALTHCARE - INDIA WIPRO GE HEALTHCARE PVT., LTD. ULTRASOUND GROUP. PLOT # 4, KADUGODI INDUSTRIAL AREA,...
4855 W. Electric Avenue TEL: (1) 262-524-5698 Milwaukee, WI 53219 FAX: +1- 414-647-4125 Customer Answer Center ASIA(Singapore/Japan): TEL: 65-291 8528 GE Ultrasound Asia (Singapore) (81) 426-48-2950 Service Department - Ultrasound 298 Tiong Bahru Road #15-01/06 FAX: 65-272-3997 Central Plaza FAX: (81) 426-482902...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 2 Pre Installation Section 2-1Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ 100 PRO. Included are descriptions of the facility and electrical needs to be met by the purchaser of the unit.
2-2-2-4 Unit Power Plug If the unit arrives without a power plug, or with the wrong plug, you must contact your GE dealer or the installation engineer must supply what is locally required. 2-2-2-5 Power Stability Requirements Voltage drop-out Max 10 ms.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 2-2-3 EMI Limitations Ultrasound machines are susceptible to Electromagnetic Interference (EMI) from radio frequencies, magnetic fields, and transients in the air or wiring. They also generate EMI. The LOGIQ™ 100 PRO complies with limits as stated on the EMC label.
RF leakage. Or, if a label has been found in such a touch metal position, move the label. Use GE specified The interconnect cables are grounded and require ferrite beads and other harnesses and shielding. Also, cable length, material, and routing are all important; do peripherals not change from what is specified.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 2-2-5 Time and Manpower Requirements Site preparation takes time. Begin Pre installation checks as soon as possible, if possible, six weeks before delivery, to allow enough time to make any changes.
Page 44
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Your Notes: Page 2-6 Section 2-2 - General Console Requirements Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 4 5 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 3 Installation Section 3-1Overview 3-1-1 Purpose of Chapter 3 This chapter contains information needed to install the unit. Included are references to a procedure that describes how to receive and unpack the equipment and how to file a damage or loss claim.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-1-3 Safety Reminders WHEN USING ANY TEST INSTRUMENT THAT IS CAPABLE OF OPENING THE AC GROUND LINE (I.E., METER’S GROUND SWITCH IS OPEN), DON’T TOUCH DANGER THE UNIT! If the unit is very cold or hot, do not turn on its power until it has had a chance to acclimate to its operating environment.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 3-2 Receiving and Unpacking the Equipment Please read that procedure before packing/unpacking the LOGIQ™ 100 PRO. We strongly advice you to store the LOGIQ™ 100 PRO packing material in undamaged condition in case of future transportation.
Page 48
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Packaging Buffe Carton Flaps Packaging Buffers Stretch Film Figure 3-2 Unpacking LOGIQ™ 100 PRO Note: The Service Manual, Operator Manual,Advanced Refernce Manual, Gel Bottle and Power Cord are wrapped in bubble sheet and are kept along the sides of the corrugated carton box.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 3-3 Preparing for Installation 3-3-1 Physical Inspection 3-3-1-1 Systems Voltage Settings • Verify that the scanner is set to the correct voltage. The Voltage settings for the LOGIQ™ 100 PRO Scanner is found on a label on the rear of the system.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Assembling LOGIQ™ 100 PRO (Contd.,) Figure 3-4 Assembling LOGIQ™ 100 PRO Section 3-5 Completing the Installation 3-5-1 Transducer Connection 1.) Connect a transducer to transducer receptacle as follows,Refer to fig 3-4 A.) Ensure that the transducer twist lock lever points towards the top of the probe...
Video Out on the rear panel of LOGIQ™ 100 PRO to the Video IN of the VGP. Connect the shutter cable from the VGP printer to the rear panel of the Logiq 100 PRO. 4.) The images and report pages of the Logiq 100 PRO can be printed on HP Laser jet printer(Model HP1020) through the USB port provided on the rear panel.Images can be...
Page 52
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Figure 3-5 Installation of Accessories Page 3-8 Section 3-5 - Completing the Installation Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 5 3 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of Two Probe Adapter & LOGIQ α200 Probe Adapter (Optional) 3-5-3 1.) Remove the adhesive tape to open the flaps of two probe port or LOGIQ α 200 Probe Adapter 2.) Remove the PU Foams inside the two probe port or LOGIQ α...
Page 54
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of Two Probe Adapter & LOGIQ α200 Probe Adapter (Optional) (contd.,) 4.) Ensure that no probe is connected to the system. If connected remove it first.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-5-4 Installation of LOGIQ™ 100 PRO with Global or Modular Trolley Note: LOGIQ™ 100 PRO has two types of Trolleys, the Global Trolley and the Modular Trolley.
Page 56
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,) 2.) Unpack the Modular Trolley packed in a box by opening the top cover, refer Figure 3-8...
Page 57
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,) Figure 3-10 Modular Trolley Assembly Chapter 3 - Installation Page 3-13 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 5 8 o f 1 9 7...
Page 58
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,) 3.) Remove two M4X10 screws from the base ofLOGIQ™ 100 PRO and fix lock plate. Refer Figure 3-11 4.) Place the unit with lock plate fixed on Trolley, ensure that lock plate gets into the projection on...
Page 59
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,) 5.) Take the belt (provided in the kit) over the unit and lock it. Refer Figure 3-12 6.) Release the Keyboard by pressing the lock release and make it rest on the handle.
Page 60
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Installation of LOGIQ™ 100 PRO with Global or Modular Trolley (Contd.,) Figure 3-13 Installation of LOGIQ™ 100 PRO with Global or Modular Trolley Page 3-16 Section 3-5 - Completing the Installation...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-5-5 Adjustment of Monitor Brightness & Contrast Turn ON the Machine and then adjust the monitor contrast and brightness. Monitor contrast and brightness should be adjusted according to the lighting in the room. For Monitor brightness and contrast adjustment refer to the appropriate section in the User Manual.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-6-1-2 Front View Figure 3-14 Front View Page 3-18 Section 3-6 - System Configuration Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 6 3 o f 1 9 7...
Page 63
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-6-1-3 Side View (Keyboard Closed) 420.6 (16.5) 405 (15 9) Figure 3-15 Side View (Keyboard Closed) 3-6-1-4 Side View (Keyboard Open) 596 (23.5) Figure 3-16 Side View (Keyboard Open)
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL SIDE VIEW 3-6-1-5 Weight without Peripherals The Weight of LOGIQ™ 100 PRO without peripherals Model Weight [kg] LOGIQ™ 100 PRO 9.00 Table 3-4 Weight of LOGIQ™ 100 PRO...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-6-2-1 Power Supply The Following Power Line requirement should be monitored one week before Installation. We recommend that a Drantez model 605-3 Power line Monitor with options 101 to be used...
Page 66
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Printing should not be done while live scanning patients. Also, the Printer cable should not be attached to the LOGIQ™ 100 PRO during the scan. For Europe Only: All devices connected to the LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-6-4 Peripherals Connections and Installation 3.6.4.1 FootSwitch Connection The footswitch which is the remote FREEZE device is connected to the rear panel of the system. this extra FREEZE switch is provided to enhance flexibility to freeze images when the system is not within reach of the user.
Page 68
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3.6.4.2 Video Graphic Printer (VGP) Connect the Video Graphic printer (with USB A to USB B cable) to the USB port located on the rear panel of the system.
Page 69
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Figure 3-22 VCR connection to Rear panel Installation of VCR: 1.) Connect the SONY VCR 2.) Switch on the VCR and the LOGIQ™ 100 PRO unit. 3.) Press the Record key to record the image.
Page 70
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3.6.4.6 USB MEMORY STICK This module serves as an interface which enables users to transfer images from LOGIQ™ 100 PRO to USB memory stick. LOGIQ™ 100 PRO in JPEG format.This can be further uploaded to any PC having windows operating system and USB port support.
Page 71
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3.6.4.7 Available probes for LOGIQ™ 100 PRO See in Specifications in the LOGIQ™ 100 PRO User manual for probes and intended use. Probe Freq in Part No.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-6-5 Video Specification Video specifications may be needed to be able to connect laser cameras to the LOGIQ™ 100 PRO. General Parameter NTSC 525[Frame] 625[Frame] Total Number of Horizontal Lines...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 3-7-1 User Manual(s) Check that the correct User Manual(s), per software (sw) revision, for the system is included. 3-7-2 Product Locator Card Fill out proper customer information on the Product Locator Installation Card. Refer to Illustration below.
Page 74
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 3-30 Section 3-7 - Paperwork Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 7 5 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ 100 PRO scanner, diagnostics by using the built-in service software, and power supply adjustments. For remaining functional checks please refer to the LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 4-2 General Procedures 4-2-1 Power On/Boot Up Note: After turning off the system, wait at least ten seconds before turning it on again. The system may not be able to boot if power is recycled too quickly.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 4-2-1-4 Front Panel ON/OFF Press the Front Panel ON/OFF key once. to Switch ON/OFF Figure 4-2 ON/OFF key on Front Panel When power is applied to the Scanner, Power is distributed to the Fan, Control panel, Monitor, Internal and External I/O’s, Boards &...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 4-3 Functional Check Procedures To perform these tests, you will need a convex or a linear probe. These procedures should be per formed during installation. These procedures are also used as basic checks to use when service of the system is required.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 4-3-1 Power Up Monitor Display (Convex Probe) Figure 4-3 Power Up Monitor Display (Convex Probe) Chapter 4 - Functional Checks Page 4-5 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 8 0 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 4-3-2 Basic Controls Table 4-2 describes basic controls which helps in checking the LOGIQ™ 100 PRO during installation. Connect the sector or convex transducer to connector on the right side of the base.
Page 81
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL TASK Expected Result CINE (Press Freeze and Rotate B/M Gain/ The Cine Gauge, menu and the Cine frame number Cine Scroll ) appears at the bottom of the display.
Page 82
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Basic Controls (Continued) Figure 4-4 B/M Mode Display Page 4-8 Section 4-1 - Overview Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 8 3 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 4-4 Diagnostics The LOGIQ™ 100 PRO System service diagnostics comprises of: 1.) Self test or power on diagnostics 2.) Service diagnostic tools The self test or power on diagnostics are run every time the system is booted. The service diagnostic tools include test procedures for testing the system at PCB level as well as block level.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 4-5 Software Configuration Checks Step Check Expected Result(s) If Not Remedy Check Date and Time setting Date and Time are correct Adjust the Date and Time setting...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 4-7 Patient contact tools 4-7-1 Probe/Connectors Check Take the probes and check them as described below:. Step Check Expected Result(s) If Not Remedy To test each delivered Probe, Connect it...
Page 86
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Step Check Expected Result(s) If Not Remedy Hold the probe connector horizontally It will display the Probe name on the with the cable pointing Sideways. Screen Turn the connector locking handle to the Vertical position.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 5-1Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ™ 100 PRO’s system concepts, component arrangement, and subsystem function. It also describes the Power Distribution System (PDS) and probes.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 5-2 General Information 1.) Probe Holder 2.) Gel Bottle Holder 3.) Brightness & Contrast 4.) Handle 5.) Monitor 6.) Keyboard 7.) Probe connector 8.) Power Switch Figure 4-5 LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 5-3 Block Diagram Block Diagram Control Davinciboard TRANSMIT Signal ECO Signal Probe Display Front End Board RECIEVE Signal Patient Patient Isolated Sensors External I/O COMPOSITE Signal...
Page 90
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL FEB Block Diagram. Figure 4-7 Block Diagram of FEB LOGIQ™ 100 PRO Page 5-4 Section 5-3 - Block Diagram Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 1 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-3-1 DAVINCI Block Diagram. Figure 4-8 Block Diagram of Davinci board LOGIQ™ 100 PRO Page 5-5 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 2 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-3-2 LOGIQ™ 100 PRO BLOCK DIAGRAM. Page 5-6 Section 5-3 - Block Diagram Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 3 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-3-3 PCB Nomenclature The following table lists the Circuit boards on LOGIQ™ 100 PRO system. Board Abbrv Description Remarks Davinci Davinci Board For LOGIQ™ 100 PRO (All S/W Versions) Keyboard For LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 5-4 Wiring Diagrams 5-4-0-1 Power Distributor/Power Supply Wiring Diagram Figure 4-9 Power Distributor/Power Supply Wiring Diagram for LOGIQ™ 100 PRO. Page 5-8 Section 5-3 - Block Diagram...
Page 95
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Table 4-10 Power Distributor Connector Voltages for LOGIQ™ 100 PRO . Connector J1 Connector J2 Connector J3 Connector J4 Power Supply Input to PDB Output Output Output...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-4-0-2 Cable Assembly Diagrams Davinci board TGC connector Figure 4-10 Cable Assembly Wiring Diagram Page 5-10 Section 5-3 - Block Diagram Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 7 o f 1 9 7...
Page 97
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Cable Assembly Diagrams (Continued) Figure 4-10 Ferrite Core Over which wires wound 2 Turns Page 5-11 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 9 8 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 5-5 Functional Subsystems 5-5-1 Front End Board The Front End Board generates the strong transmit bursts, transmitted by the probes as ultrasound into the body. It also receives weak ultrasound echoes from blood cells and body structure, amplifies the signals and converts them to an 8 bit digital signal.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-5-2 Davinci Board The Davinci CPU board uses TMS320DM6446 chip that contains an ARM Processor, DSP and Video processors into a single chip. The Major functionalities & corresponding implementation is described below.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5-5-3 Peripherals VCR, Black & White Video Graphic Printer and HP Laser Jet Printer 1020 can be connected to the scanner. For more details on recommended peripherals refer Chapter 3.
Page 101
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 5-15 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 0 2 o f 1 9 7...
Page 102
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 5-16 Section 5-7 - Rear Panel Signal List Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 0 3 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Service Adjustments Available in LOGIQ™ 100 PRO The LOGIQ™ 100 PRO does not contain service adjustable parts. In case of failure of any module like the Power Supply, HV pcb it is recommended to replace the power supply unit instead of trying to adjust the potentiometer or replacing the fuse.
Page 104
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Your Notes: Page 6-2 Section 6-1 - Overview Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 0 5 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 7 Diagnostics/Troubleshooting Section 7-1Overview 7-1-1 Purpose of Chapter 7 This section describes how to setup and run the tools and software that help maintain image quality and system operation.
Page 106
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL LED Indications and functions of PCB’s LED Location Function LED - Location on FEB This LED indicates the scan lines,the DSP is scanning.The LED glows when the cursor is moved in the B/M mode LED - Location on FEB This LED indicates the probe connect signal is good.This...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 7-2 Power On Diagnostics The power on diagnostics or self tests are run every time the system is booted. This self tests include testing the validity of the system software through EPROM checksum test and testing the system RAM.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 7-2-1 DAVINCI Board Tests When this test is selected the following screen will appear. DAVINCI BOARD TESTS 1. ALL TESTS 2. EPROM CHECK SUM TEST 3. R/W MEMORY TEST 4.
Page 109
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL DAVINCI Board Tests (Contd.,) Graphics Test The pattern shown below is displayed on the monitor when this test is enabled. Figure 7-3 Graphics Test Press any key to exit the Graphics Test...
Page 110
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL DAVINCI Board Tests (Contd.,) Flash Memory Test The Following Menu is Displayed FLASH MEMORY TESTS ALL TESTS TEST FLASH MEMORY UPDATE ARCHIVE STATUS FROM FLASH ERASE FLASH...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL DAVINCI Board Tests (Contd.,) Erase Flash memory When this option is run all the images stored in the flash is erased. Main Menu This option is used to return to the main menu.
Page 112
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Observe the FEB test pattern on the screen. Change the modes to B/B, B/M and M modes and check for the proper display of the FEB Test pattern.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 7-2-3 Keyboard Tests When this Option is selected following Menu appears KEYBOARD TESTS ALL TESTS KEYBOARD ALIVE TEST KEYS TEST LEDS TEST TRACKBALL TEST GAINKNOB TEST MAIN MENU...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 7-2-4 Monitor Test The following test patterns are displayed on the monitor when this test is enabled. Please match the patterns displayed on your monitor with the given patterns figures 7-7 to 7-13. The Gain Knob is used for selection of patterns 1-10.
Page 115
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Monitor Test (Contd.,) Figure 7-9 Monitor Test Figure 7-10 Monitor Test Chapter 7 - Diagnostics/Troubleshooting Page 7-11 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 1 6 o f 1 9 7...
Page 116
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Monitor Test (Contd.,) Figure 7-11 Monitor Test Figure 7-12 Monitor Test Page 7-12 Section 7-2 - Power On Diagnostics Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 1 7 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Monitor Test (Contd.,) Figure 7-13 Monitor Test Note: The patterns in 7-7 to 7-10 are written on the overlay where as the pattern in 7-11 to 7-13 is written on the image memory.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 7-3 Error Reporting 7-3-1 Error Reporting The following are the error messages available: Errors Remarks “THE KEYBOARD SUBSYSTEM FAILED This Message is displayed during the diagnostics of the Keyboard. When the keyboard TO RESPOND”...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 7-4 Troubleshooting The following table is provided to help the service engineer during corrective maintenance. TROUBLESHOOTING Problem Faced Cause Solution No Display, total black out No AC Power...
Page 120
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Problem Faced Cause Solution System Resets intermittently 3.3 Volt power supply may be less. a. First check for the 3.3V DC supply to the CPU board & ensure Proper Voltage.
Page 121
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Machine Not Booting Chapter 7 - Diagnostics/Troubleshooting Page 7-17 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 2 o f 1 9 7...
Page 122
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 7-18 Section 7-4 - Troubleshooting Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 3 o f 1 9 7...
Page 123
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Machine Resets Intermittently Chapter 7 - Diagnostics/Troubleshooting Page 7-19 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 4 o f 1 9 7...
Page 124
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Junk Image in the Scan Area Page 7-20 Section 7-4 - Troubleshooting Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 5 o f 1 9 7...
Page 125
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Langauge and Date not getting updated. Chapter 7 - Diagnostics/Troubleshooting Page 7-21 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 6 o f 1 9 7...
Page 126
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL No Scan Image, Probe is firing. Page 7-22 Section 7-4 - Troubleshooting Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 7 o f 1 9 7...
Page 127
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Your Notes: Chapter 7 - Diagnostics/Troubleshooting Page 7-23 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 8 o f 1 9 7...
Page 128
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 7-24 Section 7-4 - Troubleshooting Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 2 9 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter holds replacement procedures for the following modules/subsystems. CONTENTS IN CHAPTER 8 Page Section...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-2 Top Cover 8-2-1 Disassembly/Assembly of Top Cover Follow the steps below and refer Figure 8-1, For FRU Details refer 9-1-2 1.) Always Shut Down the System when disassembling the TOP COVER 2.) Disconnect the power cord from the mains and remove from the system.
Page 131
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Disassembly/Assembly (Cont’d.) Handle Cap Probe Holder Figure 8-1 Disassembly/Assembly of Top Cover Chapter 8 - Replacement Procedures Page 8-3 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 3 2 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-3KEYBOARD 8-3-1 Detaching the Keyboard Follow the steps below and refer Figure 8-2. For FRU Details refer 9-1-3 1.) Remove Front Panel. Refer 8-4-1. 2.) Remove 14-pin connector from the Davinci.
Page 133
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Detaching the Keyboard (Cont’d.) Figure 8-2 Detaching the Keyboard Chapter 8 - Replacement Procedures Page 8-5 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 3 4 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 8-3-2 Keyboard Disassembly/Assembly Follow steps below and refer Figure 8-3 . For FRU Details refer 9-1-4 1.) Remove Top Cover. Refer 8-2. 2.) Detach the keyboard. Refer 8-3-1.
Page 135
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Keyboard Assembly/Disassembly (Cont’d.) q (11 Nos) Figure 8-3 Keyboard Disassembly Chapter 8 - Replacement Procedures Page 8-7 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 3 6 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 8-3-2-1 Optical Trackball Assembly Procedure for LOGIQ™ 100 PRO 1.) Assemble TB Base Pad 2230750 (x). Refer Figure 8-4 2.) Assemble Optical Trackball PCB Assembly 2224019 (y) refer Figure 8-4 3.) Ensure that Trackball alignment is OK on the Key sheet.
Page 137
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 8-3-2-2 Optical Trackball Assembly Procedure for LOGIQ™ 100 PRO Assemble TB Base Pad2 2222115. Refer Figure 8-5 9.) Assemble Optical Trackball PCB Assembly 2224019 refer Figure 8-5 10.)Ensure that Trackball alignment is OK on the Key sheet.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-4 FRONT PANEL/MONITOR 8-4-1 Disassembly/Assembly of Front Panel Follow steps below and refer Figure 8-6 For FRU Details refer 9-1-5 1.) Remove Top Cover. Refer 8-2 2.) Lock the Keyboard before removing the front panel.
Page 139
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Disassembly / Assembly of Front Panel (Cont’d.) Lock Hold Power Switch Figure 8-6 Assembly & Disassembly of Front Panel. Chapter 8 - Replacement Procedures Page 8-11 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 0 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 8-4-2 Disassembly/Assembly of Monitor Follow steps below and refer Figure 8-7. For FRU Details refer 9-1-6 1.) Remove Top Cover. Refer 8-2 2.) Remove Front Panel. Refer 8-4-1 3.) Disconnect the communication cables between the Analog and Digital circuits.
Page 141
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Figure 8-7 Disassembly/Assembly of Monitor Chapter 8 - Replacement Procedures Page 8-13 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 2 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-5 PCBs 8-5-1 Disassembly/Assembly of Davinci Board. Follow steps below and refer 1.) Remove Top Cover. Refer 8-2. For FRU Details refer 9-1-7 to 9-1-9 2.) For disassembling both Davinci Board and FEB Board together, remove all communications and power cables from the Davinci board and FEB board.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Disassembly/Assembly of Davinci Board. Figure 8-8 Disassembly/Assembly of Davinci Board. Chapter 8 - Replacement Procedures Page 8-15 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 4 o f 1 9 7...
Page 144
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Disassembly/Assembly of Davinci Board(Cont’d). Figure 8-9 Davinci Board Disassembly Page 8-16 Section 8-5 - PCBs Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 5 o f 1 9 7...
Page 145
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Dip Switch Figure 8-10 Davinci board Dip Switch Setting for NTSC & PAL Chapter 8 - Replacement Procedures Page 8-17 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 6 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 8-5-2 Disassembly/Assembly of FEB for LOGIQ™ 100 PRO Follow steps and refer 1.) Remove Top Cover. Refer 8-2. For FRU Details refer 9-1-10 2.) Remove the screw fixing the FEB PCB to the Rear panel bracket.
Page 147
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Mating connector Davinci & FEB Figure 8-11 Disassembly/Assembly of FEB Chapter 8 - Replacement Procedures Page 8-19 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 4 8 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-6 Fan Assembly 8-6-1 Disassembly/Assembly of Fan Follow steps below and refer Figure 8-12. For FRU Details refer 1.) Remove Top Cover. Refer 8-2. 2.) Remove 2 screws (a,d) and loosen screws b,c from the chassis.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-7 Power Supply/Distributor PCB 8-7-1 Disassembly/Assembly of Power Supply/Distributor PCB Follow steps below and refer 8-13 1.) Remove Top Cover. Refer 8-2. For FRU Details refer 2.) Remove Front Panel.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 13.) For assembly follow the reverse order. Figure 8-13 Disassembly/Assembly of Power Supply/Distributor PCB Service Manual Section Functional Test/ Diagnostic Test Debrief Script 4-2-1 Power ON/ Boot up...
Page 151
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 5.) Remove the connections to the circuit breaker. 6.) Remove the rear panel from the chassis. 7.) For assembly follow the reverse order. Table 8-11 Functional tests...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-9 HV Assembly 8-9-1 Disassembly/Assembly of HV Assembly Follow steps below and refer 8-15 1.) Remove Top Cover. Refer.8-2. For FRU Details refer 2.) Remove the Shield Plate.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 8-10 Chassis 8-10-1 Disassembly/Assembly of Chassis. Follow steps below and refer 8-16 Remove Top Cover. Refer 8-2. For FRU Details refer 9-1-12 Remove Front Panel. Refer 8-4-1 Remove Monitor Assembly.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 1201 1202 Figure 8-16 Disassembly/Assembly of Chassis Page 8-26 Section 8-10 - Chassis Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 5 5 o f 1 9 7...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 9 Replacement Parts Section 9-1Overview 9-1-1 Purpose of Chapter 9 This chapter gives you an overview of Spare Parts for LOGIQ™ 100 PRO. CONTENTS IN CHAPTER 9...
Page 156
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL MATERIAL LIST - LOGIQ™ 100 PRO Part Name Part No. QTY. Description Operator Console Assembly 5304117 100-230VAC, 50/60Hz, PAL, GEMS-A. 5304414 100-230VAC, 50/60Hz, PAL, GEMS-A/AM. 5304240 100-230VAC, 50/60Hz, NTSC, GEMS-A/AM.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-2 MATERIAL LIST - Top Cover OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-1 Top Cover Table 9-1 Top Cover Item Part Name Part No. Description Top Cover Assembly 5318780 LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-3 MATERIAL LIST - KEYBOARD OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-2 Keyboard Table 9-2 Keyboard Item Part Name Part No. Description Keyboard Assembly 5305699 LOGIQ™ 100 PRO...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-4 MATERIAL LIST - KEYBOARD ASSEMBLY OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-3 Keyboard Assy Chapter 9 - Replacement Parts Page 9-5 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 6 0 o f 1 9 7...
Page 160
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Table 9-3 Keyboard Assembly Item Part Name Part No. Description Keyboard PCB 5131077 LOGIQ™ 100 PRO Keysheet With Base Plate 5135264 LOGIQ™ 100 PRO Includes bracket, screws, FPC cable & instruction.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-5 MATERIAL LIST - FRONT PANEL OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Front Panel Figure 9-4 Front Panel Table 9-4 Front Panel Item Part Name Part No. Description Front Panel Assembly 2337067 LOGIQ™...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-6 MATERIAL LIST - Monitor OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-5 Monitor Table 9-5 Monitor Item Part Name Part No. Description Monitor Assembly (CRT+PCB) with 2139789-3 LOGIQ™ 100 PRO...
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-7 MATERIAL LIST - DAVINCI OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-6 Davinci Board Table 9-6 Davinci Item Part Name Part No. Description Davinci Assy (PAL/NTSC) 5212995 LOGIQ™ 100 PRO 1.) Davinci is common for PAL &...
Page 164
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-8 Material List - FEB OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-7 FEB Assembly Table 9-7 Item Part Name Part No. Description FEB Assy 5199849 LOGIQ™ 100 PRO...
Page 165
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-9 MATERIAL LIST - Power Supply/Distributor PCB OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-8 Power Supply/Distributor PCB Table 9-8 Power Supply/Distibutor PCB Item Part Name Part No. Description...
Page 166
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-10 MATERIAL LIST - Rear Panel OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. Figure 9-9 Rear Panel Assembly Table 9-9 Rear Panel Assy Item Part Name Part No. Description Rear panel Assembly 5305191 LOGIQ™...
Page 167
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-11 MATERIAL LIST - HV Power Supply Assembly OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. 1001 Figure 9-10 HV Power Supply Assembly Table 9-10 HV Power Supply Assembly Item Part Name Part No.
Page 168
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 9-1-12 MATERIAL LIST - Cable Assembly OPERATOR CONSOLE ASSY 5304117,5304414,5304240,5304248,5304441,5304596,5306834. FRU No. PART NAME PART NO. DESCRIPTION 1301 LOGIQ™ 100 PRO Cable Set 5308178 For LOGIQ™ 100 PRO Table 9-2 Cable Assembly for LOGIQ™...
Page 169
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Notes: Chapter 9 - Replacement Parts Page 9-15 Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 7 0 o f 1 9 7...
Page 170
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Page 9-16 Section 9-1 - Overview Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 7 1 o f 1 9 7...
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Chapter 10 Quality Assurance Section 10-1 Overview 10-1-1 Purpose of Chapter 10 This chapter describes Quality Assurance & Periodic Maintenance (PM) on the scanner and its peripherals. These PM procedures are intended to maintain the quality of the ultrasound systems performance.
100 PRO periodic maintenance is performed as scheduled in order to retain its high level of safety, dependability and performance. Your GE Service Representative knows your LOGIQ™ 100 PRO best and can provide competent, efficient service. Please contact us for further information and to schedule GE Medical Systems Ultrasound to perform this service for you.
Page 173
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Service at Indicated Time Daily Weekly Monthly Annually Notes •* Clean Probes * or before each use • Clean Probe Holders • Inspect AC Mains Cable Mobile Unit Check Weekly •...
EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Standard GE Tool Kit The following is a description of the “Standard” GE tool kit in the USA. Not all tools are required for PMs Tool ID Description Tool ID Description 9-45358...
Page 175
Flashlight Mini-Mag Lite (AAA Bat.) 9-GH408 Tweezers 21576 Brush - Bristle 9-4516 Pliers 4 1/4 inch Diagonal Table 10-3 Overview of GE-1 Tool Kit Contents (Continued) GE-2 Sears Kit (#99034) Tool ID Description Tool ID Description 9-45381 Pliers, Arc Joint 9 1/2 inch 9-44067 Socket 1 1/16 in.
9-GH19512 Countersink 1/2 inch 9-44046 3/8 inch Socket Set 9-44741 12 PC Combination Wrench Set Table 10-4 Overview of GE-2 Tool Kit Contents (Continued) 10-4-2 Special Tools, Supplies and Equipment 10-4-2-1 Specific Requirements for Periodic Maintenance See Chapter 7 Tool...
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 10-5 System Periodic Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary. Step Item Description Ask & Listen Ask the customer if they have any problems or questions about the equipment.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks (See Also Chapter 4) The functional checks take about 30 minutes to perform. Refer to the system user documentation whenever necessary. 10-5-2-1 System Checks...
Verify hardcopy output of the B/W printer. Clean heads and covers if necessary. Printer Footswitch Verify that the footswitch is functioning as programed. Clean as necessary. Table 10-8 GE Approved Peripheral/Hardware Option Functional Checks 10-5-3 Input Power 10-5-3-1 Mains Cable Inspection...
The system user manuals and various probe handling cards provide a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. See the User Manual and probe care cards for more details.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-5-6-3 Basic Probe Cleaning Refer to the User’s Manual for details on probe cleaning. To help protect yourself from blood borne diseases, wear approved disposable gloves. These are made of nitrile derived from vegetable starch to prevent allergic latex reactions. Failure to follow the prescribed cleaning or disinfection procedures will void the probe’s warranty.
Page 182
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Test the system, peripherals and probes for leakage current. Excessive leakage current can cause injury or death in sensitive patients. High leakage current can also indicate degradation of insulation and a potential for electrical failure.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-2 GEMS Leakage Current Limits The following limits are summarized for NFPA 99 (For USA) and IEC 60601-1 Medical Equipment Safety Standards. These limits are GEMS standards and in some cases are lower than the above standards listed.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-4 Grounding Continuity Electric Shock Hazard. The patient must not be contacted to the equipment during this test Measure the resistance from the third pin of the attachment plug to the exposed metal parts of the case.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-5 Chassis Leakage Current Test 10-6-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible metal parts of the bedside station if the ground wire should break. The test verifies the isolation of the power line from the chassis.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-5-3 Dale 600 Meter Procedure When measuring system chassis currents with the Dale 600, always use the CHASSIS selection of the external/chassis function switch. This requires the ground clip lead and changing the meters switches in accordance with the IEC 601-1.1.
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-5-4 Data Sheet for Chassis Source Leakage Current The test passes when all readings measure less than the value shown in Table 10-13. Record all data on the PM Inspection Certificate.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-6 Probe Leakage Current Test 10-6-6-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface. 10-6-6-2 Generic Procedure for Leakage current Measurements should be made with the ground open and closed, with power line polarity normal...
Page 190
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 7.) Power ON the LOGIQ™ 100 PRO unit. 8.) Depress the LIFT GROUND rocker switch and record the highest current reading. 9.) Follow the test conditions described in Table 10-18 for every probe.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10-6-6-4 Meter Procedure using Probe Adapter to measure Probe isolation(Sink) Current The Dale 600/600E provides a method for testing probes independently from the system.The meter utilizes a probe adapter to apply a test potential commonly to all connector pins. The probe's imaging area is immersed in a saline solution along with a grounding probe from the meter to complete the current path.Saline solution is a mixture of water and salt.
Page 192
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL 10.6.6.5 Data sheet for Probe Source Leakage Current The test passes when all readings measure less than the values shown in Table 10-14 and Table 10-15. Record all data on the Planned Maintenance Inspection Certificate.
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Section 10-7 When There's Too Much Leakage Current... CHASSIS FAILS Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent. Replace any defective part. Tighten all grounds.
Page 194
(if applicable) Inspect Clean B-Mode Function Console M-Mode Function Monitor Control Panel Cables and Connectors Monitor GE Approved Peripherals (VCR, VGP) Applicable Software Options Probe Holders Applicable Hardware Options External I/O Measurement Accuracy GE Approved Peripherals COMMENTS: ________________________________________________________________ ________________________________________________________________ ________________________________________________________________...
Page 195
ALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe Patient Lead Source Leakage (Lead to Ground) Patient Lead Source Leakage (Lead to Lead)
Page 196
GE H EALTHCARE 5305880-100, R LOGIQ™ 100 PRO S IRECTION EVISION ERVICE ANUAL Your Notes: Page 10-26 Section 10-6 - Electrical Safety Tests Ap p r o v e d Do c u m e n t - 5 3 0 5 8 8 0 - 1 0 0 T PH_ r 3 . p d f Pa g e 1 9 7 o f 1 9 7...