Which Images Can Be Cad-Analyzed; Which Image Formats Can Be Displayed With Cad Markers - GE Senographe SecondLook Operator's Manual

Digital cad system
Table of Contents

Advertisement

GE Healthcare
Revision 1
4-3

Which images can be CAD-analyzed?

All images to be analyzed by the SLDU must have been generated on a licensed Senographe system.
The SLDU algorithm will analyze any raw (RAW) full breast mammographic image.
The following images cannot be analyzed:
Processed (PROCESSED) and secondary capture (SCPT) images.
System performance is not assured for the following views:
Segmented views of the breast.
Implant views with the full implant imaged. Implant Displaced views with a maximum of 25 mm (1") of
breast implant imaged can be analyzed.
Special diagnostic images (e.g., magnification views, or spot compression views).
Cleavage views.
Images with hook wires.
Collimated images.
Coned images.
4-4

Which image formats can be displayed with CAD markers?

The following image formats can be displayed with CAD markers:
RAW images
Processed images
Premium View images, for users who have the Premium View Option on the review workstation.
4-5
Selecting images for CAD analysis
You can choose to use CAD to process an individual image, a set of images, a series, or an entire
patient case.
raw
Only
images can be CAD-processed. If you send an entire patient case to the SLDU, the raw
images are CAD-processed; other images present (processed or secondary-capture images) are
ignored, but increase the time required to obtain the overlays.
To obtain the best analysis of available information, send entire patient cases (rather than individual
images) to be CAD-processed. The SLDU takes advantage of all available information during
analysis. If different views of the same breast are present, the SLDU uses all screening views in its
analysis, just as a clinician would.
However, note that this increases the time required to obtain the overlays; processing a 4-view exam
takes two minutes, as opposed to 30 seconds for a single view.
To minimize the elapsed time before CAD overlays are available, send only raw images.
CAD.fm
Senographe SecondLook Digital CAD System
Page no. 15
Operator Manual 5189820-5-C-1EN

Hide quick links: