TYPICAL CRYOGENIC VENT PIPE DETAIL
Additional pipe/tubing
WAVEGUIDE
Typical
SIGNA EXPLORER /SIGNA CREATOR
[0.5 in] mesh
DIELECTRIC
BREAK
VENTGLAS
CONNECTOR
VENT PIPE
610 mm [24 in], may be cut to a
minimum of 100 mm [4 in].
MAGNET VENT ADAPTER
Do not remove or modify the
vent adaptor bolted to the
magnet.
EN-MRI-TYP-SIGNA_EXPLORER-WEB.DWG
TYPICAL CRYOGEN SIDE WALL EXIT WITH LONG SWEEP ELBOW
KEY COMPONENTS :
RF waveguide extended from wall to magnet adapter.
Must be all same material and all welded.
Support system must withstand 8229 N (1850 lbs)
GE ventglass must be installed in vertical section directly over magnet
Mesh bird screen
RF WAVEGUIDE EXTENDED
203 [8 in]
ventglas outside
RF room
OUTSIDE
CRYOGENIC VENTING ( EXTERIOR )
OUTSIDE WALL
EXHAUST AREA
VENT CAP
6.1m x 4.6m
(customer supplied)
[20ft x 15ft] (LxW)
GROUND
VENT CAP
(customer
supplied)
EXHAUST AREA
6.1m x 4.6m
[20ft x 15ft] (LxW)
ROOF
NOT TO SCALE
Rev
A
Date
38 (1.5 in)
rigid support
insulation
Warning sign
Exclusion area
03/Aug/2018
M4 - Cryogenics (1)
25
[1 in
]
17
/24