4.) On the Utility touch panel, press System.
5.) On the monitor display, select Backup/Restore.
The Backup/Restore screen is displayed.
6.) In the Backup list,
-
-
NOTE:
The detailed section of this menu decouples the user defined configuration above. This allows you to
selectively restore what you want to restore across multiple machines.
7.) Specify where to save data in the media field.
8.) Select Backup.
The system performs the backup. As it proceeds, status information is displayed on the Backup/
Restore screen.
9.) At the end of the process, the Backup completed message is displayed on the monitor.
Press Eject (F3) for eject media/disconnect USB.
10.)Make sure to physically label the media. An identification of the system should also be noted on the
media and a backup log should be kept.
File the media in a safe place.
8-5-3

Restore Procedures

The restore procedure overwrites the existing database on the local hard drive. Make sure to
CAUTION
insert the correct media.
!! CAUTION:
You cannot restore the data between systems with different software versions.
To avoid the risk of overwriting the local patient and report archives, DO NOT check Patient
CAUTION
Archive when restoring userdefined configurations.
!! CAUTION:
1.) On the touch panel, press Utility.
2.) On the Utility touch panel, press System.
3.) On the monitor display, select Backup/Restore.
The Backup/Restore screen is displayed.
4.) In the Restore list,
-
-
or
One or several system configuration items to restore parts of the Detailed Restore of User Defined.
5.) In the Media field, select the appropriate Source device.
6.) Select Restore.
The system performs the restore. As it proceeds, status information is displayed on the Backup/
Restore screen.
7.) The LOGIQ S7 Expert/Pro restarts automatically when Restore is done.
8-34
3
DRAFT (J
Select Patient Archive and Report Archive to backup the patient records.
Select User Defined Configuration to copy system settings and user presets.
Select Patient Archive and Report Archive to restore the patient archive.
Select User Defined Configuration to restore all system settings and user presets.