  • Page 1 LOGIQ™ S7 Expert/Pro Service Manual ADVANCED SERVICE DOCUMENTATION. PROPERTY OF GE HEALTHCARE. USE OF THESE MATERIALS LIMITED TO AGENTS AND EMPLOYEES OF GE HEALTHCARE OR OTHER PARTIES EXPRESSLY LICENSED BY GE HEALTHCARE. UNLICENSED USE IS STRICTLY PROHIBITED Document Number: 5460683...
  • Page 3 • VERSUCHEN SIE NICHT, DAS GERä T ZU REPARIEREN, BEVOR DIESES WARNUNG KUNDENDIENST-HANDBUCH NICHT ZU RATE GEZOGEN UND VERSTANDEN (DE) WURDE. • WIRD DIESE WARNUNG NICHT BEACHTET, SO KANN ES ZU VERLETZUNGEN DES KUNDENDIENSTTECHNIKERS, DES BEDIENERS ODER DES PATIENTEN DURCH ELEKTRISCHE SCHLä GE, MECHANISCHE ODER SONSTIGE GEFAHREN KOMMEN.
  • Page 4 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL ESTE MANUAL DE SERVICIO Só LO EXISTE EN INGLé S. • SI ALGú N PROVEEDOR DE SERVICIOS AJENO A GEHC SOLICITA UN IDIOMA QUE NO SEA EL INGLé S, ES RESPONSABILIDAD DEL CLIENTE OFRECER UN SERVICIO DE TRADUCCIó...
  • Page 5 ΣΤΟΝ ΑΣΘΕΝΗ ΑΠΟ ΗΛΕΚΤΡΟΠΛΗΞΙΑ, ΜΗΧΑΝΙΚΟΥΣ Ή ΑΛΛΟΥΣ ΚΙΝ∆ΥΝΟΥΣ. EZEN KARBANTARTÁSI KÉZIKÖNYV KIZÁRÓLAG ANGOL NYELVEN ÉRHETŐ EL. • HA A VEVŐ SZOLGÁLTATÓJA ANGOLTÓL ELTÉRŐ NYELVRE TART IGÉNYT, AKKOR A VEVŐ FELELŐSSÉGE A FORDÍTÁS ELKÉSZÍTTETÉSE. • NE PRÓBÁLJA ELKEZDENI HASZNÁLNI A BERENDEZÉST, AMÍG A FIGYELMEZTETÉS KARBANTARTÁSI KÉZIKÖNYVBEN LEÍRTAKAT NEM ÉRTELMEZTÉK.
  • Page 6 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL ÞESSI ÞJÓNUSTUHANDBÓK ER EINGÖNGU FÁANLEG Á ENSKU. • EF ÞJÓNUSTUAÐILI VIÐSKIPTAMANNS ÞARFNAST ANNARS TUNGUMÁLS EN ENSKU, ER ÞAÐ Á ÁBYRGÐ VIÐSKIPTAMANNS AÐ ÚTVEGA ÞÝÐINGU. • REYNIÐ EKKI AÐ ÞJÓNUSTA TÆKIÐ NEMA EFTIR AÐ HAFA SKOÐAÐ OG VIÐVÖRUN...
  • Page 7 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL šĪ APKALPES ROKASGRĀMATA IR PIEEJAMA TIKAI ANGĻU VALODĀ. • JA KLIENTA APKALPES SNIEDZĒJAM NEPIECIEŠAMA INFORMĀCIJA CITĀ VALODĀ, NEVIS ANGĻU, KLIENTA PIENĀKUMS IR NODROŠINĀT TULKOŠANU. • NEVEICIET APRĪKOJUMA APKALPI BEZ APKALPES ROKASGRĀMATAS BRĪDINĀJUMS...
  • Page 8 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL ACEST MANUAL DE SERVICE ESTE DISPONIBIL NUMAI ÎN LIMBA ENGLEZĂ. • DACĂ UN FURNIZOR DE SERVICII PENTRU CLIENŢI NECESITĂ O ALTĂ LIMBĂ DECÂT CEA ENGLEZĂ, ESTE DE DATORIA CLIENTULUI SĂ FURNIZEZE O TRADUCERE.
  • Page 9 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL TA SERVISNI PRIROČNIK JE NA VOLJO SAMO V ANGLEŠČINI. • ČE PONUDNIK SERVISNIH STORITEV ZA STRANKO POTREBUJE NAVODILA V DRUGEM JEZIKU, JE ZA PREVOD ODGOVORNA STRANKA SAMA.
  • Page 10 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL BU SERVİS KILAVUZU YALNIZCA İNGİLİZCE OLARAK SAĞLANMIŞTIR. • EĞER MÜŞTERİ TEKNİSYENİ KILAVUZUN İNGİLİZCE DIŞINDAKİ BİR DİLDE OLMASINI İSTERSE, KILAVUZU TERCÜME ETTİRMEK MÜŞTERİNİN SORUMLULUĞUNDADIR. DİKKAT • SERVİS KILAVUZUNU OKUYUP ANLAMADAN EKİPMANLARA MÜDAHALE (TR) ETMEYİNİZ.
  • Page 11 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL (ZH-CN) (KO)
  • Page 12 GE Healthcare personnel. In performing all electrical work on these products, GE will use its own specially trained field engineers. All of GE’s electrical work on these products will comply with the requirements of the applicable electrical codes.
  • Page 13 The contents of this publication may not be copied or duplicated in any form, in whole or in part, without prior written permission of GE Healthcare. GE Healthcare may revise this publication from time to time without written notice. TRADEMARKS All products and their name brands are trademarks of their respective holders.
  • Page 14 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Revision History Revision Date Reason for change July 2012 Initial release August 2012 Had some feedback from Service team June 2014 Updated for R2.x.x or later List of Effected Pages (LOEP)
  • Page 15: Table Of Contents

    GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Table of Contents CHAPTER 1 Introduction Overview ............1 - 1 Purpose of Chapter 1 .
  • Page 16 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 2 Site Preparation Overview ............2 - 1 Purpose of Chapter 2 .
  • Page 17 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 3 Setup Instructions Overview............3 - 1 The Purpose of Chapter 3 .
  • Page 18 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Scanner Shutdown ........3 - 26 Transducer Connection .
  • Page 19 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Connection via a Peer-to-Peer Network ..... 3 - 50 Connection via Hospital Network .
  • Page 20 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 4 Functional Checks Overview ............4 - 1 Purpose of Chapter 4 .
  • Page 21 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL M-Mode ..........4 - 26 B-Flow Mode .
  • Page 22 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 5 Components and Functions (Theory) Overview ............5 - 1 Purpose of Chapter 5 .
  • Page 23 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Common Service Desktop (CSD)........5 - 30 Internationalization .
  • Page 24 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 6 Service Adjustments Overview ............6 - 1 Purpose of Chapter 6 .
  • Page 25 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 7 Diagnostics/Troubleshooting Overview............7 - 1 Purpose of Chapter 7 .
  • Page 26 GE Icon and Remote Connection ....... . 7 - 15...
  • Page 27 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 8 Replacement Procedures Overview............8 - 1 Purpose of Chapter 8 .
  • Page 28 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Replacement Front Cover ..........8 - 37 Manpower .
  • Page 29 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Preparations ..........8 - 48 Monitor Remove Procedure .
  • Page 30 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Removal Procedure - Separating PIL and GBF128/64 ... 8 - 74 Removal Procedure - Separating CWD Board from PIL ..8 - 75 Handling EEPROM on GFS Board .
  • Page 31 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Removal Procedure (R1.x.x) ........8 - 91 Removal Procedure (R2.x.x or later) .
  • Page 32 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Removal Procedure ........8 - 113 Installation Procedure .
  • Page 33 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Installation Procedure ......8 - 125 Functional Check .
  • Page 34 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Drawer on High Cabinet ........8 - 137 Manpower .
  • Page 35 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 9 Renewal Parts Overview............9 - 1 Purpose of Chapter 9 .
  • Page 36 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL CHAPTER 10 Care & Maintenance Overview ............10 - 1 Periodic Maintenance Inspections .
  • Page 37 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Outlet Test - Wiring Arrangement - USA & Canada ....10 - 24 Grounding Continuity .
  • Page 38 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL xxxvi Table of Contents...
  • Page 39: Overview

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 1 Introduction Section 1-1 Overview 1-1-1 Purpose of Chapter 1 This chapter describes important issues related to safely servicing the LOGIQ™ S7 ultrasound system.
  • Page 40: Typical Users Of The Basic Service Manual

    IRECTION EVISION ERVICE ANUAL 1-1-3 Typical Users of the Basic Service Manual • GE Service Personnel (installation, maintenance, etc.). • Hospital’s Service Personnel • Contractors (Some parts of Chapter 2 - Pre-Installation) 1-1-4 Models Covered by this Manual Table 1-2 LOGIQ™...
  • Page 41: Purpose Of Operator Manual(S)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-1-5 Purpose of Operator Manual(s) The Operator Manual(s) should be fully read and understood before operating the LOGIQ™ S7 and also kept near the unit for quick reference.
  • Page 42: Standard Hazard Icons

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-2-2 Standard Hazard Icons Important information will always be preceded by the exclamation point contained within a triangle, as seen throughout this chapter. In addition to text, several different graphical icons (symbols) may be used to make you aware of specific types of hazards that could cause harm.
  • Page 43: Product Icons

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-2-3 Product Icons The following table describes the purpose and location of safety labels and other important information provided on the equipment.
  • Page 44 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 1-5 Product Icons (Continued) LABEL/SYMBOL PURPOSE/MEANING LOCATION "CAUTION - Dangerous voltage" (the lightning flash with arrowhead in equilateral triangle) is used to indicate electric various shock hazards.
  • Page 45: Safety Considerations

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 1-3 Safety Considerations 1-3-1 Introduction The following safety precautions must be observed during all phases of operation, service and repair of this equipment.
  • Page 46: Electrical Safety

    DO NOT attempt to use different peripherals and accessories (brand and model; WARNING WARNING connected via USB ports) other than approved and provided by GE Healthcare! !! WARNING: The ultrasound system is an extremely sensitive and complex medical system. Any unauthorized peripherals may cause system failure or damage! The LOGIQ™...
  • Page 47 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-3-5 Auxiliary Devices Safety (cont’d) Everybody who connects additional equipment to the signal input portion or signal output portion configures a medical system, and is therefore responsible that the system complies with the requirements of the system standard IEC 60601-1-1.
  • Page 48: Battery Safety (R2.X.x Or Later)

    NOTE: Regulations vary for different countries. Dispose of a used battery in accordance with local regulations. USE ONLY BATTERIES APPROVED BY GE HEALTHCARE AS SUITABLE FOR USE WITH THE CAUTION LOGIQ™ S7 Expert/Pro ULTRASOUND SYSTEM. THE LOGIQ™ S7 EXPERT/PRO BATTERY HAS A SAFETY DEVICE.
  • Page 49: Labels Locations

    1.) LCD Caution Label 2.) Gender Caution Label (Only for India, China, Korea) 3.) Multi Caution Label 4.) LOGIQ S7 Expert/Pro Rating Label (For China, Korea, Japan) 5.) LOGIQ S7 Expert/Pro Rating Label 6.) UL Label Chapter 1 - Introduction...
  • Page 50: Dangerous Procedure Warnings

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-3-8 Dangerous Procedure Warnings Warnings, such as the examples below, precede potentially dangerous procedures throughout this manual. Instructions contained in the warnings must be followed.
  • Page 51: Electromagnetic Compatibility (Emc)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1-3-10 Returning/Shipping System, Probes and Repair Parts (cont’d) NOTE: The US Department of Transportation (DOT) has ruled that “items what were saturated and/or dripping with human blood that are now caked with dried blood;...
  • Page 52: Customer Assistance

    Latin America Applications support 1-262-524-5698 Europe OLC - EMEA (Europe, Middle East & Africa) GE Ultraschall Deutschland GmbH Phone: +49 (0) 212 2802 - 652 (-OLC) Beethovenstraße 239 +33 1 3083 1300 (English/German all segments incl. training) Postfach 11 05 60, D-42655 Solingen...
  • Page 53: Overview

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 2 Site Preparation Section 2-1 Overview 2-1-1 Purpose of Chapter 2 This chapter provides the information required to plan and prepare for the installation of a LOGIQ™ S7 Expert/Pro ultrasound unit.
  • Page 54: General Console Requirements

    Electrical Requirements NOTE: GE Healthcare requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This dedicated power shall originate at the last distribution panel before the system. The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full size Ground wire from the distribution panel to the Ultrasound outlet.
  • Page 55: Logiq™ S7 Expert/Pro Power Requirements

    The LOGIQ™ S7 ultrasound system is supplied with a main power plug, as standard. In the event that the unit arrives without a power plug, or with the wrong plug, contact your GE dealer. When necessary, the installation engineer will supply the appropriate power plug to meet the applicable local regulations.
  • Page 56: Emi Limitations

    RF gaskets touch metal. appropriate location. The interconnect cables are grounded and require ferrite beads and other shielding. Use GE- specified Cable length, material, and routing are all important; do not make any changes that do not meet harnesses and peripherals.
  • Page 57: Probe Environmental Requirements

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2-2-4 Probe Environmental Requirements Table 2-5 2D Probe Environmental Requirements Operational - 35 for 2D probe Temperature - 30 for 4D probe...
  • Page 58: System Specifications

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2-2-6 System Specifications 2-2-6-1 Physical Dimensions of LOGIQ™ S7 Expert/Pro The physical dimensions and weight (without Peripherals) of the LOGIQ™ S7 Expert/Pro unit are...
  • Page 59: Facility Needs

    The products involved (and the accompanying electrical installations) are highly sophisticated and special engineering competence is required. All electrical work on these products must comply with the requirements of applicable electrical codes. The purchaser of GE equipment must only utilize qualified personnel to perform electrical servicing on the equipment.
  • Page 60: Mandatory Site Requirements

    NOTE: All relevant preliminary network port installations at the prepared site must be performed by authorized contractors. The purchaser of GE equipment must utilize only qualified personnel to perform servicing of the equipment. 2-3-3 Site Recommendations The following are (optional) site recommendations.
  • Page 61: Recommended Ultrasound Room Layout

    Recommended Ultrasound Room Layout Figure 2-1 below shows a floor plan illustrating the recommended layout of the Ultrasound Room and depicting the minimal room layout requirements. Dedicated Power Outlets LOGIQ S7 Expert/Pro Hospital Network Cabinet for Software and Manuals (optional) 35.5 IN.
  • Page 62: Networking Setup Requirements

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2-3-4 Networking Setup Requirements 2-3-4-1 Stand-alone Unit (without Network Connection) None 2-3-4-2 Unit Connected to Hospital’s Network Supported networks: • Ethernet network connection (R1.x.x/R2 and later) •...
  • Page 63: Setup Instructions

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 3 Setup Instructions Section 3-1 Overview 3-1-1 The Purpose of Chapter 3 This chapter contains information needed to setup the LOGIQ™ S7 ultrasound system. Included are procedures to receive, unpack and configure the equipment.
  • Page 64: Installation Warnings

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-2-2 Installation Warnings 1.) Since the LOGIQ™ S7 Expert/Pro weighs approximately 90 kg (200 lbs.) without peripherals, two people are required to unpack it.
  • Page 65: Brake Pedal Operation

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-2-2-4 Brake Pedal Operation REMEMBER: If the front wheel brakes are engaged for transportation, release brake WARNING WARNING pedals (brakes on front wheels under the foot rest) to disengage the lock.
  • Page 66: Receiving And Unpacking The Equipment

    Unpack the devices such a way that packaging can be reused. For warranty purposes, storage of the above is required for one year from date of purchase. If the shipping crate is damaged, please inform the GE Healthcare sales representative immediately. NOTICE...
  • Page 67 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-3 Receiving and Unpacking the Equipment (cont’d) The envelope with delivery address, packing list and invoice is located on the front panel of the crate.
  • Page 68 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Remove the option box on front of the system.. option box on front of the system Remove the monitor pads from the both sides of the monitor and remove the 4 pads of the box bottom assy..
  • Page 69: Packing Materials - Recycling Information

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 3-5 Unpacking Procedure Step Task Unlock the casters and place the 4 caster's derection like the picture below and then using the incline plane of the box bottom assy, slowly move the system from the box.
  • Page 70: Preparing For Set Up

    Check the power cord for cuts, loose hardware, tire marks, exposed insulation, or any deterioration. Power Cord Verify continuity. Replace the power cord, as required. NOTE: Report any items that are missing, back-ordered, or damaged, to your GE Healthcare sales representative. The contact address is shown in Contact Information on page 1-14.
  • Page 71: Emi Protection

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-5-2 EMI Protection This unit has been designed to minimize the effects of Electo-Magnetic Interference (EMI). Many of the covers, shields, and screws are provided primarily to protect the system from image artifacts caused by this interference.
  • Page 72: Connection Of Auxiliary Devices

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-6 Connection of Auxiliary Devices NOTE: Normally auxiliary devices and peripherals come pre-installed with the system. Table 3-7 below outlines hardware installation procedures described in the sub-sections.
  • Page 73: Connecting The Lcd Monitor

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-1 Connecting the LCD Monitor NOTE: The LCD monitor comes pre-installed with the system. Figure 3-5 Connection Scheme - LCD Monitor Chapter 3 - Setup Instructions...
  • Page 74: Connecting The Black & White Printer

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-2 Connecting the Black & White Printer 1.) Power OFF/Shutdown the system as described in: Section 3-7-3 on page 3-25. 2.) Connect the Black & White printer according to correct connection scheme as described in...
  • Page 75: Connection Scheme: B&W Printer

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-2-1 Connection Scheme: B&W Printer Figure 3-6 B&W Printer connection Chapter 3 - Setup Instructions 3-13...
  • Page 76: Connecting The Color Printer

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-3 Connecting the Color Printer 1.) Power OFF/Shutdown the system as described in: Section 3-7-3 on page 3-25. 2.) Connect the Color printer according to correct connection scheme as described in...
  • Page 77: Connection Scheme: Color Printer

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-3-1 Connection Scheme: Color Printer Figure 3-7 Color Printer connection Chapter 3 - Setup Instructions 3-15...
  • Page 78: Connecting The Secondary "Patient" Lcd Monitor

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-4 Connecting the Secondary “Patient” LCD Monitor Secondary LCD “Patient” Monitor MUST NEVER be connected to the LOGIQ™ S7 ultrasound CAUTION systems mains supply directly!
  • Page 79 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-4 Connecting the Secondary “Patient” LCD Monitor (cont’d) appropriate Transformer Secondary Monitor HDMI to DVI / HDMI to HDMI cable connected to the...
  • Page 80: Connecting The Footswitch

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-5 Connecting the Footswitch The footswitch should be directly connected to any accessible USB-port on the LOGIQ™ S7 Expert/Pro (e.g., on rear of the system).
  • Page 81: Connecting The Usb Flash Memory Stick

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-6 Connecting the USB Flash Memory Stick Before connecting an USB device, please read General Remarks and Hints when using external USB-...
  • Page 82: General Remarks And Hints When Using External Usb-Devices

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-6-8 General Remarks and Hints when using external USB-Devices Do not connect or disconnect any external USB-devices to or from the system while...
  • Page 83: Completing The Set Up

    The connection of the LOGIQ™ S7 Expert/Pro ultrasound unit to a power source should be performed by a qualified person who has completed basic LOGIQ™ S7 Expert/Pro System User Training. Use only the power cords, cables and plugs provided by or designated by GE Healthcare to connect the unit to the power source.
  • Page 84: Back End Processor Boot Up

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-7-2-2 Back End Processor Boot Up When AC power is applied to the scanner, the ON/OFF button on the control panel illuminates green, NOTICE indicating that the System (including the Back-end Processor) is in Standby mode.
  • Page 85: During A Normal Boot, You May Observe

    3.) The Start Screen (LOGIQ™) is displayed on the monitor. 4.) Start-up progress bars indicating software loading procedures, are displayed on the monitor, as shown in Figure 3-12. and Figure 3-12 below. Figure 3-12 GE Healthcare wallpaper with progress bar (R1.x.x) Chapter 3 - Setup Instructions 3-23...
  • Page 86 Figure 3-13 GE Healthcare wallpaper with progress bar (R2.x.x or later) NOTE: Startup GE Healthcare wallpaper with progress bar may differ depending on product type and release. 5.) The software initiates and sets up the Front-End electronics and the rest of the scanner (incl.
  • Page 87: Power Off / Shutdown

    Use this button to exit from the System-Exit menu and return to the previous operation. • Exit (Only available when logged in as GE Service with Service Dongle) Select this button when you want to exit to the Windows Desktop.
  • Page 88: Scanner Shutdown

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-7-3-1 Scanner Shutdown 1.) If not already in read mode, freeze the image. 2.) Press the ON/OFF button (see: Figure 3-15 Figure 3-16) on the control panel.
  • Page 89 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5.) After complete power down, disconnect the main power cable from the system or unplug it from the AC wall outlet socket.
  • Page 90: Transducer Connection

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-7-4 Transducer Connection 3-7-4-1 Connecting the Probe Inspect the probe before and after each use for damage or degradation to the housing, strain CAUTION relief, lens, seal, cable and connector.
  • Page 91: Connecting The Cw Pencil Probe

    When deactivating the probe, the probe is automatically placed in standby mode. To deactivate a probe: 1.) Ensure the LOGIQ S7 Expert/Pro is in freeze mode. If necessary, press the Freeze key. 2.) Gently wipe the excess gel from the face of the probe.
  • Page 92: Printer Installation

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-8 Printer Installation NOTE: For Connection schemes refer to Section 3-6 "Connection of Auxiliary Devices" on page 3-10. For further installation instructions see: •...
  • Page 93: Installing The Bluetooth Deskjet Printer

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-3 Installing the Bluetooth Deskjet Printer 1.) Power off/Shutdown the system as described in: Section 3-7-3 on page 3-25. 2.) Physically connect the Bluetooth dongle as described on Section Section 3-8-3-1 on page 3-31.
  • Page 94: Verifying Bluetooth Device Driver

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-3-2 Verifying Bluetooth device driver 1.) Turn on the system. 2.) Enter maintenance mode by using service dongle. 3.) Go to Device Manager by pressing shortcut on the taskbar 4.) Check that there is “Bluetooth Radios”...
  • Page 95: Prepare The Printer And Connect The Bluetooth Printer Adapter

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-3-3 Prepare the Printer and connect the Bluetooth Printer Adapter 3-8-3-3-1 HP Officejet 100 1.) Connect the power cord to a wall outlet socket.
  • Page 96: Adding Printer To The System

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-4 Adding Printer to the system 1.) Select Utility -> Connectivity -> Service. 2.) Add the Standard Print. 3.) Highlight Standard Print in the Service list.
  • Page 97: Up-D897 - Printer Settings

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-5-1 UP-D897 - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-8-5 "Adjustment of Printer Settings" 2.) Select the...
  • Page 98: Up-D25Md - Printer Settings

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-5-2 UP-D25MD - Printer Settings 1.) Call up the ‘Printer Preferences’; operation see: Section 3-8-5 "Adjustment of Printer Settings" Settings for Paper Size MUST match with the used Paper (large/small) and also the right color ink NOTICE cartridge has to be used.
  • Page 99 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL b.) Color Correction: set Printer Hardware Color Correction1 Figure 3-26 Graphics page (Color Balance + Color Correction) c.) Lightness: Sharpness = 7; Dark = 0; Gamma = 14; Light = 0 d.) Lightness: Gamma Curve = Curve 1...
  • Page 100: Setting Printer To Print Reports

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-8-6 Setting Printer to Print Reports 3-8-6-1 Setting up the Printer to Print Reports To set up the Off-Line Printer to print reports, 1.) Press Utility ->...
  • Page 101: System Configuration

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-9 System Configuration 3-9-1 System/General Preset Menu The System/General screen allows you to specify hospital name and system date and time.
  • Page 102 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 3-10 General User Interface Preset Parameter Description Color Level (restart needed) Select System Color according to the condition of the room.
  • Page 103 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 3-15 Scan Assistant Preset Parameter Description Auto Selection Off, Category, or Description. • Off. The Scan Assistant selection on the Patient screen is completely manual. It will say “None”...
  • Page 104: External I/O Connectors

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-9-2 External I/O Connectors Figure 3-30 External I/O Connectors - on Rear of System Table 3-16 External I/O Connector - Description Item...
  • Page 105: Available Probes

    DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-10 Available Probes 3-10-1 Supported probes The LOGIQ S7 Expert/Pro supports the following types of probes: • Sector Phased Array • Convex Array • Microconvex Array • Linear Array • Matrix Array •...
  • Page 106 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Part Part Probe Name Illustration Application Feature Number Type Qty FRU 8C-D Pediatrics, B, CHI, CF, PDI, M, PW, CrossXBeam, Neonatal LOGIQView, ATO/ASO, SRI-HD,...
  • Page 107 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Part Part Probe Name Illustration Application Feature Number Type Qty FRU L8-18i-D Small Parts, B, CHI, CF, PDI, M, PW, B-Flow, B-Flow...
  • Page 108: Software/Option Configuration

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Part Part Probe Name Illustration Application Feature Number Type Qty FRU Cardiac, Vascular CW, ASO Pencil TE100024 Probe Cardiac, Vascular CW, ASO...
  • Page 109: Connectivity Setup

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-12 Connectivity Setup The LOGIQ™ S7 Expert/Pro ultrasound system can be connected to various connectivity devices. The following sections describe how to connect the system to a remote archive/work station or a DICOM service, using a TCP/IP connection.
  • Page 110: Dataflow Examples

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-12-1-2 Dataflow Examples The local database is used for patient archiving. Images are stored to internal hard drive. The local database is used for patient archiving.
  • Page 111: Stand-Alone Logiq™ S7 Expert/Pro

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-12-1-3 Stand-alone LOGIQ™ S7 Expert/Pro If digital images or 3D/4D data sets (if available) are stored, they should be saved in the Archive (Image Management System software).
  • Page 112: Wired Ethernet From Logiq™ S7 Expert/Pro To A Workstation

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-12-2 Wired Ethernet from LOGIQ™ S7 Expert/Pro to a Workstation 3-12-2-1 Direct Cable Connection from the LOGIQ™ S7 Expert/Pro to a Workstation via a Crossover Cable.
  • Page 113: Configuring Connectivity

    This greatly improves query capability. DICOM Structured Reporting creates coded clinical data that can be used for clinical research, outcomes analysis, and disease management. DICOM Structured Reporting is a standardized format for medical results. LOGIQ S7 Expert/Pro supports the following templates: •...
  • Page 114: Connectivity Functions

    The ultrasound system is pre-configured for many services, with default settings selected. You can change these services and settings as needed. You must restart the LOGIQ S7 Expert/Pro (shutdown) after making any changes to connectivity CAUTION settings in the Utility menus. This includes any changes on the TCPIP or dataflow setup !! CAUTION: screens.
  • Page 115: Tcpip

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-13-4 TCPIP This configuration category enables users with administrative rights to set the TCPIP for the system and connected remote archive. 1.) Type the name of the Ultrasound system in the Computer Name field.
  • Page 116 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 3-21 IP settings Preset Parameter Description Type the subnet mask address. NOTE: The Subnet Mask is an IP address filter that eliminates Subnet Mask communication/messages from network devices of no interest to your system.
  • Page 117: Device

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-13-5 Device To add a new device, 1.) Press Add. 2.) Type the device name in the Name field. 3.) Type the device's IP address in the IP Address field.
  • Page 118: Service

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-13-6 Service For each Device that you added to the system, you need to set up the service(s) that device supports (you must be an administrator to update these screens).
  • Page 119: Changing Parameters For A Service

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3-13-6-3 Changing parameters for a service There are certain parameters that may need to be set up for each service: Table 3-23...
  • Page 120: Connectivity Setup Worksheet

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-14 Connectivity Setup Worksheet Site System Information Comments: Floor: Site: Dept: Room: LOGIQ™ S7 Type: Expert/Pro REV: CONTACT INFORMATION Name Title...
  • Page 121 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 3-14 Connectivity Setup Worksheet (cont’d) LOGIQ™ S7 Expert/Pro IP Address Local Port Host Name AE Title Net Mask ROUTING INFORMATION GATEWAY IP Addresses...
  • Page 122: Paperwork

    Check that the correct User Manual(s) for the system and software revision, is included with the installation. Specific language versions of the User Manual may also be available. Check with your GE Sales Representative for availability. 3-60 Section 3-15 - Paperwork...
  • Page 123: Overview

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 4 Functional Checks Section 4-1 Overview 4-1-1 Purpose of Chapter 4 This chapter provides procedures for quickly checking major functions of the LOGIQ™ S7 Expert/Pro scanner diagnostics by using the built-in service software, and power supply adjustments.
  • Page 124: General Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 4-3 General Procedure SYSTEM REQUIRES ALL COVERS CAUTION Operate this unit only when all board covers and frame panels are securely in place.
  • Page 125 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 1. LCD Monitor 8. Probe holder (Option) 15. Probe port - 4 active probe ports 2. Touch panel 9. DVD Drive 16. Audio speaker 3.
  • Page 126: Appearance Inspection

    Figure 4-2 Product Labels LCD CAUTION LABEL GENDER CAUTION LABEL (ONLY FOR INDIA, CHINA, KOREA) MULTI CAUTION LABEL LOGIQ S7 Expert/Pro RATING LABEL (FOR CHINA, FOR KOREA, FOR JAPAN) LOGIQ S7 Expert/Pro RATING LABEL UL LABEL 1.) Check presence of LCD Caution Label.
  • Page 127: Mechanical Parts Functional Check

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-2 Mechanical Parts Functional Check Pay attention to moving parts and pinch point when performing mechanical parts functional CAUTION check. !! CAUTION:...
  • Page 128: Main Lcd

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-2-1 Main LCD 1.) Tilt LCD Monitor forward 2.) Verify can be tilted (see: Figure 4-4) Figure 4-4 Main LCD Chapter 4 - Functional Checks...
  • Page 129: Lcd Arm

    150mm 90degree 90degree Figure 4-6 LCD Movement • a, b : LOGIQ S7 Expert • c : LOGIQ S7 Pro 3.) Engage LCD Arm LOCK. 4.) Align LCD Arm along center line of the system. 5.) Verify LCD Panel is locked to position and LCD ARM does not move.
  • Page 130: Opio Swivel Lock

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-2-3 OPIO Swivel Lock Figure 4-7 OPIO Swivel 1.) Push the swivel button and hold it. (see: Figure 4-7) 2.) Rotate OPIO (while holding button).
  • Page 131: Swivel/Brake Lock Caster

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-2-4 Swivel/Brake Lock Caster System weighs approximately 90kg. Pay attention for system stability when working with the CAUTION casters !! CAUTION: NOTE: This procedure applies to RIGHT REAR Caster only.
  • Page 132: Swivel/Brake Lock Caster

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-2-5 Swivel/Brake Lock Caster System weighs approximately 90kg. Pay attention for system stability when working with the CAUTION casters. !! CAUTION: NOTE: This procedure applies to casters other than RIGHT REAR Caster.
  • Page 133: Power On/Off

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-3 Power On/Off After turning off a system, wait at least 10 seconds before turning it on again. NOTICE The system may not be able to boot if power is recycled too quickly.
  • Page 134 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL OFF (Stand-by) Figure 4-11 On/Off Button on Control Panel and indicator color 6.) Verify On/Off button color changes to GREEN. 7.) Verify on-board peripherals power is also turned ON.
  • Page 135: Sleep Mode Check (R1.X.x)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-3-2 Sleep Mode Check (R1.x.x) NOTE: This procedure assumes systems is already up and running. 1.) Press the ON/OFF button (see: Figure 4-12) on the control panel.
  • Page 136: Sleep Mode Check (R2.X.x Or Later)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-3-3 Sleep Mode Check (R2.x.x or later) NOTE: This procedure assumes systems is already up and running. 1.)To going into the sleep mode, enter the scan screen and lightly press the Power On/Off switch (see: Figure 4-13) once.
  • Page 137: Complete Power Down

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-3-4 Complete power down After turning off a system, wait at least 10 seconds before turning it on again. The system may not be NOTICE able to boot if power is recycled too quickly.
  • Page 138: Power Off / Shutdown Check

    Use this button to exit from the System-Exit menu and return to the previous operation. • Exit (Only available when logged in as GE Service with Service Dongle) Select this button when you want to exit to the Windows Desktop.
  • Page 139: System Information

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-4 System Information 4-3-4-1 Software Version 1.) From the touch panel, select Utility -> System -> About. 2.) Note “Software Version” in Software Group.
  • Page 140: Service Platform Confirmation

    ANUAL 4-3-4-2 Service Platform Confirmation 1.) From the touch panel, select Utility -> Service. 2.) On Service Browser, select GE Service, and enter the password. 3.) Note “SvcPform” version. Figure 4-17 Software Platform version Chapter 4 - Functional Checks 4-18...
  • Page 141: Usb Port Test

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-4-3 USB Port Test Required Equipment : Service Dongle External Port Test 1.) Plug service dongle to the following locations. Refer to Section 3-9-2 for locations.
  • Page 142: System Integration Checks

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-5 System Integration Checks 4-3-5-1 OPIO Test Figure 4-18 OPIO 1.) Power ON the system. 2.) Check backlight on AN Keyboard, B Buttons, Freeze Key.
  • Page 143: Dvd Drive Test

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL COMMENT Figure 4-20 6.) Connect any probe. (if not connected) 7.) Go to Live CF Mode. (Press “CF” Mode button, unfreeze.) 8.) Verify both B button and CF button are lit GREEN.
  • Page 144: Peripheral Checks

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-6 Peripheral Checks 4-3-6-1 DVR Test (if equipped) Required: Blank [DVD+RW] media. 1.) Make sure the tab menu of “Video” is found on the right-most touch panel tab. If not found, go to “Utility”...
  • Page 145: Gel Warmer Test

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-6-6 Gel Warmer Test 1.) Insert Gel Bottle into Gel Warmer. 2.) Power ON the warmer. 3.) Verify Gel gets warm. NOTE: Depending on environment, Gel requres 30 minute or more to warm up.
  • Page 146: Color Flow-Mode

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-7-3 Color Flow-Mode 1.) Press CF Mode button. 2.) Increase CF Gain until noise floor shows up. 3.) Verify no vertical “streaks” of color that significantly contrasts the surrounding color.
  • Page 147: B/Cf/Pw Mode

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5.) Revert to original setting in Utility -> Imaging -> General “Simultaneous”. Figure 4-24 Pulse Doppler Mode 4-3-7-5 B/CF/PW Mode 1.) Press B Mode button.
  • Page 148: M-Mode

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-7-6 M-Mode 1.) Press B Mode Button. 2.) Press M Mode Button. 3.) Verify that system scan B/M with the activated probe.
  • Page 149 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Verify BF Mode. Figure 4-28 B-Flow Mode 5.) From the touch panel, select “COLOR” 6.) Verify B-Flow Color is shown. Figure 4-29 B-Flow Color Mode...
  • Page 150: Opio Interface Check

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4-3-8 OPIO Interface Check 1.) Exit to Windows Desktop. Figure 4-30 Windows Desktop 2.) Double Click on icon “ScFpApiTest.exe”. Figure 4-31 FpApiTest.exe screen on desktop...
  • Page 151 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Select “Op-Panel Open”. 4.) Verify OP Panel button operation is reflected on diag window. Figure 4-32 Op Panel interface test screen NOTE: In above example, key “Alt”...
  • Page 152 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Select Diagnostics -> Common diagnostics -> Utilities -> Disruptive Mode Figure 4-33 Diagnostic Mode page screen 3.) Click the "YES" on Diagnostic Mode page...
  • Page 153 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Select Diagnostics -> OP Panel Interface Figure 4-35 OP Panel Interface page screen 5.) Click the "Start" on OP Panel Interface page Figure 4-36 OP Panel Interface page screen 6.) Select "Op-Panel Open"...
  • Page 154: Board Diagnostics

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 4-4 Board Diagnostics GFS- GBF128/ EEPRO Jig youhi CW Bd DC4D Battery GFS Swept Demodulator Test GFS Front-End Interface Test GFS Analog Test...
  • Page 155 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL GFS- GBF128/ EEPRO Jig youhi CW Bd DC4D Battery PC Interactive AVI Playback CD-R Test optonal optonal DVD-R DVD+RW Keyboard Microphone Monitor...
  • Page 156 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. Chapter 4 - Functional Checks 4-34...
  • Page 157: Components And Functions (Theory)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 5 Components and Functions (Theory) Section 5-1 Overview 5-1-1 Purpose of Chapter 5 This chapter explains LOGIQ™ S7 Expert/Pro’s system concepts, component arrangement, and subsystem function.
  • Page 158: General Information

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-2 General Information LOGIQ™ S7 Expert/Pro is a digital beamforming curved-, linear- and phased array ultrasound imaging system. The system supports various operating mode such as •...
  • Page 159: Operator Panel

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-2-2 Operator Panel Figure 5-2 Basic Block diagram of LOGIQ™ S7 Expert/Pro (R1.x.x) Chapter 5 - Components and Functions (Theory)
  • Page 160 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 5-3 Basic Block diagram of LOGIQ™ S7 Expert/Pro (R2.x.x or later) Section 5-2 - General Information...
  • Page 161 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Gel Warmer Left Probe Holder ON/OFF Button Right Probe Holder User Define Keys Freeze Key BT Keys P1 Key A/N Keyboard 3 DLP Probe Connectors,...
  • Page 162: System Options

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-2-3 System Options Figure 5-5 System Options 5-2-4 System Ports USB Ports Rear Panel Figure 5-6 System Ports Section 5-2 - General Information...
  • Page 163 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL System is equipped with two USB ports in front. These are customer use and Service port connected to USB controller via USB Hub.
  • Page 164: Front Usb

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-2-4-1 Front USB Customer use and Service port connected to USB controller via USB Hub. USB 2.0, 480Mbps, 5V 500mA NOTE: Front USBs are non-isolated port and not fit to connect external powered USB Devices.
  • Page 165: Hdmi

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-2-4-3 HDMI NOTE: LOGIQ™ S7 Expert/Pro External display I/O is DVI-D. However, connector shape is HDMI connector. Table 5-4 Pin Out Pin number...
  • Page 166: Rj45

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-2-4-4 RJ45 1000Base-T, 100Base-T, 10Base-T Table 5-5 Network Connector, RJ45 Modular 8 Pin Pin No Output Signal Description ETHER TD Ethernet RD+...
  • Page 167: Power On Sequence

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-3 Power ON Sequence DC-DC Controller High Votage : 24V->+/-HV1, HV2, PMX(+/-100V) 11W+11W+2.5W(14WMax) 12V->24V B/W Printer Relay 12V-> +5V, -5V, Live5V...
  • Page 168 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-6 BIOS Default (R2.x.x or later) Execute Disable Bit Enable Hardware Prefetcher Enable Adj. Cache Line Pref. Enable Intel Virtualization Enable...
  • Page 169 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-6 BIOS Default (R2.x.x or later) Serial Port 4 Enable Configuration Hardware Monitor CPU Fan Fan control Auto External Fan Fan control...
  • Page 170 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-6 BIOS Default (R2.x.x or later) USB Configuration EHCI Unit #1 Enable EHCI Unit #2 Enable USB Per-port Control Disabled PCIExpress...
  • Page 171 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-6 BIOS Default (R2.x.x or later) ASPM Support Disabled PME SCI Enable Hot plug Disabled PCIe Speed Gen1 PCIe Root port 6...
  • Page 172 Boot option #1 Boot option #2 Bios Menu : Security Administrator Password User Password * GE Password : polaris Bios Menu : Save & Exit Save Changes and Exit Discard changes and Exit Save changes and Reset Discard Changes and Reset...
  • Page 173 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-6 BIOS Default (R2.x.x or later) Discard Changes Restore defaults Save as User defaults Restore User defaults Sequence 1.) Circuit breaker ON 2.) AC/DC Unit creates Stand-by Power...
  • Page 174: Software Options

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-4 Software Options 5-4-1 Options Following options are offered for LOGIQ™ S7 Expert/Pro. Table 5-7 Software Options Software Options Description Advanced 3D For R1.x.x only;...
  • Page 175 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-7 Software Options Software Options Description 4D provides continuous, high volume acquisition of 3D images. 4D adds the dimension of “movement” to a 3D image by providing continuous, real-time displays.
  • Page 176: Hardware Options

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-5 Hardware Options 5-5-1 Options Following options are offered for LOGIQ™ S7 Expert/Pro. Table 5-8 Hardware Options Hardware Options Description ECG Option...
  • Page 177 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-8 Hardware Options (Continued) Hardware Options Description Side Probe Holder Extend number of probe holder Probe holder adapter for small probes...
  • Page 178: Regional And Peripheral Options

    LOGIQ¢‚ S7 Expert/Pro Russian Keyboard LOGIQ™ S7 Expert/Pro Swedish Keyboard LOGIQ¢‚ S7 Expert/Pro French Keyboard 5-6-2 Peripheral Options Following options are offered for LOGIQ S7 Expert/Pro. Table 5-10 Peripheral Option Printers Description LS7 Expert/Pro Sony UP-D897 Digital B/W Printer Kit...
  • Page 179 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 5-10 Peripheral Option Printers Description LS7 Expert/Pro Sony UP-D55 Digital Color A5 Printer R1.x.x only; Digital Color A5 Printer Kit (R1.x.x only)
  • Page 180: Mechanical Descriptions

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-7 Mechanical Descriptions 5-7-1 Physical Dimensions Dimentions and Weight • Height: 1,115mm (Minimum), 1,750mm (Max.) • Width: 620mm (Caster), 500mm (Keyboard) •...
  • Page 181: Air Flow Distribution

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-7-4 Air Flow Distribution NOTE: Clean the Air Filter periodically. Through the filter grid on the front of the system, air flow into the LOGIQ™ S7 Expert/Pro scanner. By means of the 1 FAN, air is blown through the nest-box, and the warm air exits the scanner through holes in the left side panel and rear of the system.
  • Page 182: Exiting To Windows Desktop

    5.) On “Maintenance Access” window, type in service password (user name not required). NOTE: Service Password is rotating password that changes on 6 months period. Only the GE service personnel or GE authorized dealers have access to maintenance password. NOTE: Maintenance Password is function of date/time.
  • Page 183 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6.) When entering Windows Desktop, press “Ctrl” and “Esc” at the same time to make windows menu bar to appear. Figure 5-11 Windows Desktop...
  • Page 184: Service Platform

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 5-9 Service Platform 5-9-1 Introduction The Service Platform will increase service productivity and reduce training and service costs. 5-9-2 Access / Security The Service Platform has different access and security user levels.
  • Page 185: How The Customer Enables/Disables Disruptive Mode And Vco

    5-9-2-2 How the Customer enables/disables Disruptive Mode and VCO 1.) If not already in read mode, FREEZE the image. 2.) Move the cursor to the GE icon and press the (= right-click). right trackball key 3.) Select Connect Clinical Lifeline (see: Figure 5-13 below).
  • Page 186: Common Service Desktop (Csd)

    5-10-1 Internationalization The user interface provided by the service platform is designed for GE personnel and as such is in English only. There is no multi-lingual capability built into the Service Interface. There are different possibilities to access the Common Service Desktop and its available features:...
  • Page 187: Csd Top Page

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-2 CSD Top Page 5-10-2-1 General Layout Table 5-11 System Information System ID, Serial, IP Address, InSite StatusÅc.etc Connected Probes Connected and System recognized probe listed...
  • Page 188: Decoding System Temperature And Voltage

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-2-2 Decoding System Temperature and Voltage NOTE: These are reference only values and therefore, parameters are not constant. Some parameters may be added/deleted depending on the system configuration, type, and variation.
  • Page 189 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 5 - Components and Functions (Theory) 5-33...
  • Page 190: Error Logs

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-3 Error Logs When the Error Logs page is selected, different log viewing options are available. Log Viewer is displayed in a separate window.
  • Page 191: Image Quality

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-5 Image Quality In the Image Quality page, you can verify image quality. Figure 5-17 Common Service Desktop - Image Quality NOTE: This page is not populated in this version.
  • Page 192: Configuration

    Figure 5-18 Common Service Desktop - Configuration Remote access is ONLY possible if the service platform is properly configured (either by the user NOTICE or a GE technician at site). Operation see: Section 7-5-6 "CSD: Configuration" on page 7-17. !! NOTICE:...
  • Page 193: Utilities

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-7 Utilities The Utilities page contains a variety of Windows utility tools to indicate the status of the system, in addition to various other tools.
  • Page 194 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Select “Disk Defragmenter” from left pane Figure 5-21 4.) Select Disk Drive to defragment on right (upper) pane 5.) Press “Defragment” to execute defragmentation of the selected disk NOTE: Press “Analyze”...
  • Page 195: Replacement

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5-10-8 Replacement There is no accessible parameters in Replacement page. 5-10-9 In the PM page, information about planned, proactive and preventive maintenance is displayed, as described in Chapter 10 - Care &...
  • Page 196 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 5-40 Section 5-10 - Common Service Desktop (CSD)
  • Page 197: Service Adjustments

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 6 Service Adjustments Section 6-1 Overview 6-1-1 Purpose of Chapter 6 This chapter describes how to test and adjust the mechanical capabilities of a scanner that may be out of specification.
  • Page 198: Lcd Monitor Adjustment

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 6-3 LCD Monitor Adjustment The LOGIQ™ S7 Expert/Pro has a free adjustable LCD monitor in relation to the user interface. The digital control panel is located at the front of the color monitor. It is NOT recommended to change the pre-adjusted settings.
  • Page 199: Brightness/Contrast

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 6-2 LCD Montor Adjustment Setting for LOGIQ™ S7 Main Menu Sub Menu Range Expert/Pro ON/OFF English/German/ French/Spanish/ Italian/Swedish/ Language Chinese/Japanese English...
  • Page 200: Contrast

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6-3-1-2 Contrast 1.) Press the MENU (middle) key of the monitor controls over 10 sec. 2.) Select the Picture -> Contrast by pressing the < LEFT or RIGHT > button to decrease/ increase cursor and the MENU (middle) key.
  • Page 201: Function

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6-3-2 Function Figure 6-4 Function 6-3-2-1 Scale 1.) Press the MENU (middle) key of the monitor controls over 10 sec. 2.) Select the FUNCTION -> Scale by pressing the < LEFT or RIGHT > button to move the cursor and the MENU (middle) key.
  • Page 202: Memory Recall

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 6-5 Information 6-3-2-3 Memory Recall 1.) Press the MENU (middle) key of the monitor controls over 10 sec. 2.) Select the FUNCTION -> MEMORY RECALL by pressing the < LEFT or RIGHT > button to move the cursor and the MENU (middle) key.
  • Page 203: Osd

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6-3-3 Figure 6-6 OSD 6-3-3-1 Language 1.) Press the MENU (middle) key of the monitor controls over 10 sec. 2.) Select the OSD -> LANGUAGE by pressing the < LEFT or RIGHT > button to move the cursor and the MENU (middle) key.
  • Page 204: Exit

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Adjust the HALF TONE by pressing the < LEFT or RIGHT > button to decrease/increase the value.(default : 50) 6-3-4 Exit When finishing the Adjusting Menu, select the EXIT (middle) and press the MENU (middle) key.
  • Page 205: Diagnostics/Troubleshooting

    7-15 7-5-2 How the Customer enables/disables Disruptive Mode and VCO 7-15 7-5-3 GE Icon and Remote Connection 7-15 7-5-4 Customer Granting Full Remote Access Permission to GE Service Technician 7-17 7-5-5 Common Service Desktop (CSD) 7-17 7-5-6 CSD: Configuration 7-18...
  • Page 206: Collect Vital System Information

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 7-2 Collect Vital System Information 7-2-1 Collecting System Information The following information is necessary to properly analyze data or images being reported as a...
  • Page 207: Request For Service (Rfs)

    There are two possibilities to contact GE: • by means of the GE “Remote Status Icon” that is displayed on the bottom of the screen. Move the cursor to the GE icon and press the (= left-click) to display the left trackball key “Request For Service”...
  • Page 208 If the service platform is not configured an Error message is displayed. The request is NOT sent! The request is saved in QUEUE for reviewing and/or resending. Figure 7-5 Contact GE - Queue Section 7-2 - Collect Vital System Information...
  • Page 209: Check Points Voltages

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 7-3 Check Points Voltages 7-3-1 Visual check 7-3-1-1 Common Power Supply (CPS/CPS2) Turn on mains switch and check green LED inside Power.
  • Page 210: Led For Power Status

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-3-2 LED for Power Status LEDs near the front cooling fan can provide information pertaining to the DC power distribution status. 1.) Remove Front Cover (no tools required)
  • Page 211: Voltage Measurement Points

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-3-3 Voltage Measurement Points 1.) Remove Front Cover (no tools required) 2.) Remove CSI Cover 3.) On CSI board, there is 12V measurement point...
  • Page 212: Screen Captures And Logs

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 7-4 Screen Captures and Logs There may be times when the customer or field engineer will want to capture a presentation on the screen so it may be recovered by the OnLine Center.
  • Page 213: System Warning/Errors And Logs

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-4-3 System Warning/Errors and Logs 7-4-3-1 Temperature warning 7-4-3-1-1 Cause CPU Temperature exceeded threshold (95°C). 7-4-3-1-2 What to do 1.Shut down the system and Clean the filter.
  • Page 214: System Voltage Failure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-4-3-3 System voltage failure 7-4-3-3-1 Cause Hardware and/or Voltage error detected. 7-4-3-3-2 What to Do 1.Press OK and reboot the system. 2.If problem persists, then shut down the system, turn OFF breaker and reboot.
  • Page 215: Hardware Configuration Error

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-4-3-4 Hardware configuration error 7-4-3-4-1 Cause Hardware(s) not detected as valid system. System will boot up but will not allow scanning. 7-4-3-4-2 What to do 1.Press OK, and shutdown the system and reboot.
  • Page 216: System Error

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-4-3-5 System Error 7-4-3-5-1 Cause Unrecoverable error occurred and unable to continue scan. 7-4-3-5-2 What to do 1.Press “OK”, turn off the console and reboot.
  • Page 217: Log Message Of System Error

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-4-3-5-3 Log message of System error Log message format “Error ; EAErrorHandler(****); Error from” + ErrorModuleName + ErrorMessage Log message of System error...
  • Page 218 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Example2 Software error Error Module Name “Error ; EAErrorHandler(2284); Error from DFEManager: DFEManager - RunDfe failed:!, DFE ERROR: InPort is not connected (2); device "DelayValues_CW.ceDelay". Severity : NonFatalSW”...
  • Page 219: Remote Access To Service Platform

    If the console is setup to connect to InSite ExC server (refer to Chapter.3 for InSite ExC Configuration), then remote access technology may provide GE technicians the possibility to view the entire customer’s desktop and operation system for diagnostics and trouble shooting.
  • Page 220: Customer Granting Full Remote Access Permission To Ge Service Technician

    Customer Granting Full Remote Access Permission to GE Service Technician 7-5-4-1 If GE Service Technician requests Remote Access Permission If a GE Service technician requests remote access to your LOGIQ™ S7 Expert/Pro scanner, following “InSite Notification” appears on the systems s creen. select [Yes] to enable “Disruptive Mode”...
  • Page 221: Common Service Desktop (Csd)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-5-5 Common Service Desktop (CSD) Refer to Section 3-13 for accessing Common Service Desktop remotely. 7-5-6 CSD: Configuration Refer to Section 3-13 for accessing Common Service Desktop remotely.
  • Page 222: Minimum Configuration To Boot/Scan

    25, 2014) IRECTION EVISION ERVICE ANUAL Section 7-6 Minimum Configuration to Boot/Scan LOGIQ S7 Expert/Pro will require at minimum GBF128/64, GFS Board (with COMExpress and HDD mounted), CPS and OPIO to boot up. 7-18 Section 7-6 - Minimum Configuration to Boot/Scan...
  • Page 223: Troubleshooting Trees, Instructions And Tech Tips

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 7-7 Troubleshooting Trees, Instructions and Tech Tips Table 7-2 below outlines LOGIQ™ S7 Expert/Pro troubleshooting trees and instructions shown in the sub- sections.
  • Page 224: System Does Not Boot Up

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-7-1 System does not boot up System doesn’t boot Verify that Is Standby Switch - Power Cord is connected to Mains illuminated ?
  • Page 225: Noise Disturbs The Image

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-7-2 Noise disturbs the Image Noise is shown in Image. Is there any electrical devices that affect the Scanner? Keep the Scanner away from those electrical devices.
  • Page 226: Trackball - Impaired Sensitivity

    EVISION ERVICE ANUAL 7-7-3 Trackball - Impaired Sensitivity The trackball has low sensitivity! Contact GE service representative Is it difficult to roll the to replace the trackball module. trackball? Set up the appropriate trackball response time (via [Trackball Speed] button in the System Setup - User Settings page).
  • Page 227: Printer Malfunction

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-7-4 Printer Malfunction Printer malfunction Is printer Check all signal and power supply cable properly connections between Printer and Scanner connected? Load paper by using the correct format.
  • Page 228: Monitor Troubleshooting

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7-7-5 Monitor Troubleshooting Fault symptom Check these items Check the power cord is properly connected. Check the video cable is properly connected.
  • Page 229: Overview

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 8 Replacement Procedures Section 8-1 Overview 8-1-1 Purpose of Chapter 8 This chapter contains replacement procedures for different modules and their subsystems.
  • Page 230 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 8-1 Chapter 8 Contents Section Description Page Number 8-11 8-47 Replacement Side B/W Printer Tray 8-12 8-48 Replacement of Monitor and LCD Arm Plastic Covers...
  • Page 231: Returning/Shipping System, Probes And Repair Parts

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-1-2 Returning/Shipping System, Probes and Repair Parts When returning or shipping the LOGIQ™ S7 Expert/Pro system in the original packaging: • system must be lowered to its minimum height with monitor flapped down (see: Figure on...
  • Page 232: Warnings And Important Information

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-2 Warnings and important information 8-2-1 Purpose of this section This section includes important information. Please read it before doing any of the procedures in this chapter.
  • Page 233: Returning/Shipping Probes And Repair Parts

    Equipment being returned must be clean and free of blood and other infectious substances. GE Healthcare policy states that body fluids must be properly removed from any part or equipment prior to shipment. GE Healthcare employees, as well as customers, are responsible for ensuring that parts/ equipment have been properly decontaminated prior to shipment.
  • Page 234: System Software - Installation/Upgrade Procedure

    USER ARCHIVE NOTE: The HDD on LOGIQ S7 Expert/Pro has three visible partitions (two other partitions are hidden). Before loading the Base System Software, you must perform the following steps: • Contact the customer to allow enough lead time to back up the system.
  • Page 235: Preparation For Software Loading

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL • Be sure all USB devices (printers only) are connected and turned on (except for USB Flash devices). If a printer is plugged in at this time, it is automatically installed.
  • Page 236 ERVICE ANUAL General, Record preset Region, Language and Regional Options. (See the latest revision of the LOGIQ S7 Expert/Pro Basic User Manual) 8-3-5-2-1-2 Connectivity - Recording the TCP/IP settings 1.) From the touch panel, select Utility -> Connectivity -> TCP/IP.
  • Page 237 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 8-4 Record settings from WLAN screen PARAMETER VALUE NETWORK NAME (SSID) NETWORK AUTHENTICATION DATA ENCRYPTION NETWORK KEY KEY INDEX Specify Yes/No to “The Key is Provided Automatically”...
  • Page 238 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 8-6 Record settings for each device listed in Device table Name IP Address 8-3-5-2-1-5 Connectivity - Recording Device Destination settings 1.) From the touch panel, select Utility -> Connectivity -> Service.
  • Page 239 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 8-4 Connectivity - Recording Dataflow settings 2.) Record all settings in Table 8-8 for each Destination Device and Service. Table 8-8...
  • Page 240 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 8-5 Connectivity - Recording the Print Key Assignments 2.) Record all settings in Table 8-9. Table 8-9 Record settings from Button screen...
  • Page 241 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Record all settings in Table 8-10. Table 8-10 Record settings from Service screen Destination Services (Types) Properties Name Device Name for each Destination Device...
  • Page 242 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 8-7 Connectivity - Recording Miscellaneous settings 2.) Record all settings in Table 8-11. Table 8-11 Record settings from Miscellaneous screen Patient/Exam Menu Options...
  • Page 243 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 8-11 Record settings from Miscellaneous screen Warn video titles exist in the internal storage Columns in examination listings ON/OFF or Value (Date | Category | Exam Description | Img.
  • Page 244 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Make sure that the Software Option Keys (alphanumeric passwords) have been WARNING WARNING recorded correctly. If the key is incorrect, you will not be able to log on after the !! WARNING: software installation has been completed.
  • Page 245 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Record all settings in Table 8-13. Table 8-13 Record settings from Users screen Group User ID Prefix Last Name First Name...
  • Page 246 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-3-5-2-1-13 System - Recording Peripheral settings 1.) From the touch panel, select Utility -> System-> Peripherals. Figure 8-11 System - Recording Peripheral settings 2.) Record all settings in...
  • Page 247: System Software - Installation Procedure

    Preset Region Language Units 8-3-6 System Software - Installation Procedure Two separate methods of software installation is offered in LOGIQ S7 Expert/Pro. A.) Software Load from DVD B.) Software Re-Load from Repository Situation • Harddrive is new • Re-installation for application or operating system issues •...
  • Page 248 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6.) The notification screen appears, and then hit the Enter key. 7.) Select either [A] “Load the complete disk”, or [B] “Load the bootable C: partition”...
  • Page 249: System Software - Installation Procedure (R2.X.x Or Later)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10.)A message appears, and then presses CTRL-ALT-DEL to reboot the system. 11.)At the first system reboot after loading the software, the OS will reboot the system automatically.
  • Page 250 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL boot. 6.) Press DVD-Drive eject button. 7.) Place the Base System Software DVD on the tray. 8.) Wait few moments for DVD drive to read DVD contents.
  • Page 251 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 11.)A confirmation window appnears, and then hit the Enter key. 12.)Approx. 20 minutes later, loading the software will be completed. 13.)Remove Base System Software DVD, and type "exit" to reboot the system.
  • Page 252 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 17.)Press “F3” and select “Yes” 18.)Go to “Boot” in BIOS menu and check that Boot Option #1 is set to “SATA 2” and Boot Option #2 is set to “SATA 0”...
  • Page 253: Application Software Installation Preparation

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 19.)Go to “Save & Exit” in BIOS menu and select “Save Changes and Exit”. Then, the system will re- boot 8-3-7-3 Application Software Installation Preparation 1.) Verify the system boot up to Windows Screen...
  • Page 254: Application Software Installation

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5.) Select “G” drive in the tree view. 6.) Double click “LoadSoftware.bat”. Software Installation will begin. 8-3-7-4 Application Software Installation 7.) Application Software Installation process will take approximately 10 minutes NOTE: The following message box will appear, but can be ignored.
  • Page 255 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8.) After the installation completes, the system will be rebooted automatically. NOTE: Do not remove Application Software DVD from drive. 9.) About 1 minute later, the following message box will appear. Click “OK” button.
  • Page 256 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10.)If SW License window appears, then enter correct Option String and click “OK”. 8-28 Section 8-3 - System Software - Installation/Upgrade Procedure...
  • Page 257: Software Installation From Common Service Desktop

    ANUAL 8-3-8 Software Installation from Common Service Desktop NOTE: This procedure requires PC with LAN connection capable of reaching to to GE Network 8-3-8-1 Involking Software Reload 1.) Select [SERVICE] from [UTILITY] Menu 2.) Select “GE Service” and enter Password 3.) Verify entering to Common Service Desktop (CSD)
  • Page 258 8.) Note “Challenge String” NOTE: The “Challenge String” is unique to each device. 9.) On PC, connect to GE site by typing the following address in internet browser http://3.62.153.99/service/ 10.)On login screen, Select “GE Service” and Enter Password 11.)On PC screen, Enter “Challenge String” noted earlier from Console...
  • Page 259 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 12.)Press [SUBMIT] on PC side 13.)On PC screen, note “Response String” 14.)Enter “Response String” into Console 15.)Press [SUBMIT RESPONSE] Chapter 8 - Replacement Procedures...
  • Page 260 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 16.)Following screen appears when [Challenge]-[Response] pair is correct 17.)Select [SYSTEM SHUTDOWN] from Common Utility Menu in CSD 18.)Select Restart System Retain Disruptive Mode 19.)Press [SUBMIT]...
  • Page 261: Software And Functional Checks After Installation/Upgrade Procedure

    For the sake of simplicity, we have used the CD-R in the examples in this section. 8-5-1-2 General Notes NOTE: When you restore backup data from the Utility menu, the LOGIQ S7 Expert/Pro application usually restarts automatically when the restoring is complete. NOTE: To perform backup and restore procedures, you must login with administrator privileges.
  • Page 262: Restore Procedures

    6.) Select Restore. The system performs the restore. As it proceeds, status information is displayed on the Backup/ Restore screen. 7.) The LOGIQ S7 Expert/Pro restarts automatically when Restore is done. 8-34 Section 8-5 - Settings - Backup and Restore...
  • Page 263: Ezbackup And Ezmove

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-5-4 EZBackup and EZMove EZBackup or EZMove allows you to manage hard disk space (move images off the hard drive) while maintaining the patient database on the scanner, as well as to back up the patient database and images.
  • Page 264: Option Keys

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-5-5 Option Keys 1.) From the touch panel, select Utility -> System-> Backup/Restore. 2.) ADD or REMOVE new option key at “Enter New Option Key” field.
  • Page 265: Replacement Front Cover

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-6 Replacement Front Cover 8-6-1 Manpower 1 person, 1 min. 8-6-2 Tools None required. 8-6-3 Removal Procedure 1.) Simply pull out Front Cover.
  • Page 266: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-6-4 Installation Procedure 1.) Front cover to be installed in reverse order of removal. 2.) Align the edges and simply push to make sure all three “latches” are engaged.
  • Page 267: Replacement Top Cover

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-7 Replacement Top Cover 8-7-1 Manpower 1 person, 5 min. 8-7-2 Tools None required. 8-7-3 Removal Procedure 1.) Top Cover shapes differ for each configuration, but removal/installation procedure are the same.
  • Page 268: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-7-4 Installation Procedure 1.) Cover to be installed in reverse order of removal. 2.) Make sure “latches” on both sides are engaged.
  • Page 269: Replacement Side Cover

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-8 Replacement Side Cover 8-8-1 Manpower 1 person, 15 min. 8-8-2 Tools Phillips Driver. 8-8-3 Removal Procedure 8-8-3-1 Removal Procedure - Side Cover (L)
  • Page 270 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Pull out Side Cover (L). There is mechanical hook on bottom of the part. Figure 8-23 Pull out Side Cover(L) 8-42...
  • Page 271: Removal Procedure - Side Cover (R)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-8-3-2 Removal Procedure - Side Cover (R) Side Cover (L) Side Cover (R) Figure 8-24 Side Cover (R) 1.) Remove Top Cover.
  • Page 272: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-8-4 Installation Procedure 1.) Cover to be installed in reverse order of removal. 8-8-5 Functional Check Visual inspection only. Test Refer to...
  • Page 273: Replacement Rear Cover

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-9 Replacement Rear Cover 8-9-1 Manpower 1 person, 15 min. 8-9-2 Tools Phillips Driver. 8-9-3 Removal Procedure Figure 8-27 Rear Cover screw 1.) Remove 5 screws.
  • Page 274: Replacement Side Tray & Footrest Cover

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-10 Replacement Side Tray & Footrest Cover 8-10-1 Manpower 1 person, 15 min. 8-10-2 Tools Phillips Driver. NOTE: “Short” type driver preferable for Foot Rest Cover removal/installation.
  • Page 275: Removal Procedure - Foot Rest Cover

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-10-3-2 Removal Procedure - Foot Rest Cover 1.) Remove 3 screws. Figure 8-30 Foot Rest Cover 8-10-4 Installation Procedure 1.) Parts to be installed in reverse order of removal.
  • Page 276: Replacement Of Monitor And Lcd Arm Plastic Covers

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-12 Replacement of Monitor and LCD Arm Plastic Covers Be careful not to fall the monitor during the replacement process. WARNING...
  • Page 277 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Unscrew Monitor Cable Holder Bracket and Cable Clamp. Figure 8-33 3.) Unscrew the Cap-tire screw to disconnect the Monitor cable. Refer to the figure below.
  • Page 278: Monitor Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Unscrew 2 screws, then separate the Monitor from the Monitor Stand Arm. Refer to the figure below. Figure 8-35 Unscrewing 2 screws & separating the Monitor...
  • Page 279: Lcd Arm Plastic Covers Remove Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-12-6 LCD Arm Plastic Covers Remove Procedure 1.) Unscrew 1 screw, than remove 3rd Arm Cover. Refer to the figure below. Figure 8-37 Removing the 3rd Arm Cover 2.) Remove 2nd Axis Cover.
  • Page 280: Lcd Arm Plastic Covers Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Unscrew 1 screw and Remove the 1st Arm Cover. Refer to the figure below. Figure 8-40 Removing the 1st Arm Cover...
  • Page 281: Replacement Of Opio And Related Parts

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-13 Replacement of OPIO and Related Parts NOTE: This section describes removal and installation of OPIO and related parts. OPIO Assy itself, Main PWA Assy, OP LCD Touch Panel Assy, Full A/N Keyboard, TGC Assy, Encoder, Sub Board Assy, Joystick Assy and Trackball Assy.
  • Page 282: Removal Procedure - Trackball Ring

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-1 Removal Procedure - Trackball Ring NOTE: Removal of Trackball Ring and Cleaning does not require tools. 1.) Turn the ring counter clock-wise 2.) Lift trackball from OP Panel...
  • Page 283: Removal Procedure - Op Panel Assy

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-2 Removal Procedure - OP Panel Assy Figure 8-43 OP Panel Assy 1.) Remove 9 screws from bottom of the OPIO Base.
  • Page 284 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Remove 2 screws from rear of the OPIO Base. Figure 8-45 Rear of OPIO Base 3.) Lift OP Panel. Figure 8-46 Lift OP Panel...
  • Page 285: Removal Procedure - An Keyboard

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Remove a connector from location [A]. 5.) Remove a connector from location [B]. 6.) Remove a connector from location [C]. Figure 8-47 Removal of cable & connectors...
  • Page 286: Removal Procedure - Trackball

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-4 Removal Procedure - Trackball 1.) Remove a connector from back of the LCD Touch Panel Assy (marked below). 2.) Remove 2 screws around trackball.
  • Page 287: Removal Procedure - Lcd Touch Panel Assy

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-6 Removal Procedure - LCD Touch Panel Assy 1.) Remove connector (Marked in picture below). 2.) Remove 9 screws around LCD Touch Panel Assy.
  • Page 288: Removal Procedure - Op Panel Main Pwa Assy

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-8 Removal Procedure - OP Panel Main PWA Assy 1.) Remove knobs from OPIO. 2.) Disconnect 7 cables marked below. 3.) Remove 27 screws around PWA board.
  • Page 289: Removal Procedure - Joystick And Encoder On Pwa

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-3-10 Removal Procedure - Joystick and Encoder on PWA 1.) Remove OP Panel Main PWA Assy. 2.) Remove “Rubber Sheet” on reverse side of PWA Assy.
  • Page 290: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-13-4 Installation Procedure 1.) Parts to be installed in reverse order of removal. Note: Make sure to assemble “Rubber Sheet” securely to PWA...
  • Page 291: Replacement Of The Caps For Hardkeys

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-14 Replacement of the Caps for Hardkeys 8-14-1 Manpower One person, approx. 1 minute/cap. 8-14-2 Tools Small-sized slotted screwdriver or tweezers.
  • Page 292 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-14-5 Functional Check Test Refer to Note System Exterior Visual Check Section 4-3-1 Control panel visual check only Power On/Off Section 4-3-3...
  • Page 293: Replacement Of Key Caps (By Special Native Language Keys)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-15 Replacement of Key Caps (by special native language keys) 8-15-1 Manpower One person, 30 minutes. 8-15-2 Tools • Regular Phillips screwdriver •...
  • Page 294: Replacement Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-15-4 Replacement Procedure 1.) Remove ‘1’ Key, ‘Left Shift’ Key, ‘F5’ Key, ‘6’ Key, ‘7’ Key, ‘H’ Key, ‘J’ Key, ‘N’ Key, ‘Back space.’...
  • Page 295 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Pullout AN Key Top Assy from Keyboard Fame. Refer to below Figure 8-62. Figure 8-62 Pullout the AN key Top Assy 4.) Replace old AN Key Top ASSY with the New AN Key Top ASSY.
  • Page 296: Replacement Checking Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-15-5 Replacement Checking Procedure 1.) Insert service dongle key in the front USB port of OPIO. 2.) Push the power button and then click the exit.
  • Page 297 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-15-6 Functional Check Test Refer to Note System Exterior Visual Check Section 4-3-1 Control panel visual check only Power On/Off Section 4-3-3...
  • Page 298: Replacement Around Acq Box

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-16 Replacement around ACQ Box NOTE: This section covers replacement of components related to ACQ Box - PIL assy, GBF128/64 Assy, GFS Assy, SOM w/ 4GB Memory, HDD, CPS Assy, CBP Assy, CSI Assy, FAN w/ Cable Assy, DC4D Assy and DVR Assy (for R1.x.x only)
  • Page 299 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Disconnect Cables. Figure 8-69 Disconnecting Cables (Signals) Description Note To main LCD OPIO To OPIO power and signal USB 2.0 For Color/BW Printers, VNAV...
  • Page 300: Removal Procedure - Pil, Gbf128/64, Gfs, Cps

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-4 Removal Procedure - PIL, GBF128/64, GFS, CPS NOTE: Before working with ACQ box, be sure to remove probes. 1.) Get ACQ Box access.
  • Page 301 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL GBF 128/64, PIL Inner shape of CPS Figure 8-72 Pulling out the board Assy Chapter 8 - Replacement Procedures 8-73...
  • Page 302: Removal Procedure - Separating Pil And Gbf128/64

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-4-1 Removal Procedure - Separating PIL and GBF128/64 1.) Unscrew 6 (six) screws and separate PIL Assy from the GBF128/64. Figure 8-73 Separating PIL Assy from the GBF128/64 NOTE: Stacking Connector is designed to always remain on PIL side.
  • Page 303: Removal Procedure - Separating Cwd Board From Pil

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-4-2 Removal Procedure - Separating CWD Board from PIL Philips screw-driver required to add this part. 1.) Locate CW connector on PIL Board.
  • Page 304 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Therefore, when GFS board is replaced, EEPROM must be moved from one board to the other to make sure vital information pertaining to the system is transferred to the new GFS board.
  • Page 305: Separating Dc4D Unit From Cps (For R1.X.x Only)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Alignment marks U32 Socket (print mark) Figure 8-78 Aligning EEPROM NOTE: Inserting EEPROM in wrong direction will result in permanent damage to EEPROM or to GFS board.
  • Page 306 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Unscrew five screws to remove the DC4D from the CPS Assy. Figure 8-80 Removing the DC4D from the CPS Assy 8-78...
  • Page 307: Installation Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-5 Installation Procedure 8-16-5-1 Installation Procedure - PIL, GBF128/64, GFS/GFS-T6, CPS/CPS2 1.) Parts to be installed in reverse order of removal. NOTE:...
  • Page 308: Removal Procedure - Acq Box

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-7 Removal Procedure - ACQ Box 1.) Remove Front Cover. 2.) Remove 2 screws. Refer to the figure below. Figure 8-81 Removing the front cover 3.) Get access to ACQ Box (Refer to Section 8-18-3 ACQ Box Access).
  • Page 309: Removal Procedure - Separating Cbp (Backplane)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-16-7-1 Removal Procedure - Separating CBP (Backplane) 1.) Disconnect CIS Connector. Refer to Section 8-18. 2.) Disconnect Fan Connector. 3.) Remove 8 Screws.
  • Page 310: Replacement Of Cps2 (R2.X.x Or Later)

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-17 Replacement of CPS2 (R2.x.x or later) 8-17-1 Manpower One person, 15 minutes for each parts. 8-17-2 Tools Phillips screwdriver. 8-17-3 Removal Procedure 1.) Remove Rear Cover.
  • Page 311 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Disconnect Cable. Figure 8-85 Disconnecting Cables (Signals) Description Note To main LCD OPIO To OPIO power and signal USB 2.0 For Color/BW Printers, VNAV...
  • Page 312: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6.) Draw CPS2 out of the NEST Frame. Figure 8-86 Separating CPS/CPS2 Assy 8-17-4 Installation Procedure 1.) Parts to be installed in reverse order of removal.
  • Page 313: Replacement Of Csi And Fan

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-18 Replacement of CSI and Fan 8-18-1 Manpower One Person, 10 minutes. 8-18-2 Tools Standard Phillips Screwdriver. 8-18-3 Removal Procedure 1.) Remove Front Cover.
  • Page 314 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Unscrew four screws to separate CSI Assy Figure 8-89 Removing the CSI Assy 5.) Remove 4 screws from Indicator Bracket. Figure 8-90 Fan Bracket...
  • Page 315: Installation Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 6.) Unscrew four screws to separate FAN Assy. Figure 8-91 Removing the FAN Assy 8-18-4 Installation Procedure 1.) Parts to be installed in reverse order of removal.
  • Page 316: Replacement Of Rear Handle

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-19 Replacement of Rear Handle 8-19-1 Manpower 1 person, 10 min. 8-19-2 Tools Hex Wrench. 8-19-3 Removal Procedure 1.) Remove Rear Cover.
  • Page 317: Replacement Of Caster

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-20 Replacement of Caster 8-20-1 Manpower Two persons, 15 minutes. 8-20-2 Tools 24mm Socket or adjustable wrench. 8-20-3 Removal Procedure Swivel Lock...
  • Page 318: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Remove Nut Figure 8-95 Removable Nut of Caster Proceed with extreme caution and avoid system tipping over. The system weighs approximately CAUTION 90kg.
  • Page 319: Replacement Of The Som,Hdd And Dvr

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-21 Replacement of the SOM,HDD and DVR 8-21-1 Manpower One person, 5 minutes. 8-21-2 Tools Phillips screwdriver. 8-21-3 Removal Procedure (R1.x.x) 1.) Unscrew eight screws and separate Hard disk from the GFS.
  • Page 320 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Remove the bracket from the hard disk. Figure 8-97 Removing the hard disk bracket 3.) Unscrew five screws to separate SOM board from the GFS.
  • Page 321 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 8-100 Push the SOM in both side of GFS Assy 5.) Unscrew four screws to separate DVR from the GFS.. Figure 8-101 Drawing up the DVR...
  • Page 322: Removal Procedure (R2.X.x Or Later)

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-21-4 Removal Procedure (R2.x.x or later) 1.) Unscrew eight screws and separate Hard disk from the GFS T6. NOTE: (*1) When screws removed, Hard Disk Drive is connected to GFS T6 board only by its connector.
  • Page 323 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Unscrew five screws to separate SOM board from the GFS T6. Figure 8-104 Unscrewing to separate SOM board 4.) Pull the SOM board out of the GFS T6..
  • Page 324: Note On Som Replacement

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-21-5 Note on SOM Replacement SOM retains its memory (date/time/BIOS setting) by receiving 3V from coin battery on GFS Board. Therefore, when SOM and GFS boards are separated, SOM is likely to loose its date/time setting. In such case, follow the procedure outlined below to update BIOS information.
  • Page 325: Note On Eeprom Replacement

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-21-6 Note on EEPROM Replacement NOTE: EEPROM on GFS contains VPD (Vital Product Data) for SOM (ComExpress). This VPD contains information such as system serial number to be matched to Software License Key, or Board revision number that is cross referenced to system software for compatibility.
  • Page 326: Accessing Eeprom

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-21-6-1 Accessing EEPROM 1.) Exit to Desktop (Refer to Section 5-11 for details). 2.) Start VPD Editor by clicking icon on the desktop “vpdedit”.
  • Page 327: Initializing Eeprom

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Note : If following windows pop up without changing any parameters (i.e. simply browsing through the VPD contents), then press “NO” before continuing.
  • Page 328 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Click “OPEN” to load selected file. NOTE: This procedure can be applied to other components with VPD, for example, for initializing PIL VPD.
  • Page 329: Modifying Contents Of Eeprom (Comexpress Vpd)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-21-6-3 Modifying contents of EEPROM (ComExpress VPD) System reads off Console Serial Number from EEPROM and match that against installed option string.
  • Page 330: Replacement Of Battery (R2.X.x Or Later)

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-22 Replacement of Battery (R2.x.x or later) 8-22-1 Manpower One person, 1 hour. NOTE: LS7 BUM 5492864 provides the information, "Refreshing the battery" in Chapter 3 Preparing to System for Use.
  • Page 331: Installation Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Disconnect two harnesses from Polaris-Battery. 4.) Unfasten "BOB Assy" using "SCREW WH M3X6" 8-22-4 Installation Procedure Install components in reverse order of removal.
  • Page 332: Replacement Of 4D

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-23 Replacement of 4D 8-23-1 Manpower One person, 1 hour. 8-23-2 Tools Phillips screwdriver. 8-23-3 Removal Procedure 1.) Remove Rear Panel.
  • Page 333: Replacement Of Wireless Lan

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-24 Replacement of Wireless Lan 8-24-1 Manpower One person, 1 hour. 8-24-2 Tools Phillips screwdriver. 8-24-3 Removal Procedure 1.) Remove Rear Cover Polaris Assy and Cable Cover BRKT Polaris.
  • Page 334: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Disassemble Wireless Lan BRKT Assy. 8-24-4 Installation Procedure 1.) Install components in reverse order of removal. 2.) Assemble Wireless Lan Label.
  • Page 335: Replacement Of The Harness Cable, Cable Duct, And Opio Cable Assy

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-25 Replacement of the Harness Cable, Cable Duct, and OPIO Cable Assy 8-25-1 Manpower One person, 45 minutes. 8-25-2 Tools Phillips screwdriver, Nipper, Allen.
  • Page 336 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5.) Unscrew 1 screw to remove Frame rear cover and Push connector toward OPIO base. Figure 8-115 8-108 Section 8-25 - Replacement of the Harness Cable, Cable Duct, and OPIO Cable Assy...
  • Page 337: Removal Procedure (Monitor Cable And Opio Cable)

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-25-5 Removal Procedure (Monitor Cable and OPIO Cable) 1.) Unscrew 1 screw to remove F-Block Bottom Cover. 2.) Remove Cable tie and Gently push connector through center of OPIO base.
  • Page 338 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Remove the Elevation cable duct cover. Figure 8-116 Remove the Elevation cable duct cover 5.) Remove Cable clamp. Figure 8-117 Remove the cable clamp 6.) Pull out the Cable through the hole.
  • Page 339 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7.) Pull out the Cable through the hole. Figure 8-119 Pull out the Cable through the hole 8.) Unscrew five screws to separate back panel and disconnect Main power cable.
  • Page 340: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10.)Disconnect Harness cables. Figure 8-122 Disconnecting Harness cables 8-25-6 Installation Procedure Install components in reverse order of removal. 8-25-7 Functional Check Power ON and Visual check on LCD monitor to see if image displays without color distortion.
  • Page 341: Replacement Of Peripheral Cable Assy

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-26 Replacement of Peripheral Cable Assy 8-26-1 DVD Cables 8-26-1-1 Manpower One person, 45 minutes. 8-26-1-2 Tools Phillips screwdriver. 8-26-1-3 Removal Procedure 1.) Remove the DVD cable connector from the rear panel.
  • Page 342: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Pull up the DVD cables from the gap. Figure 8-125 Pull up DVD cables 4.) Remove the DVD cable connectors from the DVD drive.
  • Page 343: B/W And Color Printer Cables

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-26-2 B/W and Color Printer Cables 8-26-2-1 Manpower One person, 1 hour. 8-26-2-2 Tools Phillips screwdriver. 8-26-2-3 Removal Procedure 1.) Remove the printer USB cable connector from the rear panel.
  • Page 344: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Remove the printer USB and power cable connectors from the printer. Figure 8-129 Remove printer USB and power cable connectors from printer...
  • Page 345: Replacement Of The Probe Holder(Kit), Cable Hook, And Extended Probe Holder

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-27 Replacement of the Probe Holder(Kit), Cable Hook, and Extended probe holder Replacement of Circle Key Caps only 8-27-1 Manpower 1 person, 5 minutes.
  • Page 346: Removal Procedure - Cable Hook

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL NOTE: Small Probe Cup is used for “small probe” such as L8-18i-D. For regular size probe such as 9L-D, Small Probe Cup is not required.
  • Page 347: Removal Procedure - Extended Probe Holder

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-27-3-3 Removal Procedure - Extended Probe Holder 1.) Remove two screws from Cable Hook as shown below Figure 8-134 Probe Holder 8-27-4 Installation Procedure 1.) Parts to be installed in reverse order of removal.
  • Page 348: Replacement Of Ac Cable Holder

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-28 Replacement of AC Cable Holder 8-28-1 Manpower 1 person, 5 minutes. 8-28-2 Tools Phillips Screwdriver. 8-28-3 Removal Procedure 1.) Remove 2 Screws from AC Cable Holder...
  • Page 349: Replacement Of The Peripherals

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-29 Replacement of the Peripherals 8-29-1 B/W Printer 8-29-1-1 B/W on Mid/High Console 8-29-1-1-1 Manpower 1 person, 1 hour 8-29-1-1-2 Tools...
  • Page 350: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Remove cables from printer. 4.) Unscrew four screws and pull put the B/W printer. Figure 8-138 Remove BW printer 8-29-1-1-4 Installation Procedure 1.) Parts to be installed in reverse order on removal.
  • Page 351: B/W On Side Cabinet

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-29-1-2 B/W on Side Cabinet 8-29-1-2-1 Manpower 1 person, 1 hour 8-29-1-2-2 Tools Philips screwdriver 8-29-1-2-3 Removal Procedure 1.) Remove the B/W printer cables from the system (Refer to Section 8-26-2).
  • Page 352: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-29-1-2-4 Installation Procedure 1.) Parts to be installed in reverse order on removal. 8-29-1-2-5 Functional Check Test 1: Print an image by pressing the button P1 on the operator panel. (P1 key should be configured correctly) Verify1: - Verify that image can be printed.
  • Page 353: Installation Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Draw the cabinet out of the frame Assy. Figure 8-142 Drawing out the Mid cabinet 3.) Remove the Color printer cables. (refer to...
  • Page 354: Color Printer On High Console

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-29-2-2 Color printer on High Console 8-29-2-2-1 Manpower 1 person, 1 hour 8-29-2-2-2 Tools Philips screwdriver 8-29-2-2-3 Removal Procedure 1.) Pull out the cabinet.
  • Page 355: Installation Procedure

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Unscrew four screws and pull put the Color pritner Figure 8-146 Remove the Color printer 8-29-2-2-4 Installation Procedure 1.) Parts to be installed in reverse order on removal.
  • Page 356: Replacement Of The Speaker

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-30 Replacement of the Speaker 8-30-1 Man Power 1 person, 10 minutes. 8-30-2 Tools Phillips Screwdriver. 8-30-3 Removal 1.) Remove Rear Cover.
  • Page 357: Installation

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7.) Remove 1 screw holding Ground Wire (Green Wire). 8.) Remove 8 screws holding Speaker Figure 8-149 Removal screws of Speaker 8-30-4 Installation 1.) Parts to be installed in reverse order of removal.
  • Page 358: Replacement Of The Monitor Arm

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-31 Replacement of the Monitor Arm NOTE: Some of the photos resembles consoles other than LOGIQ™ S7 Expert/Pro, but Monitor Arm is common part as Polaris Platform and therefore, service procedure is the same.
  • Page 359 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Remove the 2nd arm cover. Refer to the figure below. Figure 8-152 Removing the 2nd arm cover 4.) Remove 1 cable tie. Refer to the figure below.
  • Page 360 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 5.) Unscrew 1 screw and remove the 1st arm cover polaris. Refer to the figure below. Figure 8-154 Removing the 1st arm cover 6.) Remove 1 cable ties and extract the monitor cable through the hole.
  • Page 361 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8.) Pull out the Monitor Arm. Refer to the figure below. Figure 8-157 Pulling out the Monitor Arm Chapter 8 - Replacement Procedures...
  • Page 362: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-31-4 Installation Procedure In reverse order of removal. 8-31-5 Functional Check 1.) Visually check LCD monitor. 2.) LCD Arm Test. (Refer to Section 4-3-2-2).
  • Page 363: Replacement Of Drawer

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-34 Replacement of Drawer 8-34-1 Drawer on Mid Cabinet 8-34-1-1 Manpower 1 person, 30 minutes 8-34-1-2 Tools Philips screwdriver 8-34-1-3 Removal Procedure 1.) Pull out the cabinet.
  • Page 364: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Unscrew four screws and pull put the Drawer. Figure 8-160 Removal Drawer 8-34-1-4 Installation Procedure 1.) Parts to be installed in reverse order on removal.
  • Page 365: Drawer On High Cabinet

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 8-34-2 Drawer on High Cabinet 8-34-2-1 Manpower 1 person, 30 minutes 8-34-2-2 Tools Philips screwdriver 8-34-2-3 Removal Procedure 1.) Pull out the cabinet.
  • Page 366: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 3.) Unscrew four screws and pull put the Drawer. Figure 8-163 Removal Drawer 8-34-2-4 Installation Procedure 1.) Parts to be installed in reverse order on removal.
  • Page 367: Replacement Ecg Module

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-35 Replacement ECG Module 8-35-1 ECG Module 8-35-1-1 Manpower 1 person, 30 minutes 8-35-1-2 Tools Phillips Driver 8-35-1-3 Removal Procedure 1.) Loosen the 1 screw and remove frame rear cover Figure 8-164 Remove frame rear cover 2.) Loosen the 4 screw and dis-connect ECG cable.
  • Page 368: Replacement Of Uvc

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-36 Replacement of UVC 8-36-1 UVC ASSY 8-36-1-1 Manpower 1 person, 1 hour 8-36-1-2 Tools Standard Field Service Tool kit 8-36-1-3 Removal Procedure 1.) Remove HDMI CABLE form console.
  • Page 369 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 4.) Remove the AC Y-CABLE PYXIS power connector from adaptor. 5.) Remove the power connector from UVC module and remove the AC/DC adaptor FIX bracket with 4 screws.
  • Page 370: Installation Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 7.) Remove UVC module with the UVC FIX BRKT from UVC frame BRKT LS7/S8 with 4 screws. 8.) Remove 2 UVC FIX BRKT from UVC module with 4 screws.
  • Page 371: Replacement Of Dcwd Assy

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 8-37 Replacement of DCWD ASSY 8-37-1 DCWD ASSY 8-37-1-1 Manpower One person, 15 minutes 8-37-1-2 Tools Philips screw driver 8-37-1-3 8-32-1-3 Removal procedure 1.) Remove PIL ASSY, refer to...
  • Page 372 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 8-144 Section 8-37 - Replacement of DCWD ASSY...
  • Page 373: Overview

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 9 Renewal Parts Section 9-1 Overview 9-1-1 Purpose of Chapter 9 This chapter gives an overview of replacement parts available for the LOGIQ™ S7 Expert/Pro..
  • Page 374: List Of Abbreviations

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-2 List of Abbreviations Alternating Current Analog to Digital Converter A/N Key Alpha Numeric Keys ASIC Application Specific Integrated Circuit Assy...
  • Page 375: Parts List Groups

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-3 Parts List Groups Figure 9-1 Console Views Table 9-2 Mechanical and user accessible parts Item Part Group Name Table Number...
  • Page 376: Plastics Covers (Front/Sides/Rear)

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-4 Plastics Covers (Front/Sides/Rear) Figure 9-2 Plastics Covers (Front/Side/Rear) Table 9-3 Plastic Covers Item Part Name Part Number Description Plastics, DLP Probe Connector LOGIQ™ S7 Expert/...
  • Page 377: Lcd Monitor

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-5 LCD Monitor 201 203 Figure 9-3 LCD Monitor Table 9-4 User Interface Item Part Name Part Number Description 5392293 Full 19" monitor set with hinge 19"...
  • Page 378: Opio

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-6 OPIO Figure 9-4 OPIO Section 9-6 - OPIO...
  • Page 379 KEY LED COLOR MODE MIZAR KEY LED X MIZAR NURI FRAME REAR COVER NURI KEYBOARD LOWER NURI KEYBOARD UPPER LOGIQ S7 EXPERT LCD PANEL NURI TGC MESH NURI USB COVER NURI LCD SPONGE LOGIQ S7 PRO LCD PANEL 313 314...
  • Page 380 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Mode Switch/Encoder Parts Location Joystick Section 9-6 - OPIO...
  • Page 381 5453468 OP Sub Board Assembly Sub Board Assy 5455115 OP TGC Assembly TGC Assy 5393439 Trackball Assy (Use Cable for LOGIQ S7 Expert/Pro) Trackball Assy 5448987 OP Button+Knob+Slidepot sets OPIO Button Sets 5448176 OP Mode select Encoder 1pcs Mode Select Switch & Encoder...
  • Page 382 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-5 OPIO Item Part Name Part Number Description 5480207 A/n Key Top w/Base Assembly, Norwegian A/N KEY CAP - Norwegian 5480210...
  • Page 383: Nest Box Parts

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-7 Nest Box Parts Figure 9-5 Nest Box Parts Chapter 9 - Renewal Parts 9-11...
  • Page 384 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 9-6 Nest Box Parts(Cont‘) 9-12 Section 9-7 - Nest Box Parts...
  • Page 385 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-6 Nest Box Parts Item Part Name Part Number Description 5422540 Probe Interface Board PIL Assy 5392446 With CWD GBF128 Assy (R1.x.x)
  • Page 386: Mechanical Parts

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-8 Mechanical Parts Figure 9-7 Mechanical Parts 9-14 Section 9-8 - Mechanical Parts...
  • Page 387: Mechanical Parts

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-7 Mechanical Parts Item Part Name Part Number Description 5400391 except swivel lock caster (3 per console) Total Lock Caster 5400392...
  • Page 388: Options

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-9 Options Figure 9-8 9-16 Section 9-9 - Options...
  • Page 389 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-8 Options Item Part Name Part Number Description 5418150 Probe Holder Extention Optional Probe Holder 5380960 3 button switchs operated by the feet...
  • Page 390: Optional Peripherals And Accessories

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-10 Optional Peripherals and Accessories 9-10-1 Printers Figure 9-9 Optional Peripherals and Accessories - Printers Table 9-9 Optional Peripherals and Accessories - Printers...
  • Page 391: Probes

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-11 Probes Table 9-10 Probe Description Part Part Probe Name Illustration Application Feature Number Type Qty FRU Abdomen, C1-5-D B, CHI, CF, PDI, M, PW, B-Flow, B-Flow...
  • Page 392 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-10 Probe Description Part Part Probe Name Illustration Application Feature Number Type Qty FRU 9L-D Abdomen, Small B, CHI, CF, PDI, M, PW, B-Flow, B-Flow...
  • Page 393 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-10 Probe Description Part Part Probe Name Illustration Application Feature Number Type Qty FRU S4-10-D Pediatrics, B, CHI, CF, PDI, M, MCF, PW, CW, TVI,...
  • Page 394: Power Cord

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 9-12 Power Cord Figure 9-10 Power Cord 9-22 Section 9-12 - Power Cord...
  • Page 395 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 9-11 Power Cord Item Part Name Part Number Description Power Cord for 5182611 Power Cord for India India Power Cord for...
  • Page 396 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL This page was intentionally left blank. 9-24 Section 9-12 - Power Cord...
  • Page 397 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Chapter 10 Care & Maintenance Section 10-1 Overview 10-1-1 Periodic Maintenance Inspections It has been determined by engineering that your LOGIQ™ S7 Expert/Pro system does not have any high wear components that fail with use, therefore no Periodic Maintenance Inspections are mandatory.
  • Page 398: Why Do Maintenance

    Testing results, corrective action and the effects of corrective action must be documented and maintained on the site. Your GE service representative can help you with establishing, performing and maintaining records for a quality assurance program. Please contact us for coverage information and/or price for service.
  • Page 399 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-3-1 How often should care & maintenance tasks be performed? (cont’d) Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Recommend to clean filter at •...
  • Page 400 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 10-2 Customer Care Schedule Item Service at Indicated Time Notes Also after corrective Surface Probe Leakage • maintenance or as required by Current Checks your facilities QA program.
  • Page 401: Tools Required

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 10-4 Tools Required 10-4-1 Special Tools, Supplies and Equipment 10-4-1-1 Specific Requirements for Care & Maintenance Table 10-3 Overview of Requirements for Care & Maintenance...
  • Page 402: System Maintenance

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 10-5 System Maintenance 10-5-1 Preliminary Checks The preliminary checks take about 15 minutes to perform. Refer to the system user documentation whenever necessary.
  • Page 403: Functional Checks

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-5-2 Functional Checks The functional checks take about 60 minutes to perform. Refer to the LOGIQ™ S7 Expert/Pro Basic User Manual whenever necessary.
  • Page 404: Peripheral/Option Checks

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-5-2-2 Peripheral/Option Checks If any peripherals or options are not part of the system configuration, the check can be omitted. Table 10-6...
  • Page 405: Air Filter Cleaning

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-5-4-2 Air filter cleaning Clean the system's air filters to ensure that a clogged filter does not cause the system to overheat and reduce system performance and reliability.
  • Page 406: Physical Inspection

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-5-5 Physical Inspection Table 10-9 Physical Checks Step Item Description Labeling Verify that all system labeling is present and in readable condition.
  • Page 407: Probe Maintenance

    The Basic User Manual and/or care card provides a complete description of probe care, maintenance, cleaning and disinfection. Ensure that you are completely familiar with the proper care of GE probes. Ultrasound probes can be easily damaged by improper handling. Review the Basic User Manual of LOGIQ™...
  • Page 408: System Software Updates(Software Download)

    After you press a software download button, it may take up to three (3) seconds for the system to respond. NOTE: Software Updates through the GE service platform may not be available in all countries. 10-12 Section 10-6 - System Software Updates(Software Download)
  • Page 409: Software Download And Installation

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-6-1 Software download and installation You receive notification of available software updates via the envelope icon that appears in the status bar on the bottom of the monitor.
  • Page 410 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL NOTE: If you are not logged in as an Administrator and there is a software package to download, you will see the following dialog when shutting down the system: Figure 10-4 Software Download Dialog If Not an Administrator If you get a dialog with “Unknown status,”...
  • Page 411 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 2.) Press Download. Figure 10-6 System - Exit Dialog Window The download process starts. The progression of the download process is displayed.
  • Page 412 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Press Download. Figure 10-7 Download in Progress Options • Stop: Stops the download process; no software upgrade is performed. If you select Stop, you WILL NOT be given an opportunity to download this package again.
  • Page 413 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Figure 10-9 Paused Software Download • Resume: Press Resume to continue the software download. 3.) If auto-install was not checked, you will see the following dialog when the software download has finished (the message, “SW is ready to be installed,”...
  • Page 414 New Software Verification !! CAUTION: checklist. These verification results are tracked for regulatory purposes, sent back to GE for tracking, and approved with your signature. 10-18 Section 10-6 - System Software Updates(Software Download)
  • Page 415 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL NOTE: Press the Question mark to get information on how to check each feature. Figure 10-12 New Software Verification Checklist NOTE: You can move this dialog, but DO NOT close it. You can also minimize this dialog by clicking on the “<<“...
  • Page 416 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL When there is a failed feature noted on the New Software Verification Checklist, the following Warning dialog displays: Figure 10-14 Re-Install Previous Software Version Dialog When the New Software Verification Checklist contains a ‘Failed’...
  • Page 417: Using A Phantom

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 10-7 Using a Phantom Refer to the User Manual of the Phantom for information on using a phantom and quality assurance tests.
  • Page 418: Gehc Leakage Current Limits

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-2 GEHC Leakage Current Limits The following limits are summarized for UL60601-1 (For USA), IEC 60601-1 Medical Equipment Safety Standards and IEC 62353 Medical Electrical Equipment - Recurrent test and test after repair of medical electrical equipment.
  • Page 419 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Table 10-14 Type BF Patient Auxiliary Limits Country Conditions Others Close Protective earth and close neutral with normal polarity 0.1mA* 0.1mA*** Close Protective earth and close neutral with reverse polarity 0.1mA*...
  • Page 420: Outlet Test - Wiring Arrangement - Usa & Canada

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-3 Outlet Test - Wiring Arrangement - USA & Canada Test all outlets in the area for proper grounding and wiring arrangement by plugging in the neon outlet tester and noting the combination of lights that are illuminated.
  • Page 421: Grounding Continuity

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-4 Grounding Continuity Electric Shock Hazard! CAUTION The patient or operator MUST NOT come into contact with the equipment during this test...
  • Page 422: Enclosure Leakage Current Test

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-5 Enclosure Leakage Current Test 10-8-5-1 Definition This test measures the current that would flow in a grounded person who touched accessible non- protectively-earthed parts of equipment if the ground wire were to break.
  • Page 423: Data Sheet For Enclosure Leakage Current

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL When using Dale 600/601 or a similar test instrument, its power plug may be inserted into the wall outlet and the equipment under test is plugged into the receptacle on the panel of the meter. This places the meter in the grounding conductor and the current flowing from the case to ground will be indicated in any of the current ranges.
  • Page 424: Generic Procedure

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-6-2 Generic Procedure The testing meter is connected from accessible protectively-earthed parts of the equipment to ground. Measurements should be made with power line polarity normal and reversed, and with the neutral open and closed.
  • Page 425: Type Bf Patient Leakage Current Test - Probe

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-7 Type BF Patient Leakage Current Test - Probe 10-8-7-1 Definition This test measures the current that would flow to ground from any of the probes through a patient who is being scanned and becomes grounded by touching some other grounded surface.
  • Page 426: Type Bf Patient Leakage Current - Mains To Probe

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-8-8 Type BF Patient Leakage current - Mains to Probe Reference the procedure in the IEC 60601-1. Measure leakage current flow from mains to Probe.
  • Page 427: When There's Too Much Leakage Current

    GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Section 10-9 When There’s Too Much Leakage Current... 10-9-1 Earth and/or Enclosure Fails Check the ground on the power cord and plug for continuity. Ensure the ground is not broken, frayed, or intermittent.
  • Page 428: New Unit

    GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL 10-9-5 New Unit If the leakage current measurement tests fail on a new unit and if situation cannot be corrected, submit a Safety Failure Report to document the system problem. Remove unit from operation.
  • Page 429 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL Type BF Patient Leakage Current Test - Probes (This test should be done for all probe) Table 10-23 Tester Tester Polarity Ground...
  • Page 430 GE H RAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL ULTRASOUND INSPECTION CERTIFICATE Customer Name: System ID: Dispatch Number / Date Performed: Warranty/Contract/HBS System Type Model Number: Serial Number: Manufacture Date: Probe 1:...
  • Page 431 GE H DRAFT LOGIQ™ S7 E EALTHCARE XPERT 5460683, R DRAFT (J 25, 2014) IRECTION EVISION ERVICE ANUAL ELECTRICAL SAFETY Max Value Value Electrical Test Performed Allowed Measured Comments Outlet (correct ground &wiring config.) System Ground Continuity Chassis Source Leakage Current - Probe...
  • Page 432 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL INDEX Acoustic Noise Output, 2-6 Functional Checks, 4-1 Adjustment Required Equipment, 4-1 LCD Monitor, 6-2 Printer Settings, 3-34 Air Filter Cleaning, 10-9 Auxiliary Devices General Cleaning, 10-8 Connection Procedures &...
  • Page 433 GE H LOGIQ™ S7 E EALTHCARE XPERT 5460683, R IRECTION EVISION ERVICE ANUAL Printer Installation, 3-30 Printer Installation manually, 3-34 Warnings and Cautions, 1-7 Printer Settings What is EMC?, 1-13 Adjustment, 3-34 Worksheet for DICOM Network Information, 3-59 Probe Connector Cleaning, 10-11...
  • Page 435 2013, 2014 General Electric Company. © GE Ultrasound Korea, Ltd. 9, Sunhwan-ro 214beon-gil, Jungwon-gu, Seongnam-si, Gyeonggi-do, Korea www.gehealthcare.com...

This manual is also suitable for:

Logiq s7 pro

Table of Contents

Save PDF